# Anritsu MS4630A 10 Hz to 300 MHz Network Analyzer (Pre-Owned)

# Anritsu MS4630A 10 Hz to 300 MHz Network Analyzer (Pre-Owned)

## Key Features At A Glance

-   Frequency range 10 Hz to 300 MHz with 0.01 Hz resolution
-   Measurement range ≥120 dB dynamic range (RBW: 1 kHz) for fast out-of-band filter attenuation measurement
-   Fast 150 µs/measurement-point sweep speed using high-speed synthesizer and DSP processing
-   High-accuracy group delay measurement at 1/10,000-of-range resolution
-   Up to 10 independent multi-markers per trace with tabular marker-list display
-   Built-in filter (3 dB BW, fo, ripple, attenuation) and resonator (Fr, Zr, equivalence) analysis functions
-   Single and segmented limit-test functions for real-time pass/fail device evaluation
-   GPIB (IEEE488.2) and PTA parallel interface standard; RS-232C/Centronics via Option 02

The MS4630A is a network analyzer covering 10 Hz to 300 MHz, designed for electronics production lines that demand fast, accurate device measurements. It is well suited to high-speed evaluation of IF filter resonance and group delay characteristics, and to evaluating the impedance characteristics of resonators used in AV equipment and personal computers. A fast sweep speed of 150 µs per measurement point is achieved through a high-speed synthesizer and digital signal processing, while strengthened post-processing data-analysis macros increase total production throughput.

Its listed applications center on filter and resonator characterization on production and inspection lines. The instrument measures filter characteristics such as 3 dB bandwidth, center frequency (fo), in-band ripple, and out-of-band attenuation, which are digitally processed and analyzed at high speed. Dedicated resonator analysis functions evaluate resonance frequency (Fr) and resonance impedance (Zr), with Resonator 2 adding resonator equivalence measurement. Single and segmented limit-test functions provide real-time pass/fail evaluation of devices under test.

ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is fulfilled directly from that facility; used units are inspected and functionally verified in-house against the original manufacturer's specifications before they ship, while new units ship factory-sealed exactly as received.

## Accessories Supplied

-   J0017 Power cord, 2.5 m: 1 pc
-   F0013 Fuse, 5 A: 2 pcs
-   W1248AE MS4630A operation manual (main frame): 1 copy
-   W1249AE MS4630A operation manual (remote control): 1 copy

## Product Core & Specifications

Specification

Value

Measurement Items

Transmission characteristics (ratio measurement)

Amplitude, phase, group delay

Reflection/impedance characteristics

Amplitude, phase (with external transducer)

Level characteristics

Absolute amplitude

Frequency

Range

10 Hz to 300 MHz

Resolution

0.01 Hz

Aging rate (standard)

≤1 x 10–6/day (15 minutes after power-on)

Temperature characteristics (standard)

≤±5 x 10–6 (0° to 50°C)

Aging rate (Option 13: High-stability reference oscillator)

≤±2 x 10–8/day (24 h after power-on)

Temperature characteristics (Option 13)

≤±5 x 10–8 (0° to 50°C)

Input

Channel No.

Standard: 2 (R, TA); Option 12: 3 (R, TA, TB)

Impedance

50 Ω, 1 MΩ switchable (when combined with MA4605A: 75 Ω, 1 MΩ)

Input range (IRG)

0/+20 dBm

Max. input power

AC: +20 dBm; DC ±2.2 V (50 Ω); AC: 0 dBm; DC: ±20 V (1 MΩ)

Connector

BNC-J

Probe source

+12 ±1 V, 100 mA (with protective circuit for shorts)

Average noise level

≤–120 dBm (RBW: 1 kHz, 1 to 300 MHz), ≤–110 dBm (RBW: 1 kHz, 80 kHz to 1 MHz)

Crosstalk between channels

≥120 dB (80 kHz to 300 MHz), ≥110 dB (up to 80 kHz)

Crosstalk between transmitter and receiver

≥125 dB

Resolution bandwidth

3, 10, 30, 100, 500 Hz, 1, 2, 3, 4, 5, 10, 20 kHz and automatic setting

Output

Output level range, Output A

0 to +21 dBm; Option 10: –70 to +21 dBm

Output level range, Output B

–6 to +15 dBm (–9.5 to +11.5 dB when Option 14 added); Option 10: –76 to +15 dBm (–79.5 to +11.5 dB when Option 14 added)

Output resolution

0.01 dB

Output level accuracy

≤±1.0 dB (frequency: 100 MHz, Output A: +10 dBm)

Output level linearity

≤±0.5 dB (0 dBm reference, frequency: 100 MHz, Output A: 0 to +21 dBm)

Output level deviation

≤±1.5 dB (output A: +10 dBm, 100 MHz reference)

Step error

±0.5 dB (Option 10)

Output impedance

50 Ω (when combined with MA4605A: 75 Ω)

Connector

BNC-J

Amplitude Measurement

Measurement range

≥120 dB

Measurement resolution

0.001 dB

Display scale

0.01 dB/div to 50 dB/div (1-2-5 sequence)

Dynamic accuracy, 0 to –10 dB

±0.30 dB (80 kHz to 100 MHz) / ±0.30 dB (10 kHz to 300 MHz)

Dynamic accuracy, –10 to –60 dB

±0.05 dB (80 kHz to 100 MHz) / ±0.05 dB (10 kHz to 300 MHz)

Dynamic accuracy, –60 to –70 dB

±0.10 dB (80 kHz to 100 MHz) / ±0.30 dB (10 kHz to 300 MHz)

Dynamic accuracy, –70 to –80 dB

±0.30 dB (80 kHz to 100 MHz) / ±1.00 dB (10 kHz to 300 MHz)

