# Anritsu MS2724B 20 GHz Handheld Spectrum Analyzer (Pre-Owned)

# Anritsu MS2724B 20 GHz Handheld Spectrum Analyzer (Pre-Owned)

## Key Features At A Glance

-   Continuous 9 kHz to 20 GHz frequency coverage with 1 Hz tuning resolution
-   Resolution bandwidth from 1 Hz to 3 MHz, with 200 Hz/9 kHz/120 kHz CISPR quasi-peak bandwidths
-   Low SSB phase noise of –104 dBc/Hz typical at 10 kHz offset at 2 GHz
-   Built-in preamplifier, 65 dB step attenuator, and True RMS detection
-   Built-in AM/FM/SSB demodulator
-   LAN 10/100 (RJ45), USB 2.0 host/device, and BNC external reference and trigger interfaces
-   Ultra-portable at 3.4 kg (7.5 lb) with an 8 in. daylight-viewable SVGA color display
-   Optional GPS receiver and 3G modulation measurement packages

The Anritsu MS2724B Spectrum Master is an ultra-portable handheld spectrum analyzer providing continuous frequency coverage from 9 kHz to 20 GHz. Built for the wireless professional, it delivers the performance needed for demanding measurements in harsh RF and physical environments while remaining light enough to carry into the field at 3.4 kg (7.5 lb).

The MS2724B is suited to spectrum monitoring, WiFi and WiFi5 installation and testing, RF and microwave signal measurements, and cellular signal measurements. Its combination of wide frequency reach and portable form factor lets engineers and technicians perform these tasks on-site rather than only at a fixed bench.

ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

## Accessories Supplied

-   10580-00175 User's Guide
-   65729 Soft Carrying Case
-   40-168-R AC – DC Adapter
-   806-141 Automotive Cigarette Lighter/12 Volt DC Adapter
-   2300-498 CD ROM containing Master Software Tools
-   2000-1371 Ethernet Cable
-   3-806-152 Cross-over Ethernet Cable
-   633-44 Rechargeable battery, Li-Ion
-   1091-27 Type-N male to SMA female adapter
-   1091-172 Type-N male to BNC female adapter
-   2000-1520-R 2 GB USB Flash Drive
-   3-2000-1498 USB Type A to Mini-B Cable
-   One Year Warranty

## Product Core & Specifications

Specification

Value

Frequency

Frequency Range

9 kHz to 20 GHz

Preamp

100 kHz to 4 GHz

Tuning Resolution

1 Hz

Frequency Reference Aging

±1 ppm per 10 years

Frequency Reference Accuracy

±0.3 ppm (25º C ± 25º C) + aging

Frequency Span

10 Hz to 20 GHz plus 0 Hz (zero span)

Span Accuracy

Same as frequency reference accuracy

Sweep Time (Zero span)

10 µs to 600 s

Sweep Time (Spans >0 Hz)

Automatically optimized; can be manually increased

Sweep Time Accuracy

±2% in zero span

Sweep Trigger

Free run, Single, Video, External

Resolution Bandwidth (–3 dB)

1 Hz to 3 MHz in 1-3 sequence ±10%, 200 Hz, 9 kHz, 120 kHz when quasi-peak detector selected, 10 MHz demodulation bandwidth

Video Bandwidth (–3 dB)

1 Hz to 3 MHz in 1-3 sequence

SSB Phase Noise (Max)

10, 20 and 30 kHz offset

–95 dBc/Hz (9 kHz to 13 GHz); –91 dBc/Hz (13 GHz to 20 GHz)

100 kHz offset

–97 dBc/Hz (9 kHz to 13 GHz); –93 dBc/Hz (13 GHz to 20 GHz)

1 MHz offset

–105 dBc/Hz (9 kHz to 13 GHz); –102 dBc/Hz (13 GHz to 20 GHz)

10 MHz offset

–120 dBc/Hz (9 kHz to 13 GHz); –116 dBc/Hz (13 GHz to 20 GHz)

Amplitude

Measurement Range

DANL to +30 dBm

Display Range

1 to 15 dB/div in 1 dB steps. Ten divisions displayed.

Amplitude Units (Log Scale Modes)

dBm, dBV, dBmv, dBµV

Amplitude Units (Linear Scale Modes)

nV, µV, mV, V, kV, nW, µW, mW, W, kW

Attenuator Range

0 to 65 dB

Attenuator Resolution

5 dB steps

Overall Amplitude Accuracy (20º C to 30º C, 30 minute warmup)

±1.3 dB

Full Temperature Range (–10 to +55º C)

add ±1.2 dB

Frequency Flatness (>4 GHz)

add ±1.5 dB

Conditions

50 Ω source, single sinewave input ≤Reference Level, and ≥DANL, 60 minute warm-up, auto-attenuation

Second Harmonic Distortion (0 dB input attenuation, –30 dBm input)

50 to 500 MHz

–50 dBc

500 to 800 MHz

–45 dBc

800 to 3000 MHz

–60 dBc

\>3 GHz

–70 dBc

Third Order Intercept (TOI) (–20 dBm tones 100 kHz apart, –20 dBm Ref level, 0 dB input attenuation, preamplifier off)

2.4 GHz (Min)

+12 dBm

50 MHz to 500 MHz (Typical)

\>6 dBm

500 MHz to 2 GHz (Typical)

\>8 dBm

2 to 6 GHz (Typical)

