# Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

# Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

## Key Features At A Glance

-   Five simultaneous receiver inputs and five measurement receivers
-   Measurement speed up to 400,000 points/sec at 600 kHz IFBW, CW (Fast CW mode, no point triggering)
-   Selectable IF bandwidth from 1 Hz to 5 MHz
-   4-billion-byte data buffer; up to 500 million points per single cut (single parameter)
-   Ethernet (LAN), USB, GPIB and RS-232 host interfaces for automated test systems
-   Removable hard drive for secure environments
-   Optional internal LO output (Option 108) covering 10 MHz to 26.5 GHz
-   Five external IF inputs (A, B, C, D, R) on 50 Ohm SMA (female) connectors

The Agilent N5264A is a PNA-X measurement receiver that provides five simultaneous receiver inputs and five measurement receivers in a single instrument. It is built around external IF inputs that accept signals from remote mixers or frequency converters, with a selectable IF bandwidth spanning 1 Hz to 5 MHz.

The external IF inputs allow the use of external IF signals from remote mixers or frequency converters, presented as five inputs labeled A, B, C, D, and R on 50 Ohm SMA (female) connectors. The instrument exposes Ethernet, USB, and GPIB host computer interfaces, and its frequency source control interface uses a TTL handshake for coordinating an external source.

ValueTronics International stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

## Brand Heritage

Hewlett-Packard's test and measurement business was spun out as Agilent Technologies in 1999. Agilent's electronic measurement business was subsequently established as Keysight Technologies in 2014.

## Product Core & Specifications

Specification

Value

Key Specifications

Measurement Speed (max) points/sec @ 600 KHz IFBW, CW frequency

400,000 points/sec (Fast CW mode - no point triggering)

Receiver Inputs

5 (simultaneously)

Measurement Receivers

5 (simultaneously)

Data Buffer Size

4 billion bytes

Data Buffer size (max. points for single cut)

500 million points (for single parameter; two parameters are 250 million points each)

IF Bandwidth

1 Hz to 5 MHz

Frequency Source Control Interface

TLL hand shake

Trigger In / Out

Three pairs

Host Computer Interface

Ethernet, USB and GPIB

Security

Hard drive removable

External IF Inputs

Function

Allows use of external IF signals from remote mixers or frequency converters

Connectors

SMA (female); A, B, C, D, R

Input Frequency

7.605634 MHz

Input Impedance

50 Ω

RF Damage Level

+23 dBm

DC Damage Level

1 VDC

0.1 dB Compression Point

\-9.0 dBm

Compression @ -10 dBm, Magnitude

0.03 dB

Compression @ -10 dBm, Phase

0.23°

Noise Floor, 10 Hz IF BW

\-143 dBm

Noise Floor, 10 KHz IF BW

\-113 dBm

Crosstalk

\-134 dB

Dynamic Range @ 10 Hz

134 dB @ 0.1dB compression to noise floor

Dynamic Accuracy (-40 dBm reference, over range set by compression and noise floor @ IF Frequencies)

\-10 dBm

0.037 dB

\-20 dBm

0.024 dB

\-30 dBm

0.016 dB

\-40 dBm

0.010 dB

\-50 dBm

0.013 dB

\-60 dBm

0.021 dB

\-70 dBm

0.032 dB

\-80 dBm

0.041 dB

\-90 dBm

0.049 dB

\-100 dBm

0.057 dB

\-110 dBm

0.072 dB

\-120 dBm

0.188 dB

LO output 2 (Option 108)

Frequency Stability

+/- 0.05 ppm, -10 to 70C, +/- 0.1ppm/yr max

Frequency Accuracy

+/- 1 ppm

Frequency Range

10 MHz to 26.5 GHz

Frequency Switching Speed

< 100 microsecond/point

Frequency Resolution

1 Hz

Power Flatness

+/- 1.0 dB

Power Output

+10 dBm

2nd Harmonics, 20 MHz to 2.0 GHz

\-23 dBc

2nd Harmonics, 2.0 GHz to 5.0 GHz

\-28 dBc

2nd Harmonics, 5.0 GHz to 23.0 GHz

\-35 dBc

2nd Harmonics, 23.0 GHz to 26.5 GHz

\-27 dBc

3rd Harmonics, 30 MHz to 8.0 GHz

\-32 dBc

3rd Harmonics, 8.0 GHz to 15.0 GHz

\-38 dBc

3rd Harmonics, 15.0 GHz to 26.5 GHz

\-48 dBc

Phase Noise, 10 MHz to 500 MHz (1/10/100 kHz, 1 MHz offset)

