# Keysight E9327A E-Series E9320 Peak and Average Power Sensor, 50 MHz to 18 GHz, 5 MHz Video Bandwidth

# Keysight E9327A E-Series E9320 Peak and Average Power Sensor, 50 MHz to 18 GHz, 5 MHz Video Bandwidth

## Key Features At A Glance

-   Peak, average, and time-gated power measurement using the E-series E9320 sensor family in Normal and Average-only modes
-   Frequency coverage from 9 kHz to 110 GHz and power range of –70 to +44 dBm, sensor dependent
-   GPIB (IEEE 488.2), RS-232, and RS-422 serial interfaces supplied as standard for ATE integration
-   SCPI command language with 20 Msample/s continuous sampling for automated measurement systems
-   Available in single-channel (E4416A) and dual-channel (E4417A) configurations
-   Predefined setups for GSM900, EDGE, NADC, iDEN, Bluetooth, IS-95 CDMA, W-CDMA, and cdma2000 standards
-   Internal 1 mW (0.0 dBm) 50 MHz power reference, factory set to ±0.5% traceable to the NPL
-   Operates with E9320, E-series CW, E9300, 8480, and N8480 series power sensors

The Keysight E4416A and E4417A are EPM-P series power meters that operate with the E-series E9320 family of peak and average power sensors for peak, average, and time-gated power measurements. The E4416A is a single-channel instrument, while the E4417A provides two measurement channels. In addition to the E9320 sensors, the EPM-P meters also operate with the 8480 and N8480 series, E-series CW, and the E9300 range of power sensors for average power measurements. Two measurement modes are provided: Normal mode, the default when using E9320 sensors, for peak, average, and time-related measurements; and Average only mode, intended primarily for average power measurements on low-level signals and used with the 8480, N8480, E-series CW, and E9300 sensors.

A power meter quantifies the RF or microwave power present at its sensor input and displays the result in absolute units (watts or dBm) or relative units (percent or dB). The EPM-P series adds predefined setups for common wireless standards — GSM900, EDGE, NADC, iDEN, Bluetooth, IS-95 CDMA, W-CDMA, and cdma2000 — simplifying configuration for those measurement tasks. With a 20 Msample/second continuous sampling rate and a 5 MHz video bandwidth, the system follows the power envelope of an input signal; because that envelope is, in some cases, set by the signal's modulation bandwidth, the video bandwidth is sometimes referred to as modulation bandwidth.

ValueTronics holds and stocks its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, so the instruments offered here are inventory we keep on hand rather than items sourced after the fact. That combination lets engineers and labs source the right power meter and sensor configuration with confidence in both condition and availability.

## Brand Heritage

Keysight Technologies traces its roots to Hewlett-Packard, whose test and measurement business was spun off as Agilent Technologies in 1999 and then established as the independent Keysight Technologies in 2014.

## Compare Other Models in This Series

The EPM-P series pairs the E4416A and E4417A power meters with the E-series E9320 family of peak and average power sensors. The E4416A is a single-channel meter and the E4417A is a dual-channel meter; both share the same measurement modes, interfaces, and reference architecture. Measurement capability — frequency coverage, power range, and dynamic range — depends on which sensor is connected.

The E9320 sensors come in six models that differ in frequency coverage and video bandwidth. The E9321A, E9322A, and E9323A cover 50 MHz to 6 GHz, while the E9325A, E9326A, and E9327A cover 50 MHz to 18 GHz, with video bandwidths of 300 kHz, 1.5 MHz, and 5 MHz spanning the pairs.

Each new model below is its own dedicated product page with condition-matched pricing. Select the meter or sensor that fits your application, and review that page for its specific configuration and availability.

The primary meter difference is channel count: the E4416A provides a single measurement channel, while the E4417A provides two channels and adds a second recorder output for channel B. Net weight differs slightly (approximately 4.0 kg versus 4.1 kg), and the E4417A ships with two E9288A sensor cables versus one for the E4416A.

Across the E9320 sensors, the differences are frequency range and video bandwidth: the E9321A/E9322A/E9323A reach 6 GHz and the E9325A/E9326A/E9327A reach 18 GHz, with video bandwidths of 300 kHz, 1.5 MHz, and 5 MHz respectively. The comparison table that follows details the per-model frequency range, video bandwidth, and power range.

