# Keysight Technologies N3306A 60V/120A 600W DC Electronic Load Module

# Keysight Technologies N3306A 60V/120A 600W DC Electronic Load Module

## Key Features At A Glance

-   Modular system: a 6-slot full-rack mainframe and a 2-slot half-rack mainframe accept any mix of six load modules
-   Module power ranges across the series from 150 W to 600 W, with current from 0–10 A to 0–120 A and voltage from 0–60 V to 0–240 V
-   Stable operation down to zero volts for testing low-voltage power supplies
-   16-bit voltage, current, and power measurement with simultaneous, linked V/I readings
-   GPIB, RS-232, and front-panel control, plus analog programming and monitoring ports
-   Constant current, constant voltage, and constant resistance modes with separately controlled rising and falling slew rates
-   Built-in transient generator (0.25 Hz–10 kHz) and downloadable List sequences for automated test
-   SCPI command set with VXIplug&play drivers for ATE integration

The Keysight N3300 Series is a family of modular DC electronic loads built around two mainframes and six interchangeable load modules. Designed to sink and dissipate power from a device under test, these loads let engineers exercise power supplies, DC-to-DC converters, and battery chargers without building custom dissipation hardware. The series emphasizes fast command processing—commands are handled more than ten times faster than previous-generation loads—along with configurable measurement and stable operation across a wide operating envelope.

Intended applications across the series include high-volume manufacturing test, where the goal is to maximize production throughput without expanding floorspace. Specific listed uses include testing low-voltage power supplies down to zero volts, testing multiple-output DC power supplies and DC-to-DC converters, and validating high-current supplies by operating load modules in parallel for additional current-sinking capability. The loads also serve as the programmable load element in power supply and battery charger test systems, with a DC connection terminal provided for ATE applications.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. That combination lets engineers source the right mainframe-and-module configuration with confidence in both availability and condition.

## Brand Heritage

Keysight Technologies traces its roots to Hewlett-Packard, whose test and measurement business was spun off as Agilent Technologies in 1999. Agilent's electronic measurement division was then established as Keysight Technologies in 2014, the corporate name under which this product line is sold today.

## Compare Other Models in This Series

The N3300 Series is built as a modular system of two mainframes and six load modules. The full-rack N3300A provides six slots and the half-rack N3301A provides two, and any assortment of modules can be installed up to each mainframe's slot capacity. This lets a single platform be tailored to a specific mix of power, current, and voltage requirements.

Four modules each occupy a single slot, while the two highest-power modules each occupy two slots. Because modules can be added or swapped, the same mainframe can grow or change as test requirements evolve, and modules can be operated in parallel to increase total current-sinking capability.

Each model below links to its own dedicated product page with condition-matched pricing. Whether you need a new mainframe or a new load module, select the specific model that matches your configuration to see its current availability and price.

The modules differ primarily in maximum power, input current, and input voltage. Maximum power per module ranges from 150 W up to 600 W, input current from 0–10 A up to 0–120 A, and input voltage from 0–60 V up to 0–240 V. The comparison table below lists each module's ratings side by side.

Slot count is the other key difference: the single-slot modules pack into either mainframe more densely, while the two 2-slot modules deliver the highest power and current at the cost of additional mainframe space. Selecting a module is a matter of matching its current, voltage, and power envelope to the supply or converter under test.

Model

Max Voltage

Max Current

Max Power

**N3306A**

60 V

120 A

600 W

N3300A

—

—

—

N3301A

—

—

—

N3302A

60 V

30 A

150 W

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Specification

Value

Input Ratings

Current

0–120 A

Voltage

0–60 V

Maximum power @ 40 °C

600 W (derated linearly from 100% at 40 °C to 75% at 55 °C)

Specified Current @ Low Voltage Operation

2.0 V

120 A

1.5 V

90 A

1.0 V

60 A

0.5 V

30 A

0 V

0 A

Typical minimum operating voltage @ full scale current

1.4 V

Constant Current Mode

Low range / High range

12 A / 120 A

Regulation

10 mA

Low range accuracy

0.1% + 15 mA

High range accuracy

0.1% + 37.5 mA

Constant Voltage Mode

Low range / High range

6 V / 60 V

Regulation

20 mV

Low range accuracy

0.1% + 3 mV

High range accuracy

0.1% + 8 mV

Constant Resistance Mode

Range 1 (I > 10% of current rating)

