# Agilent N3383A 300 kHz to 9 GHz 3-Port PNA Series Network Analyzer (Pre-Owned)

# Agilent N3383A 300 kHz to 9 GHz 3-Port PNA Series Network Analyzer (Pre-Owned)

## Key Features At A Glance

-   Three-port network analyzer family spanning 300 kHz to 9 GHz across the N3381A, N3382A, and N3383A models
-   GPIB (IEEE-488), 10/100BaseT LAN, USB, and RS-232 interfaces standard for automated test system integration
-   Type-N (female) 50 Ω test ports; Type-N, 3.5 mm, and 7-16 calibration kits and ECal modules supported
-   8.4-inch (21.3 cm) color active-matrix LCD at 640 x 480 resolution, with VGA video output for an external monitor
-   Selectable IF bandwidth from 1 Hz to 40 kHz to trade sweep speed against noise floor
-   Corrected system dynamic range up to 128 dB at the test port (25 MHz to 3 GHz)
-   Optional step attenuator (Opt 1E1) extends source power down to -85 dBm; optional high-stability timebase (Opt 1E5)

The PNA Series Network Analyzers comprise three models — the N3381A, N3382A, and N3383A — covering 300 kHz to 3.0 GHz, 300 kHz to 6.0 GHz, and 300 kHz to 9.0 GHz respectively. Each is a three-port S-parameter measurement system that characterizes the reflection and transmission behavior of RF components through full two-port calibrated measurements.

A vector network analyzer stimulates a device under test with a swept RF source and measures the magnitude and phase of the signals reflected from and transmitted through that device, yielding the S-parameters that describe its linear behavior. Across the series these measurements are made over selectable IF bandwidths, with error correction applied through a calibration of the test setup using known standards.

ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received, giving buyers a verified, accountable source for both new and pre-owned test and measurement equipment.

## Brand Heritage

Agilent Technologies originated as the test-and-measurement business of Hewlett-Packard, which spun it out as a separate company in 1999. In 2014, Agilent's electronic measurement business was established as Keysight Technologies, which carries the product line forward today.

## Compare Other Models in This Series

The PNA Series Network Analyzers are offered as three models that share a common architecture and differ in their upper frequency limit: the N3381A reaches 3.0 GHz, the N3382A reaches 6.0 GHz, and the N3383A reaches 9.0 GHz, each starting at 300 kHz. All three are three-port S-parameter systems built on the same source, receiver, and display platform.

Because the models share the same calibration-kit support, connectivity, IF-bandwidth range, and front- and rear-panel layout, the choice between them is driven primarily by the frequency coverage a given device under test requires.

Each pre-owned model below links to its own dedicated product page with condition-matched pricing. Select the model whose frequency coverage fits your application, then review the pre-owned listing for that specific unit.

The N3381A covers 300 kHz to 3.0 GHz, the lowest of the three upper-frequency limits in the series, while sharing the three-port test set, source power range, and IF-bandwidth settings common to the family.

The primary difference across the family is maximum frequency — 3.0 GHz, 6.0 GHz, and 9.0 GHz for the N3381A, N3382A, and N3383A respectively — as shown in the comparison table below. Source power range reflects this as well: all models provide -15 to +10 dBm up to 6 GHz, with the 6 to 9 GHz band (applicable to the N3383A) specified at -15 to +7 dBm.

Options apply across the family, including the 1E1 step attenuator that extends the source power range down to -85 dBm, the 1E5 high-stability timebase, and Option 014 for receiver access. Refer to the comparison table for model-by-model frequency coverage.

Model

Frequency Range

Test Ports

Connector Type

**N3383A**

300 kHz to 9.0 GHz

3

Type-N, 50 Ω

N3381A

300 kHz to 3.0 GHz

3

Type-N, 50 Ω

N3382A

300 kHz to 6.0 GHz

3

Type-N, 50 Ω

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Specification

Value

Test Port Output Frequency

Range (N3383A)

300 kHz to 9.0 GHz

Resolution

1 Hz

Source Stability

±1 ppm, 0° to 40 °C, typical; ±1 ppm/year maximum

Source Stability (Option 1E5)

±0.05 ppm, 0° to 40 °C, typical; ±0.1 ppm/year maximum

CW Accuracy

±3 ppm

CW Accuracy (Option 1E5)

±1 ppm

System Dynamic Range

Dynamic range at test port, 300 kHz to 25 MHz

125 dB (specification)

Dynamic range at test port, 25 MHz to 3 GHz

128 dB (specification)

Dynamic range at test port, 3 GHz to 6 GHz

118 dB (specification)

Dynamic range at test port, 6 GHz to 9 GHz

115 dB (specification)

Dynamic range at receiver input, 300 kHz to 25 MHz

140 dB (characteristic)

Dynamic range at receiver input, 25 MHz to 3 GHz

143 dB (characteristic)

Dynamic range at receiver input, 3 GHz to 6 GHz

133 dB (characteristic)

Dynamic range at receiver input, 6 GHz to 9 GHz

130 dB (characteristic)

Corrected System Performance — 85032F Type-N Calibration Kit (full 2-port)