\-80 / -85 / -76 / -113 dBc/Hz

Phase Noise, 500 MHz to 1 GHz (1/10/100 kHz, 1 MHz offset)

\-90 / -110 / -106 / -115 dBc/Hz

Phase Noise, 1 GHz to 2 GHz (1/10/100 kHz, 1 MHz offset)

\-85 / -105 / -101 / -110 dBc/Hz

Phase Noise, 2 GHz to 4 GHz (1/10/100 kHz, 1 MHz offset)

\-80 / -100 / -96 / -105 dBc/Hz

Phase Noise, 4 GHz to 8 GHz (1/10/100 kHz, 1 MHz offset)

\-74 / -94 / -90 / -99 dBc/Hz

Phase Noise, 8 GHz to 16 GHz (1/10/100 kHz, 1 MHz offset)

\-68 / -88 / -84 / -93 dBc/Hz

Phase Noise, 16 GHz to 26.5 GHz (1/10/100 kHz, 1 MHz offset)

\-62 / -82 / -78 / -87 dBc/Hz

10 MHz Reference

Reference In Connector

BNC, female

Reference In Input Frequency

10 MHz ± 10 ppm, typical

Reference In Input Level

\-15 dBm to +20 dBm, typical

Reference In Input Impedance

200 Ω, nom.

Reference Out Connector

BNC, female

Reference Out Output Frequency

10 MHz ± 1 ppm, typical

Reference Out Signal Type

Sine Wave, typical

Reference Out Output Level

+10 dBm ± 4 dB into 50 Ω

Reference Out Output Impedance

50 Ω, nominal

Reference Out Harmonics

<-40 dBc, typical

External Monitor Information

VGA Video Output Connector

15-pin mini D-Sub; Drives VGA compatible monitors

Flat Panel (TFT) Resolutions

1024 X 768, 800 X 600, 640 X 480

Flat Panel (DSTN) Resolutions

800 X 600, 640 X 480

CRT Monitor Resolutions

1280 X 1024, 1024 X 768, 800 X 600, 640 X 480

Test Set IO

25-pin D-Sub connector, available for external test set control

Power IO

9-pin D-Sub, female; analog and digital IO

Handler IO

36-pin parallel I/O port; all input/output signals are default set to negative logic; can be reset to positive logic via GPIB command

Trigger Information

Trigger In/Meas Trigger, Nominal Input Impedance

5K Ohms

Trigger In/Meas Trigger, Minimum Pulse Width

1 us

Trigger In/Meas Trigger, DC Damage Level

5.5 volts

Trigger In/Meas Trigger, Drive Voltage

TTL (0, +5.0) Volts

Trigger Out/Meas Trigger Ready, Nominal Input Impedance

5K Ohm

Trigger Out/Meas Trigger Ready, Pulse Width

\= Data acquisition

Trigger Out/Meas Trigger Ready, Polarity

Selectable with sweep or point mode

Trigger Out/Meas Trigger Ready, Drive Voltage

TTL (0, +5.0) Volts

Trigger Inputs/Outputs (Aux. 1 & 2)

BNC(f), TTL/CMOS compatible

GPIB

24-pin D-sub (Type D-24), female; compatible with IEEE-488 (two ports - dedicated controller and dedicated talker/listener)

Parallel Port (LPT1)

25-pin D-Sub miniature connector, female; provides connection to printers or any other parallel port peripherals

Serial Port (COM 1)

9-pin D-Sub, male; compatible with RS-232

USB Port

Four ports on front panel (all Host) and five ports (four hosts and one Device) on rear panel. Type A configuration (eight hosts) and Type B configuration (one Device), USB 2.0 compatible

LAN

10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates

Line Power

Frequency, Voltage

50/60 Hz for 100 240 VAC (auto switching)

Max

450 watts

> **Important:** External IF input damage levels: +23 dBm RF and 1 VDC.

> **Important:** Simultaneous operation of the internal and external displays is allowed, but with 640 X 480 resolution only. If you change resolution, you can only view the external display (internal display will "white out").

> **Important:** All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument has been turned on.

> **Important:** When configuring the N5264A Option 108 as the LO source, you may improve system measurement sensitivity by using a method of AM noise suppression.

> **Important:** For Regulatory and Environmental information, refer to the PNA Series Installation and Quick Start Guide, located online at http://cp.literature.agilent.com/litweb/pdf/E8356-90001.pdf.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Agilent
- **Type:** Net Ana
- **Tags:** product_manual_group_3

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/n5264a-keysight-network-analyzer-used)
> Updated: 2026-06-20