# SSA5083A Siglent Spectrum Analyzer New

### Key Features at a Glance

-   **Frequency:** 9 kHz to 13.6 GHz
-   **DANL:** -165 dBm/Hz typical (preamp on)
-   **SSB Phase Noise:** -105 dBc/Hz typ. at 1 GHz, 10 kHz offset
-   **Analysis BW:** 25 MHz std., 40 MHz opt. (SSA5000-B40)
-   **Real-Time:** 100% POI 7.20 µs (SSA5000A-RTA1)
-   **Measurements:** CHP, ACPR, OBW, Harmonic, TOI (SSA5000-AMK)
-   **Modulation:** Analog AM/FM/PM, digital ASK/FSK/MSK/PSK/QAM
-   **Display:** 12.1 inch multi-touch, HDMI, web-browser remote

## About the Siglent SSA5083A

In summary, the Siglent SSA5083A is a benchtop spectrum analyzer covering 9 kHz to 13.6 GHz with a -165 dBm/Hz typical displayed average noise level, -105 dBc/Hz typical SSB phase noise at 1 GHz / 10 kHz offset, and 25 MHz of standard analysis bandwidth (40 MHz optional). It supports real-time spectrum analysis with a 7.20 µs 100% probability-of-intercept minimum signal duration, analog modulation analysis (AM, FM, PM), vector digital modulation analysis (ASK, FSK, MSK, PSK, QAM), and CISPR 16-1-1 EMI pre-compliance, on a 12.1 inch multi-touch screen with HDMI output and web-browser remote control.

### What is the Siglent SSA5083A used for?

The SSA5083A is used for RF spectrum and signal analysis including channel power, ACPR, occupied bandwidth, harmonic, and third-order intercept measurements, plus real-time captures of hopping signals, transient interference, and intermittent emissions that a strictly swept analyzer cannot reliably catch. Per the datasheet, applications include broadcast monitoring, cellular site work, IoT device characterization, WLAN and Bluetooth surveying, education, production test, and field maintenance.

### How does the SSA5083A compare to a swept-only analyzer?

A swept analyzer measures one narrow segment at a time and steps through the span sequentially, so short events between sweeps are missed. The SSA5083A with the SSA5000A-RTA1 option performs up to 150,000 FFTs per second over a 40 MHz analysis bandwidth and guarantees 100% probability of intercept for signals as short as 7.20 µs — suitable for frequency-hopping radios, radar pulses, and intermittent interference. Per the datasheet, the SSA5083A delivers -165 dBm/Hz typical DANL with preamp on, +15 dBm typical TOI from 50 MHz to 7.22 GHz, residual response under -90 dBm, RBW from 1 Hz to 10 MHz in 1-3-10 sequence, and CISPR 16-1-1 EMI filters at 200 Hz, 9 kHz, 120 kHz, and 1 MHz. The 10 MHz reference can be locked externally, and SCPI / IVI / LabVIEW remote control is supported over USB-TMC, VXI-11, socket, and Telnet, with an optional USB-GPIB adapter.

### Where does the SSA5083A fit in an RF lab?

-   Transmitter verification, spurious checks, harmonic and TOI characterization on RF designs up to 13.6 GHz.
-   Real-time monitoring of hopping radios, radar pulses, and intermittent interference.
-   EMI pre-compliance with CISPR 16-1-1 quasi-peak detection and pre-loaded EN 55022, GB 9254, and FCC Part 15 limit standards (SSA5000-EMI).
-   Modulation quality analysis (EVM, magnitude error, phase error, IQ offset, SNR) for digital wireless designs.
-   Education and research benches that benefit from a 12.1 inch multi-touch screen and web-browser remote control.

## Accessories Supplied

Per the manufacturer datasheet, the standard ship kit includes: printed Quick Start documentation; USB Type-B device cable; region-appropriate AC power cord; wireless USB mouse; and a 2.92 mm female-to-female through adapter (2.92F-2.92F-40A, DC to 40 GHz).

## Core Specifications

### Frequency, Bandwidths, Sweep, Trigger

Frequency range / resolution

9 kHz to 13.6 GHz / 1 Hz

Internal reference

10.000000 MHz, < 1 ppm initial, < 0.5 ppm/yr aging

OCXO option 10M\_OCXO\_L

< 0.1 ppm initial, < 50 ppb/yr aging

RBW / VBW (-3 dB)

1 Hz to 10 MHz, 1-3-10 sequence

EMI filter (-6 dB)

200 Hz, 9 kHz, 120 kHz, 1 MHz (CISPR 16)

Analysis bandwidth

25 MHz std., 40 MHz opt. (SSA5000-B40)

Sweep time, Span = 0 Hz / ≥ 100 Hz

1 µs to 6000 s / 1 ms to 4000 s

Sweep mode

Swept (RBW 3 kHz-10 MHz); FFT (1 Hz-10 kHz)

Trigger sources / external

Free, Video, External, Period / 5 V TTL

Gate methods

Gated LO, Gated Video, Gated FFT

### Amplitude, Level, Phase Noise

Measurement range (preamp off, 1 MHz to 13.6 GHz)

DANL to +23 dBm

Reference level

\-170 dBm to +23 dBm, 1 dB steps

Preamplifier

20 dB nominal, 100 kHz to 13.6 GHz

Input attenuation

0 to 50 dB, 2 dB steps

Maximum input DC voltage

± 50 VDC

Maximum average power

27 dBm, 3 min, fc ≥ 10 MHz, att > 20 dB

Traces / detectors

6 / Pos-peak, Neg-peak, Sample, Normal, Average (V/RMS/Video)

SSB phase noise (fc = 1 GHz), 10 kHz offset

\-103 dBc/Hz, -105 typ.