Dynamic accuracy, –80 to –90 dB

±1.20 dB (80 kHz to 100 MHz) / ±4.00 dB (10 kHz to 300 MHz)

Dynamic accuracy, –90 to –100 dB

±4.00 dB (80 kHz to 100 MHz) / — (10 kHz to 300 MHz)

Phase Measurement

Measurement range

±180°

Measurement resolution

0.001°

Display scale

0.01° to 50°/div (1-2-5 sequence)

Dynamic accuracy, 0 to –10 dB

±6.0° (80 kHz to 100 MHz) / ±6.0° (10 kHz to 300 MHz)

Dynamic accuracy, –10 to –60 dB

±0.3° (80 kHz to 100 MHz) / ±0.3° (10 kHz to 300 MHz)

Dynamic accuracy, –60 to –70 dB

±0.8° (80 kHz to 100 MHz) / ±2.0° (10 kHz to 300 MHz)

Dynamic accuracy, –70 to –80 dB

±2.0° (80 kHz to 100 MHz) / ±6.0° (10 kHz to 300 MHz)

Dynamic accuracy, –80 to –90 dB

±6.0° (80 kHz to 100 MHz) / ±20.0° (10 kHz to 300 MHz)

Dynamic accuracy, –90 to –100 dB

±20.0° (80 kHz to 100 MHz) / — (10 kHz to 300 MHz)

Group Delay Measurement

DRG

Δθ/(360 x ΔF)

Measurement resolution

2.78 x 10–5/ΔF

Display scale

1 ps/div to 50 ms/div

Dynamic accuracy

Phase measurement accuracy/(360 x aperture frequency)

Calibration, Correction

Calibration types

Frequency response, 1 port, 1 path-2 port, frequency response/isolation calibration, π-NET calibration

Calibration data interpolation

Measurement frequency, when number of measurement points changed, based on calibration data before change, new calibration data interpolation calculation possible (except at log frequency measurement and 1001 measurement points)

Normalize

X–S

Electrical length calibration range

0 to ±999999.9999999 m, Resolution: 100 nm

Phase offset range

±180°

Sweeping

Frequency sweep

LIN (CENTER/SPAN, START/STOP), LOG (START/STOP)

Level sweep

LIN (START/STOP/STEP)

Number of measurement points

11, 21, 51, 101, 251, 501, 1001

Break point

Anywhere between 1 and 1001

Sweep time

150 µs/point, 38 ms/250 points full sweep (RBW: 20 kHz, normalize calibration, 1 trace)

Setting range

1 ms to 27.5 h

Sweep range

Full sweep, part sweep (between markers)

Sweep control

REPEAT/SINGLE, STOP/CONT

Sweep trigger

INT/EXT (RISE, FALL, LEVEL)

Display

Max. display screens

2 channels, 4 traces

Display format

LOG MAG (M), PHASE (P), DELAY (D), M/P, M/D, LIN MAG (LIN), LIN/P, LIN/D, REAL (R), IMAG (I), R/I, Z, Z/θ, Q, Z/Q, POLAR, VSWR, IMPD (Z∠θ, Rs + Ls/Cs, Q/D, R + jx), ADMT (Y∠θ, Rp + Lp/Cp, Q/D, G + jB)

Display

640 x 480 dots, 6.5” color LCD

Markers

Marker functions

NORMAL MKR, ΔMKR, 0 MKR, MKR → MAX, MKR → MIN, MKR → CF, Δ→ SPAN, MKR → +PEAK, MKR →–PEAK, MKR TRACK + PEAK, MKR TRACK–PEAK, MKR CHANGE, MKR OFFSET

Setting

Set marker position to frequency or point

Multi-marker

Max. 10 markers for each trace

Measurement Parameters

Filter function

F0, IL, passband (L, R), attenuation band (L, R), Ripple, Q, SF

Resonator function RESON 1

Fr, Fa, Zr, Za (0 PHASE), Fm, Fn, Zm, Zn (MAX/MIN)

Resonator function RESON 2

Fs, Fr, Fa, Zr, Za, Q, equivalence constant (R1, L1, C1, C0)

Averaging functions

Method: SUM, MAX, MIN, Count: 1 to 1000

Limit line

Single or segment (10) limit line, pass/fail evaluation against limit line

Receive bandwidth and sweep time auto-setting

Receive bandwidth set automatically for set sweep time; automatically set to give minimum sweep time at set receive bandwidth

Trace Data Calculation

Measurement data memory

Max. 1001 points each memory in same format as display format

> **Important:** Maximum input power: +20 dBm AC / ±2.2 V DC (50 Ω); 0 dBm AC / ±20 V DC (1 MΩ).

> **Important:** Reflection/impedance measurements are made with an external transducer. The standard delivery includes only the power cord, fuses, and operation manuals; a Reflection Bridge (62-series or MA2000-series) or the MA4605A Impedance Adapter must be ordered separately to perform reflection/impedance measurements (see Optional accessories).

> Recommended pairing: use the MA4605A Impedance Adapter (10 Hz to 300 MHz, 50/75 Ω) for impedance measurement, or one of the listed reflection bridges for reflection/impedance characteristics.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 600.0 USD
- **Vendor:** Anritsu
- **Type:** Net Ana
- **Tags:** Analyzers, Anritsu, Manufacturers, Network, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 525.00 USD | In stock |
| NIST Traceable with Full Data | 562.50 USD | In stock |
| ISO IEC 17025 Accredited | 600.00 USD | In stock |

## Images

- Anritsu MS4630A 10 Hz to 300 MHz Network Analyzer (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/ms4630a-anritsu-network-analyzer-used)
> Updated: 2026-07-06