Model

Type

Frequency Range

Video Bandwidth

**E9327A**

Power Sensor

50 MHz to 18 GHz

5 MHz

E4416A

Power Meter

9 kHz to 110 GHz, sensor dependent

5 MHz

E4417A

Power Meter

9 kHz to 110 GHz, sensor dependent

5 MHz

E9321A

Power Sensor

50 MHz to 6 GHz

300 kHz

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Calibration Factor (CF) and Reflection Coefficient (Rho) data sheet, unique to each sensor (matched by serial number)

## Product Core & Specifications

Specification

Value

Sensor Specification

Frequency range

50 MHz to 18 GHz

Video bandwidth

5 MHz

Power range (average only mode)

–60 dBm to +20 dBm

Power range (normal mode)

–40 dBm to +20 dBm

Maximum power

+23 dBm average; +30 dBm peak (< 10 µsec duration)

Connector type

Type N (m)

Measurement Ranges — Normal Mode

Lower range (min. power)

–40 dBm

Lower range (max. power)

–5 dBm

Upper range (min. power)

–30 dBm

Upper range (max. power)

+20 dBm

Auto range point (normal mode)

–15 dBm

Measurement Ranges — Average Only Mode

Lower range (min. power)

–60 dBm

Lower range (max. power)

–10.5 dBm

Upper range (min. power)

–35 dBm

Upper range (max. power)

+20 dBm

Auto range point (average only mode)

–11.5 dBm

Maximum SWR (≦ 0 dBm)

50 MHz to 2 GHz

1.14

\>2 GHz to 16 GHz

1.22

\>16 GHz to 18 GHz

1.26

Sensor Linearity

Normal mode (25 ± 10 °C / 0 to 55 °C)

± 4.2% / ± 5.0%

Average only mode (25 ± 10 °C / 0 to 55 °C)

± 3.7% / ± 5.0%

Additional linearity error if temperature changes after calibration (25 ± 10 °C / 0 to 55 °C)

± 1.0% / ± 1.0%

Zero Set

Normal mode / average only mode

60 nW / 0.6 nW

Zero Drift and Measurement Noise

Zero drift (normal mode / average only mode)

< 40 nW / < 100 pW

Measurement noise, normal mode (free run) / normal mode (VBW off, no averaging) / average only mode

< 25 nW / < 550 nW / < 400 pW

Measurement Uncertainty (typical, E9323A/E9327A, 5 MHz bandwidth, 25 °C)

–30 to –20 dBm

0.9%

–20 to –10 dBm

0.8%

–10 to 0 dBm

0.65%

0 to +10 dBm

0.55%

+10 to +20 dBm

0.45%

Rise and Fall Times (Low / Medium / High / Off)

Rise time (< µs)

0.9 / 0.4 / 0.2 / 0.2

Fall time (< µs)

0.9 / 0.4 / 0.2 / 0.2

Settling time, rising (< µs)

4.5 / 3.5 / 1.5 / 0.4

Settling time, falling (< µs)

4.5 / 3.5 / 2 / 0.4

Physical

Dimensions (L x W x H)

150 mm x 38 mm x 30 mm (5.9 in x 1.5 in x 1.2 in)

Net weight

0.2 kg (0.45 lbs)

Shipping weight

0.55 kg (1.2 lbs)

> **Important:** E9320 power sensors must be used with an E9288A, E9288B, or E9288C sensor cable.

> **Important:** A compatible power sensor is required to make measurements and is ordered separately — the EPM-P power meter's standard-shipped accessories include only the power sensor cable, power cord and manuals (no sensor). The EPM-P series operates with the E-series E9320 family for peak/average measurements and with the 8480, N8480, E-series CW and E9300 sensors for average power measurements.

> **Important:** The E9288A, B and C sensor cables will also operate with 8480, N8480 and E-series power sensors.

> **Important:** Rise and fall time specifications are only valid when used with the E9288A sensor cable (1.5 meters).

> **Important:** Calibration service options (R-50C-011-3 / R-50C-011-5) are not available in all countries.

> **Important:** To preserve the EMC performance of the product, any cable which becomes worn or damaged must be replaced with the same type and specification.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 7282.0 USD
- **Vendor:** Keysight
- **Type:** Pow Met
- **Tags:** Agilent, E9327, E9327A, Factory New, Inc., Keysight, Keysight Technologies, Manufacturers, Model 9327, Other Equipment, Pow Met, product_manual_group_3, Product_new, RF, Rfp Met, Sensor

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 7282.00 USD | In stock |

## Images

- E9327A Keysight Technologies RF Sensor

---

> Source: [ValueTronics](https://valuetronics.com/products/e9327a-keysight-rf-sensor-new)
> Updated: 2026-06-02