0.017–1 Ω

Range 2 (I > 1% of current rating)

0.9–10 Ω

Range 3 (I > 0.1% of current rating)

9–100 Ω

Range 4 (I > 0.01% of current rating)

90–1000 Ω

Current Measurement

Low range / High range

12 A / 120 A

Low range accuracy

0.05% + 10 mA

High range accuracy

0.05% + 20 mA

Voltage Measurement

Low range / High range

6 V / 60 V

Low range accuracy

0.05% + 3 mV

High range accuracy

0.05% + 8 mV

Power Measurement

Accuracy

0.1% + 1.3 W

Programming Resolution

Constant current mode

0.2/2 mA

Constant voltage mode

0.1/1 mV

Constant resistance mode

0.0175/0.175/1.75/17.5 mΩ

Readback Resolution

Current

0.2/2 mA

Voltage

0.1/1 mV

Programmable Slew Rate — Current

Slow band

2–100 kA/s

Fast band ≥3 V

200 k–10 MA/s

Fast band <3 V

200 k–1 MA/s

Programmable Slew Rate — Voltage

Slow band

1 k–50 kV/s

Fast band ≥3 V

100 k–500 kV/s

Fast band <3 V

100 k–50 kV/s

Programmable Slew Rate — Resistance Range 1

Slow band

11–280 Ω/s

Fast band ≥3 V

560–8.5 kΩ/s

Fast band <3 V

560–850 Ω/s

Programmable Slew Rate — Resistance Range 2

Slow band

110–2800 Ω/s

Fast band ≥3 V

5600–85 kΩ/s

Fast band <3 V

5600–8.5 kΩ/s

Programmable Slew Rate — Resistance Range 3

Slow band

1.1 k–28 kΩ/s

Fast band ≥3 V

56 k–850 kΩ/s

Fast band <3 V

56 k–85 kΩ/s

Programmable Slew Rate — Resistance Range 4

Slow band

11 k–280 kΩ/s

Fast band ≥3 V

560 k–8.5 MΩ/s

Fast band <3 V

560 k–850 kΩ/s

Programmable Short

Maximum

17 mΩ

Typical

12 mΩ

Programmable Open

Resistance

≥ 20 kΩ

Ripple and Noise (20 Hz–10 MHz)

Current (rms / peak to peak)

6 mA / 60 mA

Voltage (rms)

8 mV rms

External Analog Programming

Voltage programming accuracy

0.5% + 12 mV

Current programming accuracy

0.25% + 18 mA

External Monitor Ports

Voltage monitor accuracy

0.25% + 12 mV

Current monitor accuracy

0.1% + 18 mA

> **Important:** Maximum continuous power is derated linearly from 100% of maximum at 40 °C to 75% of maximum at 55 °C.

> **Important:** The N3302A through N3307A are plug-in load modules that require an N3300A (6-slot) or N3301A (2-slot) mainframe to operate; the mainframe is not included with a module and must be ordered separately.

> **Important:** No service manual is available for this product.

> **Important:** Calibration interval is one year for modules; the N3300A and N3301A mainframes do not require calibration.

> **Important:** Below 3 volts, the maximum bandwidth of the electronic load is reduced by a factor of ten to one; a fast pulse train with large transitions in transient mode may not be achievable at low voltages because the reduced slew rate can prevent the load from reaching the upper programmed value.

> **Important:** Specifications may degrade when the unit is subject to an RF field of 3 V/meter, line spikes of 500 V, or an 8 kV electrostatic discharge.

> **Important:** DC current accuracy specifications apply 30 seconds after input current is applied.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 5032.0 USD
- **Vendor:** Keysight Technologies
- **Type:** Dc Loads
- **Tags:** 3306, Agilent, Agilent Keysight, Agilent Technologies, consignment-active, DC Electronic Load, Factory New, Inc., Keysight, Keysight Technologies, Manufacturers, Model 3306, N3306, N3306A, Power Supplies & Loads, product_manual_group_2, product_manual_group_3, Product_new

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 5032.00 USD | In stock |

## Images

- N3306A Keysight Technologies DC Electronic Load

---

> Source: [ValueTronics](https://valuetronics.com/products/n3306a-keysight-dc-electronic-load-new)
> Updated: 2026-06-02