# Agilent E5100A Network Analyzer (Pre-Owned)

# Agilent E5100A Network Analyzer (Pre-Owned)

## Key Features At A Glance

-   Source and receiver covering 10 kHz to 300 MHz in a single network analyzer
-   GPIB (ANSI/IEEE 488.2), parallel I/O, and EXT TRIGGER with built-in Instrument BASIC for automated systems
-   1 to 4 measurement channels (depends on option) with 2 to 1,601 measurement points per sweep
-   Frequency accuracy ±20 ppm at 23 ±5 °C; optional ovenized reference (Option 1D5) for ±1 ppm
-   Phase noise < -90 dBc/Hz at 10 kHz offset from a 0 dBm fundamental
-   6.5 inch color LCD (640 x 480), built-in flexible disk drive, and VGA monitor output
-   Frequency resolution 1 mHz; output power resolution 0.1 dB
-   Measurement calibration: response, response and isolation, and 1-port 3-term

The Agilent E5100A is a network analyzer covering 10 kHz to 300 MHz, integrating a signal source and a receiver in a single instrument. It supports one to four measurement channels and performs both magnitude and phase measurements, with measurement calibration types including response, response and isolation, and 1-port 3-term.

A network analyzer characterizes how a device behaves across frequency by comparing a stimulus signal against the device's response. The E5100A offers ratio measurement modes such as A/R, R/A, and, with the appropriate input configuration, B/R, B/A, A/B, and additional pairings, alongside absolute A, B, and R measurements. Both frequency and power sweep parameters are available, using linear (step, ramp) or list sweep types over a Cartesian display format.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units leave the warehouse factory-sealed exactly as received. Holding inventory directly means the instrument you order is one we physically have on hand and have prepared to the standard appropriate for its condition.

## Brand Heritage

The test and measurement line carrying the Agilent name originated with Hewlett-Packard, whose test and measurement business was spun out as Agilent Technologies in 1999. That business was subsequently transferred to Keysight Technologies in 2014.

## Accessories Supplied

-   Power cable (Qty 1)
-   Sample program disk (Qty 1, E5100-180X01)
-   CD-ROM (manuals) (Qty 1, E5100-905XX)
-   Function Reference (Qty 1, E5100-900X02)
-   Programming Manual (Qty 1, E5100-900X72)
-   User's Guide (Qty 1, E5100-900X12)
-   Instrument BASIC Users Handbook (Qty 1, 04155-90150)
-   Instrument BASIC Users Handbook Supplement (Qty 1, E5100-900X52)

## Product Core & Specifications

Specification

Value

Source — Frequency Characteristics

Range

10 kHz to 300 MHz

Accuracy (at 23 ±5 °C)

±20 ppm

Accuracy with Option E5100A-1D5 (at 0 to 55 °C, 20 minutes after power on)

±1 ppm

Stability (at 23 ±5 °C)

±5 x 10–6/day (SPC)

Stability with Option E5100A-1D5 (48 hours after power on)

±2.5 x 10–9/8 hours (SPC)

Resolution

1 mHz

Source — Output Power Characteristics (measured at RF OUT 1, RF OUT 2 terminated with 50 Ω)

Range (nominal) with Option E5100A-001

–9 dBm to +11 dBm

Range (nominal) with Option E5100A-002

–15 dBm to +5 dBm

Range (nominal) with Option E5100A-003

–12 dBm to +8 dBm

Range (nominal) with Option E5100A-801

–48 dBm to +22 dBm

Range (nominal) with Option E5100A-802

–54 dBm to +16 dBm

Range (nominal) with Option E5100A-803

–51 dBm to +19 dBm

Range (nominal) with Option E5100A-600 (at RF OUT 1)

–52 dBm to +18 dBm

Range (nominal) with Option E5100A-600 (at RF OUT 2)

–65 dBm to +5 dBm

Resolution

0.1 dB

Level accuracy (at 23 ±5 °C, 0 dBm output level, 50 MHz)

±1 dB

Flatness (at 23 ±5 °C, relative to 0 dBm output level at 50 MHz)

+2 dB, –4 dB

Flatness with Option E5100A-803

+2.5 dB, –4.5 dB

Linearity (at 23 ±5 °C, relative to 0 dBm output level at 50 MHz)

±1 dB

Source — Spectral Purity Characteristics

Non–harmonic spurious signals (at < 300 MHz)

< –45 dBc

Phase noise (at 10 kHz offset from 0 dBm fundamental)

< –90 dBc/Hz

Other Source Information

Reverse power protection

20 dBm, 25 Vdc (SPC)

Output connector

BNC female

Output impedance

50 Ω (nominal)

Receiver — Input Characteristics

Frequency range

10 kHz to 300 MHz

1 MΩ input for Options E5100A-703/704/707/708

10 kHz to 5 MHz

IF bandwidth (IF BW)

10 Hz to 30 kHz, 1, 1.5, 2, 3, 4, 5, 8 step (nominal)

Impedance

50 Ω (nominal)

1 MΩ input for Option E5100A-703/704/707/708

1 MΩ // 30 pF (nominal)

Return loss (at 50 Ω input, 10 kHz ≤ frequency < 100 MHz)

20 dB (SPC)

Return loss (at 50 Ω input, 100 MHz ≤ frequency ≤ 300 MHz)

15 dB (SPC)

