# Keysight DSOX2014A 100 MHz 4-Channel InfiniiVision 2000 X-Series Digital Storage Oscilloscope

# Keysight DSOX2014A 100 MHz 4-Channel InfiniiVision 2000 X-Series Digital Storage Oscilloscope

## Key Features At A Glance

-   Bandwidth options from 70 MHz to 200 MHz across the series with calculated rise times from ≤5 ns down to ≤1.75 ns
-   MegaZoom IV custom ASIC delivers waveform update rate up to 200,000 waveforms per second with 2 GSa/s sample rate and 1 Mpt memory per channel
-   8.5-inch WVGA color display (800 x 480) with 64 levels of intensity grading for signal visibility
-   Available as 2-channel or 4-channel DSO models, with field-upgradable MSO option adding 8 integrated digital timing channels
-   Five instruments in one: oscilloscope, 8-channel logic timing analyzer (MSO upgrade), 20 MHz WaveGen function generator (option), serial protocol triggering and decode (option), and integrated 3-digit digital voltmeter with 5-digit frequency counter
-   Hardware-based serial protocol decode and triggering available for I²C, SPI, UART/RS232/422/485, CAN, and LIN (optional licenses)
-   Standard segmented memory acquisition with up to 250 segments and re-arm time of 19 μs minimum
-   Field-upgradable bandwidth, digital channels, WaveGen, and serial protocol analysis — invest now and add capability later

The Keysight InfiniiVision 2000 X-Series is a family of entry-level benchtop oscilloscopes designed to deliver more capability for budget-conscious engineering and educational benches. The series spans six base models — the 2002A, 2004A, 2012A, 2014A, 2022A, and 2024A — covering 70 MHz, 100 MHz, and 200 MHz bandwidth tiers in both 2-channel and 4-channel DSO configurations, with mixed-signal (MSO) variants that add 8 integrated digital timing channels for time-correlated analog-and-digital debug.

Keysight positions the 2000 X-Series as a five-instruments-in-one platform for engineers, educators, and design labs that need bench versatility without buying separate instruments. The same chassis serves as a digital storage oscilloscope, an 8-channel logic timing analyzer (with the MSO upgrade), a 20 MHz WaveGen function generator (with the WaveGen option, including AM, FM, and FSK modulation capability), a hardware-based serial protocol triggering and decode tool for embedded buses (I²C, SPI, UART/RS232/422/485, CAN, LIN), and an integrated 3-digit digital voltmeter with a 5-digit frequency counter that operates through the same probes as the scope channels.

ValueTronics has supported professional test and measurement procurement since 1992, with all inventory housed and tested in our 20,000 sq ft secure facility at 1675 Cambridge Drive, Elgin, Illinois. {{CONDITION}} InfiniiVision 2000 X-Series oscilloscopes sold through ValueTronics ship from our domestic warehouse with the documentation and accessory verification that professional engineering and education customers expect when sourcing a multi-function bench instrument.

## Brand Heritage

Keysight Technologies was established as the electronic measurement business of Agilent Technologies in 2014, following Agilent's 1999 spinoff from Hewlett-Packard. The InfiniiVision product family carries the test and measurement engineering heritage of that HP-to-Agilent-to-Keysight lineage, which remains the longest continuous design history of any major oscilloscope brand in the industry.

## Compare Other Models in This Series

The InfiniiVision 2000 X-Series is offered as six base configurations spanning two channel counts (2 or 4 analog channels) and three bandwidth tiers (70 MHz, 100 MHz, and 200 MHz). Each base model is available as a DSO (analog-only) or as an MSO with 8 integrated digital timing channels added.

All six models share the same MegaZoom IV ASIC, 2 GSa/s maximum sample rate, 1 Mpt-per-channel memory, 8.5-inch WVGA display, 200,000 waveforms-per-second update rate, integrated 3-digit DVM and 5-digit counter, and standard segmented memory. The differences across the family are bandwidth, calculated rise time, channel count, and time base range.