\>10 dBm

6 to 20 GHz (Typical)

\>12 dBm

Dynamic Range 2/3 (TOI-DANL) in 1 Hz RBW

2.4 GHz

101 dB min

Displayed Average Noise Level (DANL) in 1 Hz RBW (Preamplifier On)

10 MHz to 1 GHz

–159 dBm (Equivalent Noise Figure 15 dB, 23º C)

1 GHz to 3 GHz

–156 dBm (18 dB)

3 to 4 GHz

–154 dBm (20 dB)

Displayed Average Noise Level (DANL) in 1 Hz RBW (Preamplifier Off)

10 MHz to 4 GHz

–139 dBm (35 dB)

4 GHz to 10 GHz

–136 dBm (38 dB)

10 GHz to 13 GHz

–130 dBm (44 dB)

13 GHz to 20 GHz

–136 dBm (38 dB)

Spurious Response

Input-Related Spurious (–30 dBm input, 0 dB input attenuation, Span <1.7 GHz)

–70 dBc typical; –60 dBc max; except input frequency 3275 MHz, –50 dBc max

Residual Spurious (Preamplifier off, RF input terminated, 0 dB input attenuation)

–90 dBm max; –85 dBm max, >13 GHz

Residual Spurious (Preamplifier on, RF input terminated, 0 dB input attenuation)

–100 dBm max

General

RF Input VSWR (≥10 dB input attenuation)

1.5:1 typical <13 GHz; 2:1 typical 13 to 20 GHz

Maximum Continuous Input (≥10 dB input attenuation)

+30 dBm

Input Damage Level

≥10 dB input attenuation, >+30 dBm, ±50 Vdc

ESD Damage Level (≥10 dB input attenuation)

\>10 kV

External Reference Frequencies

1, 1.2288, 1.544, 2.048, 2.4576, 4.8, 4.9152, 5, 9.8304, 10, 13 and 19.6608 MHz at –10 dBm to +10 dBm

Battery Life

2.3 hours typical

Display

Display

Bright daylight-viewable color transmissive LCD: Full SVGA, 8 in.

Languages

Built-in Languages

English, Spanish, Italian, French, German, Japanese, Korean, and Chinese. Two customized languages can be installed from Master Software Tools.

Marker Modes

Markers

6 Markers, 9 Modes: Normal, Delta, Marker to Peak, Marker to Center, Marker to Reference Level, Next Peak Left, Next Peak Right, All Markers Off, Noise Marker, Frequency Counter Marker (1 Hz resolution), Markers Tracking or Fixed, Marker 1 reference for all deltas.

Sweeps

Sweeps

Full span, Zero span, Span Up/Span Down

Detection

Detection

Peak, Negative peak, Sample, RMS, Quasi-peak

Memory

Trace and Setup Storage

Limited only by the capacity of the installed Compact Flash card or USB Flash drive. For a 256 MB card, storage is greater than 13000 spectrum analyzer traces and over 10000 setups.

Traces

Displayed Traces

Three Traces with trace overlay. Trace A is always the live data; Traces B and C can be either stored data or traces which have been mathematically manipulated. Trace C can show max hold or min hold.

Interfaces

Interfaces

Type N female RF connector for Spectrum Analyzer input; Reverse polarity BNC jack for optional GPS antenna connector; BNC female connectors for ext. reference and ext. trigger; 5-pin Mini-B USB 2.0 for data transfer to a PC; USB 2.0 Host connector used with PSN50 High Accuracy Power Meter and USB Flash Drives; RJ45 connector for Ethernet 10/100 Base T; 2.5 mm 3-wire headset connector

Size and Weight

Size

313W x 211H x 77D mm (12W x 8H x 3D in.)

Weight

3.4 kg (<7.5 lbs.) typical

Environmental

Standard

MIL-PRF-28800F class 2

Operating

–10º C to 55º C, humidity 85% or less

Storage

–51º C to 71º C

Altitude

4600 meters, operating and non-operating

Safety

Safety

Conforms to EN 61010-1 for Class 1 portable equipment

Electromagnetic Compatibility

EMC

Meets European Community requirements for CE marking.

> **Important:** Maximum continuous input: +30 dBm (≥10 dB input attenuation). Input damage level: >+30 dBm, ±50 Vdc.

> **Important:** Most over-the-air, RF measurement, and demodulator options for the MS2724B require Option MS2724B-009 (IQ Demodulation Hardware); Options MS2724B-033 and MS2724B-034 additionally require Option MS2724B-031 (GPS).

> **Important:** Option MS2724B-019 High Accuracy Power Meter does not include the PSN50 sensor; the sensor (PSN50 or MA24106A) must be obtained separately.

> **Important:** Discrete spurious signals are not included in the measurement of DANL as they are covered by the residual spurious specification.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 675.0 USD
- **Vendor:** Anritsu
- **Type:** Spec An
- **Tags:** Analyzers, Anritsu, Manufacturers, product_manual_group_1, Product_used, Spectrum

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 525.00 USD | In stock |
| NIST Traceable with Full Data | 600.00 USD | In stock |
| ISO IEC 17025 Accredited | 675.00 USD | In stock |

## Images

- Anritsu MS2724B 20 GHz Handheld Spectrum Analyzer (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/ms2724b-anritsu-spectrum-analyzer-used)
> Updated: 2026-07-06