# HIOKI RM3545-02 Resistance Meter with Multiplexer Support

# HIOKI RM3545-02 Resistance Meter with Multiplexer Support

## Key Features At A Glance

-   Four-terminal DC resistance measurement with guard terminal to minimize the effects of external noise
-   Accuracy guaranteed immediately at startup — no warm-up or zero-adjustment required
-   Bench-top meters for manual production-line testing and integration into automated inspection systems
-   Temperature correction, comparator (ABS/REF%) judgments, scaling, and auto hold across the series
-   RS-232C, USB, and isolated EXT I/O handler interfaces available for ATE integration (GP-IB on RM3545-01)
-   User-selectable, high-volume judgment tones plus panel save/load of measurement settings
-   Applicable standards: safety EN61010; EMC EN61326, EN61000-3-2, EN61000-3-3

The HIOKI RM3545 and RM3544 are high-accuracy, bench-top resistance meters that measure resistance using a 4-terminal direct-current (constant-current) method with a guard terminal. The RM3545 is positioned as a super-high-accuracy, multi-channel instrument for advanced development and production applications, while the RM3544 is an easy-to-use meter suited to both manual operation and integration with automatic lines. Both deliver guaranteed accuracy immediately after startup, with no warm-up or zero-adjustment required.

Listed measurement targets span motors, solenoids, choke coils, transformers, and wire harnesses; contacts, relay contacts, and switches; fuses, resistors, heaters, wires, and welds; conductive rubber and paint; and compact magnetic components such as EMC filters and ferrite beads. Application examples include measuring a 3-phase motor winding, testing a connector or wiring harness, testing a relay, and testing battery terminal welds. The RM3545 additionally addresses multi-contact resistance measurement of motor and transformer windings as well as small-signal contacts, including compact fuses and airbag inflators.

ValueTronics stocks and holds its own test-and-measurement inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, and has specialized in new and used instrumentation since 1992.

## Compare Other Models in This Series

The RM3545 and RM3544 form a two-model resistance-meter range, and the right choice depends on the application. The RM3545 is the super-high-accuracy, multi-channel model intended for advanced development and production applications, while the RM3544 is an easy-to-use, high-accuracy meter built for both manual operation on production lines and integration into automatic test lines.

Both share the same 4-terminal DC measurement method with a guard terminal, along with temperature correction, comparator judgments with user-selectable tones, scaling, and auto hold, and both run on a 100–240 VAC supply. The differences lie in measurement range, resolution, accuracy, and the advanced functions and interfaces each model offers, detailed in the comparison table below.

Each new model linked below is its own dedicated product page with condition-matched pricing. Select the model and configuration that fits your application to view the specifications and availability for that unit.

The RM3545 is distinguished within the range by functions the RM3544 does not provide, including low-power (LP) resistance measurement, offset voltage compensation (OVC), a temperature-conversion function for temperature-rise testing, D/A output of the measured value, and — in the RM3545-02 configuration — multiplexer-based scanning of up to 20 locations with 4-terminal measurement.

In measurement performance, the RM3545 provides a 6.5-digit (1,200,000-count) display, 0.01 μΩ maximum resolution, 0.006% basic accuracy, a measurement range to 1200 MΩ, and up to 1 A of measurement current. The RM3544 provides up to a 4.5-digit display, 1 μΩ maximum resolution, 0.02% basic accuracy, a range to 3.5 MΩ, and up to 300 mA of measurement current.

The models also differ by configuration. The base RM3544 has no communication interfaces, while the RM3544-01 adds EXT I/O, RS-232C, and USB. Among the RM3545 variants, the base RM3545 and RM3545-01 cannot accept multiplexer units; the RM3545-01 adds a GP-IB connector, and only the RM3545-02 supports the Z3003 Multiplexer Unit for 4-terminal scanning. Refer to the comparison table for a model-by-model breakdown.

