# Agilent E4400A Analog RF Signal Generator, 250 kHz to 1000 MHz (Pre-Owned)

# Agilent E4400A Analog RF Signal Generator, 250 kHz to 1000 MHz (Pre-Owned)

## Key Features At A Glance

-   Frequency range 250 kHz to 1000 MHz with 0.01 Hz resolution
-   Output level range +13 to -136 dBm at 0.02 dB resolution
-   AM, FM, phase, and pulse modulation with an internal modulation source
-   SSB phase noise <-120 dBc/Hz at 500 MHz (typical, 20 kHz offset)
-   HP-IB (IEEE-488.2) and RS-232 remote control with SCPI 1992.0
-   Optional high-stability timebase (Opt 1E5): aging <±0.1 ppm/yr
-   Reverse power protection to 50 W max and 50 V dc max
-   Conducted/radiated interference meets MIL-STD-461B RE02 Part 2 and CISPR 11

The Agilent E4400A is an analog RF signal generator covering 250 kHz to 1000 MHz, also identified as the Agilent ESG-1000A. It produces calibrated RF output across that span with 0.01 Hz frequency resolution, and pairs that frequency agility with a broad amplitude range and a full suite of analog modulation modes driven from an internal source.

An analog RF signal generator supplies a known, controllable RF stimulus so that an engineer can drive a device under test and measure its response. It serves as the source half of a stimulus-response test setup, generating carrier signals at defined frequencies, levels, and modulation states for bench and system measurements.

ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house against the original manufacturer specifications before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

## Brand Heritage

This instrument carries Agilent branding from the era when Hewlett-Packard spun off its test and measurement business as Agilent Technologies in 1999. Agilent's electronic measurement division later became Keysight Technologies in 2014.

## Product Core & Specifications

Specification

Value

Frequency

Frequency Range (Agilent ESG-1000A)

250 kHz to 1000 MHz

Resolution

0.01 Hz

Switching Speed

Modulation On: <45 ms, typical; Modulation Off: <35 ms, typical

Accuracy

Same as timebase

Sweep Modes

Operating modes

Step: frequency & power, and arbitrary list

Dwell Time

2 ms to 60 sec

Number of Points

2 to 401

Internal Reference Oscillator

Aging Rate (Standard, typical)

<±2 ppm/yr

Aging Rate (High Stability, Opt 1E5)

<±0.1 ppm/yr or <±0.0005 ppm/day after 45 days

Temperature (Standard)

<±1 ppm

Temperature (High Stability, Opt 1E5)

<±0.05 ppm, typical (0° to 55°C)

Line Voltage (Standard)

<±0.1 ppm (+5%, -10%)

Line Voltage (High Stability, Opt 1E5)

<±0.002 ppm, typical (+5%, -10%)

Timebase Reference Output

Frequency

10 MHz

Amplitude

\>0.35 Vrms into 50 ohm load

External Reference Input

Frequency

1, 2, 5, 10 MHz ± typ. 10 ppm; Option 1E5: 1 ppm, typical

Amplitude

\>0.15 Vrms

Input Impedance

50 ohm

Output

Range

250 kHz to 1000 MHz: +13 to -136 dBm

Resolution

0.02 dB

Level Accuracy (at 23 ±5°C), +7 to -127 dBm

250 kHz to 1 GHz: ±0.5 dB

Level Accuracy (at 23 ±5°C), <-127 dBm

250 kHz to 1 GHz: ±1.5 dB

Attenuator Hold Level Range

\>17 dB

Switching Speed

<25 ms typical; With Power Search Mode: <210 ms typical

Reverse Power Protection

250 kHz to 1000 MHz: 50 Watts; Max dc Voltage: 50 V

SWR (typical)

250 kHz to 1000 MHz: <1.4:1

Output Impedance

50 ohms

Level Accuracy footnote

Accuracy degrades by 0.02 dB/°C over full temperature range and by 0.3 dB above +7 dBm.

Frequency Bands

Band 1

250 kHz to <=249.999 MHz; N# 1

Band 2

\>249.999 to <=500 MHz; N# 0.5

Band 3

\>500 MHz to <=1 GHz; N# 1

Spectral Purity

SSB Phase Noise (typical, at 20 kHz offset)

at 500 MHz: <-120 dBc/Hz; at 1000 MHz: <-116 dBc/Hz

Residual FM (CW mode, 0.3-3 kHz BW, CCITT, rms)

Phase Noise Mode 1:

Harmonics (<=+4 dBm output level)

<-30 dBc

Nonharmonics (>3 kHz offset, <+7 dBm output level)

250 kHz to 1000 MHz: <-65 dBc

Subharmonics (<=1000 MHz)

None

Frequency Modulation

Maximum Deviation

N x 10 MHz

Resolution

0.1% of deviation or 1 Hz, whichever is greater

Deviation Accuracy (1 kHz rate, dev.