SSB phase noise, 100 kHz / 1 MHz offset

\-103, -105 typ. / -116, -119 typ. dBc/Hz

### Displayed Average Noise Level (DANL, dBm)

Conditions: 20-30 °C, att = 0 dB, RBW = 1 Hz, sample detector, trace average > 50.

Range

Preamp Off

Preamp On

100 kHz to 1 MHz

\-130, -143

\-135, -148

1 MHz to 10 MHz

\-143, -148

\-153, -165

10 MHz to 1.22 GHz

\-144, -148

\-159, -163

1.22 to 3.15 GHz

\-140, -144

\-158, -162

3.15 to 7.22 GHz

\-137, -141

\-154, -158

7.22 to 13.6 GHz

\-136, -140

\-154, -158

Values shown as guaranteed, typical.

### Accuracy, Frequency Response, Distortion

Frequency response, 10 MHz to 7.5 GHz

± 1.5 dB

Frequency response, 7.5 to 13.6 GHz

± 2.0 dB

Absolute amplitude accuracy at 50 MHz

± 0.4 dB

RBW switching uncertainty

± 0.2 dB nominal

RF input VSWR, 10 MHz to 13.6 GHz

1.6 nominal

Second harmonic distortion (SHI), 10 MHz to 7.22 GHz

42 dBm nominal

Third-order intercept (TOI), 50 MHz to 7.22 GHz

11 dBm, 15 dBm nominal

Third-order intercept (TOI), 7.22 GHz to 13.6 GHz

10 dBm, 14 dBm nominal

Residual response / spurious

< -90 dBm / < -65 dBc (mixer -30 dBm)

1 dB gain compression

\> 5 dBm nominal

### Real-Time, EMI, and Modulation Analysis (options)

Real-time bandwidth (SSA5000A-RTA1)

25 MHz default, 40 MHz with SSA5000A-B40

100% POI minimum signal duration

7.20 µs, Kaiser window, FMT

FFT per second (40 MHz BW) / max stored frames

150,000 / 50,000 (loop)

Spectrogram dynamic range

200 dB

FMT types

Greater Than, Less Than, Outside Mask, Inside Mask

Advanced Measurement Kit (SSA5000-AMK)

CHP, ACPR, OBW, T-Power, CNR, Harmonic (up to 10), TOI, Spectrogram

EMI (SSA5000-EMI) detectors / limit standards

Peak, Voltage Average, Quasi-Peak / EN 550xx, GB 9254, FCC Part 15, user

Analog Mod. Analysis (SSA5000-AMA)

AM, FM, PM, ± 4% accuracy; carrier 2 MHz to 13.6 GHz

AM / FM / PM range

20 Hz-100 kHz, 5-95% / 20 Hz-200 kHz, 1-400 kHz / 50 Hz-50 kHz, 0.2-100 rad

Digital Mod. Analysis (SSA5000-DMA)

ASK, 2/4/8/16 FSK, GMSK MSK, BPSK/QPSK/OQPSK/8PSK, DBPSK/DQPSK/D8PSK/π-4-DQPSK/π-8-D8PSK, 16/32/64/128/256 QAM

Symbol rate

1 ksps to 32 Msps; Rate × Pts/Sym ≤ 150 Msps

Error metrics

EVM (rms, peak), magnitude error, phase error, IQ offset, carrier offset, SNR, quadrature error, gain imbalance

### Inputs, Outputs, Remote Control

RF input

2.92 mm male, 50 Ω nominal (front)

USB / LAN

4 USB-A + 1 USB-B 2.0 / 10/100 Base, RJ-45

Video / Audio

HDMI / 3.5 mm jack

10 MHz reference I/O

Out: > 0 dBm; In: -5 to +10 dBm; BNC-F, 50 Ω

External trigger I/O

Out: 3.3 V TTL; In: 5 V TTL, 1 kΩ

Remote protocols

SCPI / LabVIEW / IVI over USB-TMC, VXI-11, Socket, Telnet; NI-MAX; HTML5 web browser; GPIB via USB-GPIB adapter

### Display, Storage, General

Display

TFT LCD, 1200 × 800, 12.1 inch multi-touch

Storage

eMMC 4 GB internal, 32 GB USB external

Dimensions (W×H×D) / weight

378 × 284 × 126 mm / 7.40 kg (16.3 lb)

AC source

100-240 V, 50/60 Hz or 100-120 V, 400 Hz; 80 W max

Operating temperature / humidity

0 to 50 °C; ≤ 95% RH (0-30 °C), ≤ 75% RH (30-50 °C)

Altitude / safety

< 3000 m / IEC, EN, UL, CSA 61010-1 and 61010-2-30

EMC / RoHS

EN 61326-1:2013, EN 61000-3-2/3 / 2011/65/EU

#### Note on Probes and Options

The SRF5030T near-field probe kit (3 H-probes at 20/10/5 mm, 1 E-probe at 5 mm, 300 kHz to 3 GHz) is an optional accessory for EMI source-location work and is ordered separately. The SSA5000-AMK / SSA5000A-RTA1 / SSA5000-B40 / SSA5000-EMI / SSA5000-AMA / SSA5000-DMA / 10M\_OCXO\_L are software / factory-installed options. Confirm required options at the time of order.

This webpage was written by a human with an A.I. "Intern", which may contain errors. Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

## Details

- **Price:** 21712.0 USD
- **Vendor:** Siglent
- **Type:** Spec An
- **Tags:** Analyzers, Manufacturers, New, Siglent, Spectrum

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 21712.00 USD | In stock |

## Images

- SSA5083A Siglent Spectrum Analyzer New

---

> Source: [ValueTronics](https://valuetronics.com/products/ssa5083a-siglent-spectrum-analyzer-new)
> Updated: 2026-05-21