Model

Resistance Range

Basic Accuracy

Max Resolution

**RM3545-02**

0.00 μΩ to 1200 MΩ

0.006%

0.01 μΩ

RM3545

0.00 μΩ to 1200 MΩ

0.006%

0.01 μΩ

RM3545-01

0.00 μΩ to 1200 MΩ

0.006%

0.01 μΩ

RM3544

0.000 mΩ to 3.5 MΩ

0.02%

1 μΩ

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Power cord ×1
-   CLIP TYPE LEAD L2101 ×1
-   Temperature Sensor Z2001 ×1
-   Male EXT I/O connector ×1
-   Instruction manual ×1
-   Application disc ×1
-   USB cable (A-to-B type) ×1
-   Spare fuse ×1

## Product Core & Specifications

Specification

Value

Measurement Types

Resistance measurement

0.000 00 mΩ (10 mΩ range) to 1200.0 MΩ (1000 MΩ range), 12 ranges

Low power resistance measurement (LP)

0.00 mΩ (1000 mΩ range) to 1200.00 Ω (1000 Ω range), 4 ranges

Temperature measurement (thermistor)

\-10.0 to 99.9 °C

Temperature measurement (analog input)

\-99.9 to 999.9 °C

General

Measurement method

4-terminal direct current (constant current), banana plug, with guard terminal

Range switching

Auto or Manual

Temperature correction

Reference temperature setting range: -10 °C to 99.9 °C; Temperature coefficient setting range: -99,999 ppm/°C to 99,999 ppm/°C

Zero-adjustment

By range, by step (RM3545-02 only); within ±50% f.s. of each range (not required for 100 MΩ or greater ranges)

Trigger

Internal or external

Measurement speed

FAST / MED / SLOW1 / SLOW2

Delay

Internal fixed value / 0 to 9999 ms (1 ms step)

Functions

Temperature correction, Temperature conversion, Self-calibration, offset voltage compensation (OVC), comparator (ABS/REF%), BIN, key-lock (OFF, menu lock, all lock), display digit count selection (7/6/5 digits), automatic power supply frequency settings (AUTO/50Hz/60Hz), scaling, judgment sound setting, auto hold, statistical calculations, clock, self-test, L2105 LED Comparator Attachment output

Measurement fault detection functions

Contact check, over detection, current fault detection

Averaging

OFF, 2 to 100 averaging iterations (variable in 1-iteration steps)

Panel store / panel load

30 (front terminals), 8 (MUX)

D/A output

Output: resistance measured value; Output voltage: 0 V DC to 1.5 V DC; Output impedance: 1 kΩ; Number of bits: 12 bit

EXT I/O

TRIG and other, BIN, BCD

Communication interfaces

Select from RS-232C, PRINTER (RS-232C), or USB (no GP-IB on RM3545-02); remote function, communications monitor function, data output function, memory (50 data)

RS-232C

Bit rates: 115,200 / 38,400 / 19,200 / 9,600 bps

USB

Class: CDC (COM mode), HID (USB keyboard mode)

Operating temperature and humidity

0 to 40 °C, 80% rh or less (non-condensating)

Storage temperature and humidity

\-10 to 50 °C, 80% rh or less (non-condensating)

Operating environment

Indoors, Pollution Degree 2, up to 2,000 m ASL

Power supply

Rated supply voltage: 100 to 240 VAC ±10%; Rated supply frequency: 50/60 Hz

Rated power consumption

40 VA

Insulation withstand potential

1.62 kV AC for 1 min. (with 10 mA cutoff current)

Dimensions

Approx. 215 W × 80 H × 306.5 D mm (8.46 in. W × 3.15 in. H × 12.07 in. D) (without projections)

Mass

Approx. 3.2 kg (112.9 oz) (not including Z3003)

Applicable standards

Safety: EN61010; EMC: EN61326, EN61000-3-2, EN61000-3-3

Multiplexer (RM3545-02)

Number of installed units

Max. 2

Measurement terminal settings

Front terminals / MUX (when MUX is selected, leads cannot connect to front terminals)

Support unit

Z3003

Number of channels

42; switching time 30 ms (reference value); up to 20 locations with 4-terminal measurement