Directivity, 300 kHz to 1.3 GHz

49 dB

Directivity, 1.3 GHz to 3 GHz

46 dB

Directivity, 3 to 6 GHz

40 dB

Directivity, 6 to 9 GHz

38 dB

Source Match, 300 kHz to 1.3 GHz

41 dB

Source Match, 1.3 GHz to 3 GHz

40 dB

Source Match, 3 to 6 GHz

36 dB

Source Match, 6 to 9 GHz

35 dB

Load Match, 300 kHz to 1.3 GHz

49 dB

Load Match, 1.3 GHz to 3 GHz

45 dB

Load Match, 3 to 6 GHz

39 dB

Load Match, 6 to 9 GHz

37 dB

Reflection Tracking, 300 kHz to 1.3 GHz

±0.011 dB

Reflection Tracking, 1.3 GHz to 3 GHz

±0.021 dB

Reflection Tracking, 3 to 6 GHz

±0.032 dB

Reflection Tracking, 6 to 9 GHz

±0.054 dB

Transmission Tracking, 300 kHz to 1.3 GHz

±0.012 dB

Transmission Tracking, 1.3 GHz to 3 GHz

±0.020 dB

Transmission Tracking, 3 to 6 GHz

±0.055 dB

Transmission Tracking, 6 to 9 GHz

±0.083 dB

Corrected System Performance — 85092C Type-N ECal Module (full 2-port)

Directivity, 1.3 GHz to 3 GHz

54 dB

Directivity, 3 to 6 GHz

52 dB

Source Match, 300 kHz to 1.3 GHz

45 dB

Source Match, 3 to 6 GHz

41 dB

Source Match, 6 to 9 GHz

36 dB

Load Match, 6 to 9 GHz

39 dB

Reflection Tracking, 300 kHz to 1.3 GHz

±0.040 dB

Reflection Tracking, 3 to 6 GHz

±0.060 dB

Transmission Tracking, 1.3 GHz to 3 GHz

±0.039 dB

Transmission Tracking, 6 to 9 GHz

±0.136 dB

Corrected System Performance — 85033E 3.5 mm Calibration Kit (full 2-port)

Directivity, 300 kHz to 1.3 GHz

46 dB

Directivity, 1.3 GHz to 3 GHz

44 dB

Directivity, 3 to 6 GHz

38 dB

Directivity, 6 to 9 GHz

38 dB

Source Match, 300 kHz to 1.3 GHz

43 dB

Source Match, 1.3 GHz to 3 GHz

40 dB

Source Match, 3 to 6 GHz

37 dB

Source Match, 6 to 9 GHz

36 dB

Load Match, 300 kHz to 1.3 GHz

46 dB

Load Match, 1.3 GHz to 3 GHz

44 dB

Load Match, 3 to 6 GHz

38 dB

Load Match, 6 to 9 GHz

38 dB

Reflection Tracking, 300 kHz to 1.3 GHz

±0.006 dB

Reflection Tracking, 1.3 GHz to 3 GHz

±0.007 dB

Reflection Tracking, 3 to 6 GHz

±0.009 dB

Reflection Tracking, 6 to 9 GHz

±0.010 dB

Transmission Tracking, 300 kHz to 1.3 GHz

±0.012 dB

Transmission Tracking, 1.3 GHz to 3 GHz

±0.021 dB

Transmission Tracking, 3 to 6 GHz

±0.057 dB

Directivity, 300 kHz to 1.3 GHz

52 dB

Directivity, 6 to 9 GHz

47 dB

Source Match, 1.3 GHz to 3 GHz

44 dB

Load Match, 300 kHz to 1.3 GHz

47 dB

Load Match, 1.3 GHz to 3 GHz

47 dB

Load Match, 3 to 6 GHz

44 dB

Reflection Tracking, 1.3 GHz to 3 GHz

±0.040 dB

Reflection Tracking, 6 to 9 GHz

±0.070 dB

Transmission Tracking, 300 kHz to 1.3 GHz

±0.039 dB

Transmission Tracking, 3 to 6 GHz

±0.068 dB

> **Important:** Test port damage level: +30 dBm or ±30 VDC, typical (ports 1, 2, 3). With Option 014, the R, A, B, and C receiver inputs are rated at +15 dBm or ±5 VDC, typical.

> **Important:** Test port dynamic range specification applies only when power is sourced from Port 1; if power is sourced from either Port 2 or Port 3, dynamic range decreases by 3 dB.

> **Important:** Source output power performance is specified on Port 1 only; Port 2 and Port 3 output performance is typically 3 dB less.

> **Important:** Dynamic range may be limited at particular frequencies below 750 MHz due to spurious receiver residuals (to 100 dB at the test port and 115 dB at the receiver input).

> **Important:** Line power requires a third-wire ground; the power supply has a voltage autoswitching feature.

> **Important:** Option H08 and Option H11 specifications are not provided in this N3381/2/3A specifications document.

> Recommended pairing: full two-port calibration uses the N6314A test port cable with a compatible kit or ECal module (85032F or 85092C for Type-N, 85033E or 85093C for 3.5 mm, 85038A for 7-16).

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 600.0 USD
- **Vendor:** Agilent Technologies
- **Type:** Net Ana
- **Tags:** Agilent, Agilent Keysight, Analyzers, Manufacturers, Network, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 525.00 USD | In stock |
| NIST Traceable with Full Data | 562.50 USD | In stock |
| ISO IEC 17025 Accredited | 600.00 USD | In stock |

## Images

- Agilent N3383A 300 kHz to 9 GHz 3-Port PNA Series Network Analyzer (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/n3383a-agilent-network-analyzer-used)
> Updated: 2026-07-06