# 9518-01 Hioki Module New

**Features:**

**GP-IB Interface for 35XX LCR meters**

-   All of the functions of the LCR HiTESTER/C HiTESTER, except for powering on and off, can be controlled via the GP-IB interface
-   The beeper sound can be turned on and off.
-   The unit can be reset.
-   IEEE 488.2-1987 standard (essential) commands can be used.

**GP-IB Interface**

By connecting the 9518-01 GP-IB interface to the LCR HiTESTER/C HiTESTER, it is possible to control all the functions of the main unit (except for powering on and off) via the GP-IB bus. This unit is compliance wiht the following standard.

**Compliance standard: IEEE 488.1-1987**

Furthermore, the 9518-01 is designed with reference to the following standard:

**Reference standard: IEEE 488.2-1987**

On the 9518-01, if the output queue becomes full, it is cleared and a query error is generated. This differs from the IEEE 488.2 specification, which only stipulates the clearing of the output queue and the outputting of a query error when a deadlock state occurs, that is, when both the processing has become impossible.