Damage level DC

25 Vdc

Damage level AC

20 dBm

Receiver — Averaging Noise Level (magnitude, 23 ±5 °C, RF attenuator: 0 dB, 50 Ω input)

IF BW 30 kHz (at > 1 MHz)

–100 dBm

IF BW 10 kHz (at > 300 kHz)

–105 dBm

IF BW 3 kHz (at > 100 kHz)

–110 dBm

IF BW 1 kHz (30 kHz ≤ frequency < 100 kHz)

–95 dBm

IF BW 1 kHz (100 kHz ≤ frequency ≤ 300 MHz)

–115 dBm

IF BW 300 Hz (10 kHz ≤ frequency < 100 kHz)

–100 dBm

IF BW 300 Hz (100 kHz ≤ frequency ≤ 300 MHz)

–120 dBm

IF BW 100 Hz (10 kHz ≤ frequency < 100 kHz)

–105 dBm

IF BW 100 Hz (100 kHz ≤ frequency ≤ 300 MHz)

–125 dBm

Receiver — Magnitude Characteristics

Absolute amplitude accuracy (at 23 ±5 °C, –30 dBm input for RF attenuator: 0 dB or –5 dBm input for RF attenuator: 25 dB, 50 Ω input)

±1 dB

Frequency response (50 Ω input, 10 kHz ≤ frequency < 100 kHz)

±1 dB

Frequency response (50 Ω input, 100 kHz ≤ frequency ≤ 100 MHz)

±0.5 dB

Frequency response (50 Ω input, 100 MHz < frequency ≤ 300 MHz)

±1 dB

Frequency response (1 MΩ input for Option E5100A-703/704/707/708, using 50 Ω feedthrough)

±3 dB

Trace noise (at 1 kHz IF BW, frequency > 305 kHz)

< 10 dBm rms

Stability

0.02 dB/°C (SPC)

Receiver — Phase Characteristics

Measurement mode

Normal/Expanded

Frequency response (10 kHz ≤ frequency < 100 kHz)

±5°

Frequency response (100 kHz ≤ frequency ≤ 100 MHz)

+2.5°

Frequency response (100 MHz < frequency ≤ 300 MHz)

±5°

Trace noise (at 1 kHz IF BW, frequency > 305 kHz)

< 50 mdeg rms

Stability

0.15 deg/°C (SPC)

General Characteristics

Operating temperature (disk drive in operation)

10 to 40 °C

Operating humidity (disk drive in operation, wet bulb ≤ 29 °C, without condensation)

15% ≤ RH ≤ 80%

Operating temperature (disk drive not in operation)

5 to 40 °C

Operating humidity (disk drive not in operation, wet bulb ≤ 29 °C, without condensation)

15% ≤ RH ≤ 80%

Operating altitude

0 to 2,000 meters

Warm-up time

30 minutes

Non-operating temperature

–20 to 60 °C

Non-operating humidity (wet bulb ≤ 40 °C, without condensation)

15% ≤ RH ≤ 90%

Non-operating altitude

0 to 4,572 meters

Safety

Certified by CSA-C22.2 No.1010.1-92, Based on IEC 1010-1 (1990) including Amendment 1 (1992)

EMC

Complies with CISPR 11(1990)/EN 55011(1991): Group 1, Class A; IEC 801-2 (1991)/EN 55082-1(1992): 4 kV CD, 8 kV AD; IEC 801-3 (1984)/EN 55082-1(1992): 3 V/m; IEC 801-4 (1988)/EN 55082-1(1992): 1 kV power lines, 0.5 kV signal lines

Power requirement

90 to 132 V or 198 to 264 V, 47 to 63 Hz, 400 VA max

Weight (depending on option)

12 kg (SPC)

Cabinet dimensions

425(W) x 177(H) x 425(D) mm (SPC)

Supplemental — Measurement Function

Number of measurement channels

1 to 4

Display format

Cartesian

Sweep parameter

frequency, power

Sweep type

linear (step, ramp), list

Measurement points per sweep

2 to 1,601

Measurement calibration

Response, response and isolation, 1-port 3-term

Display

6.5 inch color LCD, 640 x 480 dots

Flexible disk drive

720 Kbytes/1.2 Mbytes/1.44 Mbytes, DOS format, binary or ASCII format

Flash disk

256 Kbytes

Ram disk

256 Kbytes

Programming

Instrument BASIC

GPIB

ANSI/IEEE 488.2 compatible

Parallel I/O port

16 bit output, 8 bit input/output, TTL level

External video monitor output

VGA

> **Important:** Damage level: 25 Vdc (DC) and 20 dBm (AC). Reverse power protection: 20 dBm, 25 Vdc.

> **Important:** When the analyzer is equipped with Option E5100A-001 or E5100A-003, the power splitter section does not apply (deleted). When equipped with Option E5100A-600, the power-splitter insertion-loss item does not apply.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 600.0 USD
- **Vendor:** Agilent
- **Type:** Net Ana
- **Tags:** Agilent, Agilent Keysight, Analyzers, Manufacturers, Network, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 525.00 USD | In stock |
| NIST Traceable with Full Data | 562.50 USD | In stock |
| ISO IEC 17025 Accredited | 600.00 USD | In stock |

## Images

- Agilent E5100A Network Analyzer (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/e5100a-agilent-network-analyzer-used)
> Updated: 2026-07-06