# B&K Precision DAS60 6-Channel High Speed Multi-Function Recorder

# B&K Precision DAS60 6-Channel High Speed Multi-Function Recorder

## Key Features At A Glance

-   2, 4, or 6 isolated universal analog channels for voltage, current, temperature, and frequency on a common time base
-   CAT III 600 V rated isolated input channels per IEC 61010
-   Wide input range from ±5 mV to ±500 VDC / 424 VRMS at 14-bit resolution
-   Fast 1 MSa/s sample rate (memory mode) and 100 kHz bandwidth for capturing intermittent events
-   16 logic input channels plus thermocouple and Pt100/Pt1000 temperature measurement
-   10-inch TFT touchscreen display with battery life up to 9.5 hours
-   2 USB host ports and a LAN interface with license-free Sefram software and built-in VNC
-   32 GB (DAS30/DAS50) or 64 GB (DAS60) built-in solid-state memory

The B&K Precision DAS30, DAS50, and DAS60 are high speed multi-function recorders built to measure and record signals ranging from small sensors to large electrical systems on a single instrument. Each model combines isolated universal analog channels, logic inputs, and temperature measurement with high-speed sampling and large internal solid-state memory, presenting captured data on a wide 10-inch touchscreen TFT display.

These recorders are applied to measuring signals ranging from small sensors to large electrical systems, to maintenance and failure analysis, and to power analysis of single- and three-phase systems. Each channel can simultaneously record a different signal — voltage, temperature, current, or frequency — using a common time base, allowing high voltage and current waveforms, temperature, and logic data to be captured together in one instrument.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, so the instruments we list are equipment we keep on hand.

## Compare Other Models in This Series

The DAS30, DAS50, and DAS60 form a single series of high speed multi-function recorders that share the same core measurement architecture — isolated universal analog channels, 14-bit resolution, 100 kHz bandwidth, a 1 MSa/s memory-mode sample rate, CAT III 600 V isolation, and a 10-inch touchscreen — while differing in channel count and several capability tiers. The DAS30 provides 2 isolated universal channels, the DAS50 provides 4, and the DAS60 provides 6.

The models also differ in file-mode sampling interval, internal memory size, power-analysis networks, alarm count, and which temperature and analysis options are factory-installed versus standard. The comparison table below details these differences model by model.

Each new model linked below is its own dedicated product page with condition-matched pricing. Select the model whose channel count and capability tier fit your application, then choose the available new configuration on that model's page.

Channel count scales across the series — 2 channels on the DAS30, 4 on the DAS50, and 6 on the DAS60 — and the file-mode sampling interval is 5 µs (200 kSa/s) on the DAS30 and 2 µs (500 kSa/s) on the DAS50 and DAS60. Internal solid-state memory is 32 GB on the DAS30 and DAS50 and 64 GB on the DAS60.

Power-analysis capability also differs by model: the DAS30 covers single-phase, the DAS50 covers single-phase and Delta (Aron), and the DAS60 covers single-phase, Delta (Aron), and Star, adding 400 Hz power-analysis frequency and PWM analysis. The DAS60 includes two Pt100/Pt1000 inputs as standard and provides four alarms, whereas the DAS30 and DAS50 offer Pt100/Pt1000 as a factory option and two alarms. Refer to the comparison table for the full per-model breakdown.

Model

Analog Channels

Internal Memory

File Mode Sample Rate

**DAS60**

6

64 GB

2 µs (500 kSa/s)

DAS30

2

32 GB

5 µs (200 kSa/s)

DAS50

4

32 GB

2 µs (500 kSa/s)

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   AC mains adapter 100 / 240 V
-   Rugged carrying case
-   CAT III sheathed banana test leads (one set per channel)
-   Alligator clips (one set per channel)
-   Bare wire to banana adapters (one set per channel)
-   25 pin male connector and backshell
-   Soft wipe
-   Stylus
-   Screwdriver
-   Roll of thermal printer paper (-P models)
-   Calibration certificate & test report

## Product Core & Specifications

Specification

Value

Universal Inputs

Number of Channels

DAS30: 2; DAS50: 4; DAS60: 6

Voltage

Maximum Input Voltage

±500 VDC or 424 VRMS

Maximum Offset

± 5 ranges (up to ± 500 V)

Accuracy

0.1% of the full scale ±10 µV

True RMS AC/DC Ranges

200 mV to 424 V

Response Time

100 ms typical (40 ms to 50 Hz)

Crest Factor

2.2 and 600 V Max

Input Impedance (DC)

\> 25 MΩ for upper ranges < 1 V, 1 MΩ for upper ranges, 150 pF typical

Channel Isolation

\> 100 MΩ at 500 VDC

Bandwidth and Filters

Bandwidth (-3 dB), > 1 V

100 kHz

Bandwidth (-3 dB), > 50 mV

50 kHz

Bandwidth (-3 dB), 5 mV

20 kHz

True RMS AC/DC Bandwidth

5 Hz to 500 Hz

Internal Analog Filters

10 kHz, 1 kHz, 100 Hz, 10 Hz

Slope

20 dB/decade

Programmable Digital Filters

10 Hz, 1 Hz, 0.1 Hz, 0.01 Hz, 0.001 Hz

Frequency Counter

Sensitivity

100 mV RMS min.