Each new 2000 X-Series model below is its own dedicated product page with condition-matched pricing and the specific accessory and probe configuration listed for that page. Use the comparison table to identify the bandwidth and channel count that fits the application, then navigate to the model page for full details.

Bandwidth and rise time scale across the three tiers: the 2002A and 2004A are rated at 70 MHz with ≤5 ns calculated rise time, the 2012A and 2014A at 100 MHz with ≤3.5 ns, and the 2022A and 2024A at 200 MHz with ≤1.75 ns. The 200 MHz models also have an extended time base range starting at 2 ns/div, compared to 5 ns/div on the 70 MHz and 100 MHz models. Standard included passive probes are bandwidth-matched: N2841A 150 MHz probes ship with the 70 and 100 MHz models, and N2842A 300 MHz probes ship with the 200 MHz models.

Channel count is the second differentiator: the 2002A, 2012A, and 2022A are 2-channel DSOs, while the 2004A, 2014A, and 2024A are 4-channel DSOs. MSO variants (MSOX prefix) of each model add 8 digital timing channels for a total of 10 or 12 simultaneous channels. All other performance and feature characteristics — sample rate, memory depth, display, update rate, integrated DVM and counter, segmented memory, and upgrade paths — are identical across the family.

Model

Bandwidth

Analog Channels

Digital Channels

**DSOX2014A**

100 MHz

4

—

DSOX2002A

70 MHz

2

—

DSOX2004A

70 MHz

4

—

DSOX2012A

100 MHz

2

—

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   N2841A 150 MHz 10:1 passive probe (one per channel)
-   Localized power cord
-   Certificate of Calibration
-   Documentation CD

## Product Core & Specifications

Category

Specification

Value

General

Maximum sample rate

2 GSa/s half-channel interleaved, 1 GSa/s per channel

General

Maximum memory depth

1 Mpt per channel (standard)

General

Display

8.5-inch WVGA color TFT LCD with 64 levels of intensity grading, 800 x 480 resolution

General

Waveform update rate

200,000 waveforms per second

Vertical (Analog)

Input coupling

AC, DC

Vertical (Analog)

Input sensitivity range

1 mV/div to 5 V/div

Vertical (Analog)

Input impedance

1 MΩ ± 2% (11 pF)

Vertical (Analog)

Vertical resolution

8 bits (12 bits with averaging)

Vertical (Analog)

Dynamic range

± 8 divisions from center screen

Vertical (Analog)

Maximum input voltage

135 Vrms; 190 Vpk (not rated for CAT II, III, IV; no transient overvoltage allowed)

Vertical (Analog)

DC vertical gain accuracy

± 3% full scale (≥ 10 mV/div); ± 4% full scale (< 10 mV/div)

Vertical (Analog)

DC vertical offset accuracy

± 0.1 div ± 2 mV ± 1% of offset setting

Vertical (Analog)

Channel-to-channel isolation

≥ 40 dB from DC to maximum specified bandwidth

Vertical (Analog)

Position/offset range (1 MΩ)

1 mV to 200 mV/div: ± 2 V; > 200 mV to 5 V/div: ± 50 V

Vertical (Analog)

Hardware bandwidth limit

Approximately 20 MHz (selectable)

Horizontal

Time base accuracy

25 ppm ± 5 ppm per year (aging)

Horizontal

Pre-trigger range

Greater of 1 screen width or 200 µs (400 µs in interleaving mode)

Horizontal

Post-trigger range

1 s to 500 s

Horizontal

Channel-to-channel deskew range

± 100 ns

Acquisition

Modes

Normal, Peak detect, Averaging (2 to 65,536), High resolution (12 bits when ≥ 20 µs/div), Segmented (re-arm time 19 µs)

Acquisition

Peak detect

Capture glitch as narrow as 500 ps at all timebase settings

Trigger

Modes

Normal, Auto, Single, Force

Trigger

Coupling

AC, DC, noise reject, LF reject, HF reject

Trigger

Sensitivity (internal)

< 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div

Trigger

Sensitivity (external)

200 mV (DC to 100 MHz); 350 mV (100 to 200 MHz)

