# Agilent E4433A Digital RF Signal Generator, 250 kHz to 4000 MHz (Pre-Owned)

# Agilent E4433A Digital RF Signal Generator, 250 kHz to 4000 MHz (Pre-Owned)

## Key Features At A Glance

-   Frequency range 250 kHz to 4000 MHz with 0.01 Hz resolution
-   RF output from +13 dBm down to -136 dBm with 0.02 dB level resolution
-   Standard FM, phase, AM, and pulse modulation with an internal sine/square/ramp/triangle/pulse/noise source
-   External I/Q modulation inputs (50 Ω) with 20 MHz typical RF bandwidth
-   Optional I/Q baseband generator (Opt UN3/UN4) for DECT, GSM, NADC, PDC, PHS, and TETRA formats
-   SSB phase noise <-120 dBc/Hz at 500 MHz carrier, 20 kHz offset (typical)
-   Optional high-stability timebase (Opt 1E5): <±0.1 ppm/yr aging rate
-   Reverse power protection to 50 W with 50 V max DC at the 50 Ω RF output

The Agilent E4433A is a digital RF signal generator from the ESG-D series, delivering a calibrated, controllable RF carrier from 250 kHz to 4000 MHz. Also designated the ESG-D4000A, it pairs a synthesized frequency source with both analog and digitally modulated output in a single instrument, setting frequency to 0.01 Hz resolution and adjusting output amplitude in 0.02 dB steps.

A signal generator supplies the known, controllable stimulus that a device under test responds to. The E4433A produces standard analog modulation — frequency, phase, amplitude, and pulse — and, when fitted with the optional I/Q baseband generator (Option UN3 or UN4), generates digitally modulated signals for the DECT, GSM (including DCS1800/PCS1900), NADC, PDC, PHS, and TETRA standards. External I and Q inputs also allow the instrument to be driven by externally supplied baseband signals.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Pre-owned instruments are inspected and functionally verified in-house by our technicians before they ship, while new units leave the warehouse factory-sealed exactly as received from the manufacturer — we never open or test a new unit. Buying directly from our held stock means the instrument you order is the instrument we ship.

## Brand Heritage

Hewlett-Packard's test and measurement operation was spun off as Agilent Technologies in 1999, and in 2014 that electronic measurement business was established as Keysight Technologies. Instruments originally branded under these names share a continuous engineering lineage across the corporate transitions.

## Product Core & Specifications

Specification

Value

Frequency

Frequency Range (Agilent ESG-D4000A)

250 kHz to 4000 MHz

Resolution

0.01 Hz

Switching Speed (Modulation On)

<45 ms, typical

Switching Speed (Modulation Off)

<35 ms, typical

Accuracy

Same as timebase

Frequency Bands

Band 1 (N# 1)

250 kHz to <=249.999 MHz

Band 2 (N# 0.5)

\>249.999 to <=500 MHz

Band 3 (N# 1)

\>500 MHz to <=1 GHz

Band 4 (N# 2)

\>1 to <=2 GHz

Band 5 (N# 4)

\>2 to 4 GHz

Sweep Modes

Operating modes

Step: frequency & power, and arbitrary list

Dwell Time

2 ms to 60 sec

Number of Points

2 to 401

Internal Reference Oscillator

Aging Rate (Standard, typical)

<±2 ppm/yr

Aging Rate (High Stability, Opt 1E5)

<±0.1 ppm/yr or <±0.0005 ppm/day after 45 days

Temperature (Standard)

<±1 ppm

Temperature (High Stability, Opt 1E5)

<±0.05 ppm, typical (0° to 55° C)

Line Voltage (Standard)

<±0.1 ppm (+5%, -10%)

Line Voltage (High Stability, Opt 1E5)

<±0.002 ppm, typical (+5%, -10%)

Timebase Reference Output Frequency

10 MHz

Timebase Reference Output Amplitude

\>0.35 Vrms into 50 ohm load

External Reference Input Frequency

1, 2, 5, 10 MHz ± typ. 10 ppm (Option 1E5: 1 ppm, typical)

External Reference Input Amplitude

\>0.15 Vrms

External Reference Input Impedance

50 ohm

Output

Range (250 kHz to 1000 MHz)

+13 to -136 dBm

Range (>1000 MHz to 3000 MHz)

+10 to -136 dBm

Range (>3000 MHz to 4000 MHz)

+7 to -136 dBm

Resolution

0.02 dB

Level Accuracy (+7 to -127 dBm, 250 kHz to 2 GHz)

±0.5 dB

Level Accuracy (+7 to -127 dBm, 2 GHz to 4 GHz)

±0.9 dB

Level Accuracy (<-127 dBm, 250 kHz to 2 GHz)

±1.5 dB

Level Accuracy (<-127 dBm, 2 GHz to 4 GHz)

±2.5 dB

Attenuator Hold Level Range

\>17 dB

Switching Speed

<25 ms typical

Switching Speed (With Power Search Mode)

<210 ms typical

Reverse Power Protection (250 kHz to 2000 MHz)

50 Watts

Reverse Power Protection (>2000 MHz to 4000 MHz)

25 Watts

Max DC Voltage

50 V

SWR (typical, 250 kHz to 2000 MHz)

<1.4:1

SWR (typical, >2000 to 4000 MHz)