Duty Cycle

10% min.

Frequency Range

0.1 Hz to 100 kHz

Basic Accuracy

0.02% of full scale

Data Acquisition and Trigger

Resolution

14 bit

Fastest Sampling Interval — File mode (single channel)

DAS30: 5 µs (200 kSa/s); DAS50, DAS60: 2 µs (500 kSa/s)

Fastest Sampling Interval — Memory mode (single channel)

1 µs (1 MSa/s)

Memory Length (memory mode)

32 M word segments up to 128 blocks

Triggering

Positive edge, negative edge, on logic input, delay, go no go

Pre-trigger

±100%

Temperature with Thermocouples (cold junction compensation: ±0.5 °C)

Type J

410 °F to 2192 °F (210 °C to 1200 °C)

Type K

482 °F to 2498 °F (250 °C to 1370 °C)

Type T

392 °F to 752 °F (200 °C to 400 °C)

Type S

122 °F to 3200 °F (50 °C to 1760 °C)

Type B

392 °F to 3308 °F (200 °C to 1820 °C)

Type E

482 °F to 1832 °F (250 °C to 1000 °C)

Type N

482 °F to 2372 °F (250 °C to 1300 °C)

Type C

32 °F to 4208 °F (0 °C to 2320 °C)

Type L

392 °F to 1652 °F (200 °C to 900 °C)

Power Analysis Function

Networks

Single phase, 3 phase

Display

Fresnel diagram, oscilloscope, data

Measurements

Mean value, RMS, peak, crest factor, THD and DF for voltage & current, active, reactive and apparent power, power factor (ø)

Harmonics

Calculated up to rank 50, with display and record

Logic Input

Channels

16

TTL Maximum Voltage

24 V

Sampling Interval

1 µs (1 MSa/s) per channel

Sensor Supply

9 to 15 VDC

Alarms

Alarms

2, A & B 0 to 5 V output

Pt100/Pt1000 (factory option for DAS30 & DAS50)

Number of Channels

2

Current

1 mA for Pt100, 100 µA for Pt1000

Resolution

20 bits

Temperature Range

\-392 °F to 1562 °F (-200 °C to +850 °C)

Measurements

2, 3, 4 wires

Accuracy @ 20 ºC

±0.2 °C

Printer (factory option for all models)

Paper Width

110 mm

Paper Speed

1 mm/min. to 25 mm/s

Paper Speed (memory mode)

10 mm/s max.

Resolution, Y axis

8 dots/mm

Resolution, X axis

16 dots/mm

XY mode

8 dots/mm (both axis)

General

Internal Solid State Memory (file mode)

32 GB (DAS30, DAS50); 64 GB (DAS60)

Operating Temperature

0 °C to 40 °C, 80% RH (no condensation)

Storage Temperature

\-68 °F to 140 °F (-20 °C to 60 °C)

Display

10” TFT touchscreen LCD, backlit, 1024 x 600 dots

Power Supply

15 V / 4 A max with main adapter (100 / 240 VAC)

Interfaces

2 x USB host, LAN (10/100 base-T with RJ45 socket)

Battery

Non removable, Lithium-ion

Typical Battery Life

9.5 hours with standby mode, 4 hours without standby mode

Safety

IEC 61010 - CAT III 600 V

Weight

5.5 lbs (2.5 kg)

Dimensions (W x H x D)

8.25” x 11.5” x 4.1” (210 x 295 x 105 mm)

Warranty

Two Years

> **Important:** Current clamps are not included with the recorder.

> **Important:** The 110 mm thermal printer is a factory-installed option on all models and must be specified at time of order (-P configurations).

> **Important:** The 2 Pt100/Pt1000 temperature inputs are a factory-installed option on the DAS30 and DAS50 (-T configurations) and are included as standard on the DAS60.

> Recommended pairing: the 16-channel isolated logic adapter (part 984405500) listed in the optional accessories, for capturing logic-channel data alongside the analog inputs.

## About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 7200.0 USD
- **Vendor:** B&K Precision
- **Type:** Data Log
- **Tags:** BK Precision, DAQ, DAS Series, DAS30-50-60 Series, Data Acquisition, Data Logger, Manufacturers, Meters & DAQ, Multi-Channel Recorder, Other Equipment, Power Analysis, product_manual_group_2, Product_new, Recorder, Sefram

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 7200.00 USD | In stock |

## Images

- B&K Precision DAS60 6-Channel High Speed Multi-Function Recorder

---

> Source: [ValueTronics](https://valuetronics.com/products/das60-b-k-precision-recorder-new)
> Updated: 2026-06-27