<±(3.5% of FM deviation + 20 Hz)

Modulation Frequency Response (deviation = 100 kHz), FM1

1 dB BW: dc/20 Hz to 100 kHz; 3 dB BW, typical: dc/5 Hz to 10 MHz

Modulation Frequency Response (deviation = 100 kHz), FM2

1 dB BW: dc/20 Hz to 100 kHz; 3 dB BW, typical: dc/5 Hz to 1 MHz

Distortion (1 kHz rate, THD, dev. = N x 100 kHz)

<1%

Phase Modulation

Maximum Deviation

N x 90 radians

Resolution

0.1% of set deviation

Deviation Accuracy (1 kHz rate)

<±(5% of deviation + 0.01 radians)

Modulation Frequency Response, Normal BW (N x 90 rad)

PM1: dc to 100 kHz; PM2: dc to 100 kHz

Modulation Frequency Response, High BW (N x 2 pi rad)

PM1: dc to 1.5 MHz (typ); PM2: dc to 1 MHz (typ)

Modulation Frequency Response, High BW (N x pi/2 rad)

PM1: dc to 4 MHz (typ); PM2: dc to 0.9 MHz (typ)

Distortion (1 kHz rate, THD, dev

<1%

Amplitude Modulation (fc>500 kHz)

Range (envelope peak <=max specified power)

0 to 100%

Resolution

0.1%

Rates (3 dB Bandwidth)

dc/10 Hz to 10 kHz

Distortion (1 kHz rate, THD), 30% AM

<1.5%

Distortion (1 kHz rate, THD), 90% AM

<4%

Accuracy (1 kHz rate)

<±(5% of setting + 1%)

Pulse Modulation

On/Off Ratio (<=1 GHz)

\>80 dB

Rise/Fall Times

150 ns, typical

Minimum Width (typical)

ALC On: 2 µs; ALC Off: 0.4 µs

Pulse Repetition Frequency (typical)

ALC On: 10 Hz to 250 kHz; ALC Off: DC to 1.0 MHz

Level Accuracy (relative to CW)

± 0.5 dB, typical

Internal Pulse Generator (Squarewave only)

Squarewave Rates

0.1 Hz to 50 kHz

Pulse Period

16 µs to 30 seconds

Width

8 µs to 30 seconds

Resolution

4 µs

Internal Modulation Source

Provides

FM, PM, and AM Modulation Signals and LF Out

Waveforms

sine, square, ramp, triangle, pulse, noise

Rate Range

Sine: 0.1 Hz to 50 kHz; Square, Ramp, Triangle: 0.1 Hz to 10 kHz; Pulse: 4 µs

Resolution

0.1 Hz

Frequency Accuracy

0.005%

Swept Sine Mode (Frequency, Phase Continuous), Operating Modes

Triggered or Continuous Sweeps

Swept Sine Mode, Frequency Range

0.1 Hz to 50 kHz

Swept Sine Mode, Sweep Time

1 ms to 65 seconds

Swept Sine Mode, Resolution

1 ms

Dual Sinewave Mode, Frequency Range

0.1 Hz to 50 kHz

Dual Sinewave Mode, Amplitude Ratio

0 to 100%

Dual Sinewave Mode, Amplitude Resolution

0.1%

LF Out (Internal Modulation Source)

Amplitude

0 to 3 Vpeak into 50 ohms

Output Impedance

<1 ohm

External Modulation Inputs

Modulation Types

Ext1: FM, PM, AM, and Burst Envelope; Ext2: FM, PM, AM, and Pulse

Simultaneous Modulation

All modulation types may be simultaneously enabled, except: FM with PM, AM with Burst. AM, FM, and PM can sum simultaneous inputs from any two sources (INT, EXT1, and EXT2.) Any given source (INT, EXT1, or EXT2) may only be routed to one activated modulation type.

Remote Programming

Interface

HP-IB (IEEE-488.2-1987) with Listen and Talk, RS-232

Control Languages

SCPI version 1992.0, also compatible with Agilent 8656B & 8657A/B/D/J mnemonics

Functions Controlled

All front panel functions except power switch and knobs

IEEE-488 Functions

SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C0, E2

General

Power Requirements

90 to 254 V; 50, 60, or 400 Hz; 200 W maximum

Operating Temperature Range

0 to 55° C

Leakage

Conducted and radiated interference meets MIL-STD-461B RE02 Part 2 and CISPR 11

> **Important:** Reverse power protection: 50 W maximum, 50 V dc maximum.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 1050.0 USD
- **Vendor:** Agilent
- **Type:** Rf Gen
- **Tags:** Agilent, Agilent Keysight, Manufacturers, product_manual_group_1, Product_used, RF, Signal Generators

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 700.00 USD | In stock |
| NIST Traceable with Full Data | 840.00 USD | In stock |
| ISO IEC 17025 Accredited | 1050.00 USD | In stock |

## Images

- E4400A Agilent RF Generator

---

> Source: [ValueTronics](https://valuetronics.com/products/e4400a-agilent-rf-generator-used)