Trigger

Types

Edge, Edge then edge (B trigger), Pulse width, Pattern, Video, Runt, Rise/Fall time, Nth edge burst, Or, I²C, SPI, CAN, LIN, RS232/422/485/UART (serial types optional)

Math

Operators

Add, subtract, multiply, divide, FFT, Ax + B, Square, Absolute, Common Log, Natural Log, Exponential, Base 10 Exponential, LP Filter, HP Filter, Magnify, Measurement Trend, Chart Logic Bus (Timing or State)

Math

FFT windows

Hanning, flat top, rectangular, Blackman-Harris – up to 64 kpts resolution

MSO (Digital Channels)

Channels

8 channels (D0 to D7) on MSO models or via DSOX2MSO upgrade

MSO (Digital Channels)

Maximum sample rate

1 GSa/s

MSO (Digital Channels)

Maximum record length

500 kpts/ch (digital only); 125 kpts/ch (analog and digital)

MSO (Digital Channels)

Threshold selections

TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-definable (± 8.0 V in 10 mV steps)

MSO (Digital Channels)

Threshold accuracy

± (100 mV + 3% of threshold setting)

MSO (Digital Channels)

Maximum input voltage

± 40 V peak CAT I

MSO (Digital Channels)

Maximum input dynamic range

± 10 V about threshold

MSO (Digital Channels)

Minimum voltage swing

500 mVpp

MSO (Digital Channels)

Input impedance

100 kΩ ± 2% at probe tip, ~8 pF

MSO (Digital Channels)

Minimum detectable pulse width

5 ns

MSO (Digital Channels)

Channel-to-channel skew

2 ns (typical), 3 ns (maximum)

WaveGen (Optional)

Waveforms

Sine, square, pulse, triangle, ramp, noise, DC

WaveGen (Optional)

Sine frequency range

0.1 Hz to 20 MHz

WaveGen (Optional)

Square/pulse frequency range

0.1 Hz to 10 MHz

WaveGen (Optional)

Ramp/triangle frequency range

0.1 Hz to 100 kHz

WaveGen (Optional)

Amplitude range

20 mVpp to 5 Vpp into Hi-Z; 10 mVpp to 2.5 Vpp into 50 Ω

WaveGen (Optional)

DC offset range

± 2.5 V into Hi-Z; ± 1.25 V into 50 Ω

WaveGen (Optional)

Modulation types

AM, FM, FSK

DVM (Standard)

Functions

ACrms, DC, DCrms, frequency

**Important:** WaveGen, MSO digital channels, serial protocol decode, and bandwidth are all field-upgradable via license — a 2-channel 70 MHz DSO can be expanded after purchase up to a 4-channel 200 MHz MSO with full protocol decode and WaveGen using the model-specific upgrade licenses (DSOX2BW, DSOX2MSO, DSOX2WAVEGEN, D2000GENB/AUTB/BDLB).**Important:** The digital channels and serial protocol analysis cannot be used simultaneously on the 2000 X-Series.**Important:** Maximum input voltage on analog channels is 135 Vrms; 190 Vpk. This instrument is not rated for CAT II, III, or IV measurement categories, and no transient overvoltage is allowed. Use only for measurements within the specified measurement category.**Important:** Only one connectivity module (DSOXLAN or DSOXGPIB) may be installed and used at a time.Recommended pairing: the included N2841A 150 MHz 10:1 passive probes (70 and 100 MHz models) or N2842A 300 MHz 10:1 probes (200 MHz models) are matched to the bandwidth tier of each instrument; the N2755A 8-channel logic probe kit ships standard with all MSO models and with the DSOX2MSO upgrade.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 3831.0 USD
- **Vendor:** Keysight
- **Type:** Ms Oscilloscope
- **Tags:** Digital, Keysight, Keysight Technologies, Manufacturers, Oscilloscopes, product_manual_group_3, Product_new

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 3831.00 USD | In stock |

## Images

- DSOX2014A Keysight Technologies Digital Oscilloscope

---

> Source: [ValueTronics](https://valuetronics.com/products/dsox2014a-keysight-digital-oscilloscope-new)
> Updated: 2026-05-30