<1.9:1

Output Impedance

50 ohms

Level Accuracy with Digital Modulation (ALC on; relative to CW)

pi/4 DQPSK or QPSK Formats

±0.15 dB (raised cosine or root-raised cosine filter, alpha >=0.35; 10 kHz 25 MHz; power

Constant Amplitude Formats (FSK, GMSK, etc.)

no degradation in power level accuracy

Level Accuracy with ALC Off

±0.5 dB, typical (after power search executed; relative to CW)

Spectral Purity

SSB Phase Noise (typ, 20 kHz offset) at 500 MHz

<-120 dBc/Hz

at 1000 MHz

<-116 dBc/Hz

at 2000 MHz

<-110 dBc/Hz

at 3000 MHz

<-104 dBc/Hz

at 4000 MHz

<-104 dBc/Hz

Residual FM (CW mode, 0.3-3 kHz BW, CCITT, rms) Phase Noise Mode 1

Residual FM Phase Noise Mode 2

Harmonics (<=+4 dBm output level)

<-30 dBc

Nonharmonics (250 kHz to 1000 MHz)

<-65 dBc

Nonharmonics (>1000 MHz to 2000 MHz)

<-59 dBc

Nonharmonics (>2000 MHz)

<-53 dBc

Subharmonics (<=1000 MHz)

None

Subharmonics (>1000 MHz)

<-40 dBc

IQ Modulation

I&Q Input Impedance

50 ohms

Full Scale Input

√(I²+Q²) = 0.5 Vrms

External Input RF Bandwidth (1 dB)

20 MHz, typical

DC Offset (I + Q independently adjustable)

±100%

I/Q Gain Ratio

±4 dB

DC Vector Accuracy - Static EVM (rms)

<0.75% (<0.6 GHz), <0.5% (0.6-2 GHz), 0.75% (2-3.7 GHz), <1% (<=4 GHz)

DC Vector Accuracy - Magnitude Error (rms)

<0.5% (<0.6 GHz), <0.35% (0.6-2 GHz), <0.5% (2-3.7 GHz), <0.75% (<=4 GHz)

DC Vector Accuracy - Phase Error (rms)

<0.35° (<0.6 GHz), <0.25° (0.6-2 GHz), <0.35° (2-3.7 GHz), <0.5° (<=4 GHz)

DC Vector Accuracy - Origin Offset (dBc)

<-46 (<0.6 GHz), <-46 (0.6-2 GHz), <-40 (2-3.7 GHz), <-40 (<=4 GHz)

Frequency Modulation

Maximum Deviation

N x 10 MHz

Resolution

0.1% of deviation or 1 Hz, whichever is greater

Deviation Accuracy (1 kHz rate, dev.

<±(3.5% of FM deviation + 20 Hz)

FM1 1 dB Bandwidth (deviation = 100 kHz)

dc/20 Hz to 100 kHz

FM1 3 dB Bandwidth, typical

dc/5 Hz to 10 MHz

FM2 1 dB Bandwidth

dc/20 Hz to 100 kHz

FM2 3 dB Bandwidth, typical

dc/5 Hz to 1 MHz

Distortion (1 kHz rate, THD, dev. = N x 100 kHz)

<1%

Phase Modulation

Maximum Deviation

N x 90 radians

Resolution

0.1% of set deviation

Deviation Accuracy (1 kHz rate)

<±(5% of deviation + 0.01 radians)

Normal BW (N x 90 rad)

PM1 dc to 100 kHz; PM2 dc to 100 kHz

High BW (N x 2pi rad)

PM1 dc to 1.5 MHz (typ); PM2 dc to 1 MHz (typ)

High BW (N x pi/2 rad)

PM1 dc to 4 MHz (typ); PM2 dc to 0.9 MHz (typ)

Distortion (1 kHz rate, THD, dev

<1%

Amplitude Modulation (fc>500 kHz)

Range (envelope peak <=max specified power)

0 to 100%

Resolution

0.1%

Rates (3 dB Bandwidth)

dc/10 Hz to 10 kHz

Distortion (1 kHz rate, THD) 30% AM

<1.5%

Distortion (1 kHz rate, THD) 90% AM

<4%

Accuracy (1 kHz rate)

<±(5% of setting + 1%)

Wide Band AM Rate (1 dB Bandwidth, typ) ALC On

400 Hz to 10 MHz

> **Important:** Maximum DC voltage at the RF output is 50 V; reverse power protection is rated to 50 W from 250 kHz to 2000 MHz and 25 W from above 2000 to 4000 MHz.

> **Important:** Level accuracy degrades by 0.02 dB/°C over the full temperature range and by 0.3 dB above +7 dBm.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 560.0 USD
- **Vendor:** Agilent
- **Type:** Rf Gen
- **Tags:** Agilent, Agilent Keysight, Manufacturers, product_manual_group_1, Product_used, RF, Signal Generators

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 455.00 USD | In stock |
| NIST Traceable with Full Data | 490.00 USD | In stock |
| ISO IEC 17025 Accredited | 560.00 USD | In stock |

## Images

- Agilent E4433A Digital RF Signal Generator, 250 kHz to 4000 MHz (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/e4433a-agilent-rf-generator-used)
> Updated: 2026-07-06