Resistance Measurement Accuracy, LP OFF — ±(% rdg. + % f.s.); values in parentheses = OVC on

10 mΩ (max 12.000 00 mΩ, res 10 nΩ)

FAST 0.060+0.050 (0.060+0.015); MED 0.060+0.020 (0.060+0.002); SLOW1 0.060+0.020 (0.060+0.001); current 1 A; max open-terminal voltage 5.5 V

100 mΩ (max 120.000 0 mΩ, res 100 nΩ)

High: FAST 0.060+0.010, MED 0.060+0.010, SLOW1 0.060+0.010, current 1 A; Low: FAST 0.014+0.050, MED 0.014+0.020, current 100 mA

1000 mΩ (max 1200.000 mΩ, res 1 μΩ)

High: FAST 0.012+0.010, MED 0.012+0.008, current 100 mA; Low: FAST 0.008+0.050, MED 0.008+0.020, current 10 mA

10 Ω (max 12.000 00 Ω, res 10 μΩ)

High: FAST 0.008+0.010, MED 0.008+0.008, current 10 mA; Low: FAST 0.008+0.050, MED 0.008+0.020, current 1 mA

100 Ω (max 120.000 0 Ω, res 100 μΩ)

High: FAST 0.007+0.005, MED 0.007+0.002, SLOW1 0.007+0.001, current 10 mA; Low: FAST 0.008+0.010, MED 0.008+0.010, current 1 mA

1000 Ω (max 1200.000 Ω, res 1 mΩ)

FAST 0.007+0.005, MED 0.006+0.002, SLOW1 0.006+0.001, current 1 mA

10 kΩ (max 12.000 00 kΩ, res 10 mΩ)

FAST 0.008+0.005, MED 0.007+0.002, SLOW1 0.007+0.001, current 1 mA, 20 V

100 kΩ (max 120.000 0 kΩ, res 100 mΩ)

FAST 0.008+0.005, MED 0.007+0.002, SLOW1 0.007+0.001, current 100 μA

1000 kΩ (max 1200.000 kΩ, res 1 Ω)

FAST 0.015+0.005, MED 0.008+0.002, SLOW1 0.008+0.001, current 10 μA

10 MΩ (max 12.000 00 MΩ, res 10 Ω)

FAST 0.030+0.005, MED 0.030+0.002, SLOW1 0.030+0.001, current 1 μA

100 MΩ, high-precision ON (max 120.000 0 MΩ, res 100 Ω)

FAST 0.200+0.005, MED 0.200+0.002, SLOW1 0.200+0.001, current 100 nA

100 MΩ, high-precision OFF (max 120.00 MΩ, res 10 kΩ)

10.00 MΩ or less: 0.50+0.02; 10.01 MΩ or more: 1.00+0.02; current max. 1 μA

1000 MΩ, high-precision OFF (max 1200.0 MΩ, res 100 kΩ)

100.0 MΩ or less: 1.00+0.02; 100.1 MΩ or more: 10.00+0.02

Multiplexer Unit Z3003 — Additional Accuracy (added to RM3545 specifications)

30 mΩ (max 35.000 mΩ)

FAST 0.030+0.080, MED/SLOW 0.030+0.070, current 300 mA, open-circuit 5.5 V max.

300 mΩ (max 350.00 mΩ)

FAST 0.025+0.017, MED/SLOW 0.025+0.014, current 300 mA

3 Ω (max 3.500 0 Ω)

FAST 0.025+0.017, MED/SLOW 0.025+0.014, current 30 mA

30 Ω (max 35.000 Ω)

FAST 0.020+0.010, MED/SLOW 0.020+0.007, current 10 mA

300 Ω (max 350.00 Ω)

FAST 0.020+0.010, MED/SLOW 0.020+0.007, current 1 mA

3 kΩ (max 3.500 0 kΩ)

FAST 0.020+0.010, MED/SLOW 0.020+0.007, current 1 mA

30 kΩ (max 35.000 kΩ)

FAST 0.020+0.010, MED/SLOW 0.020+0.007, current 100 μA

300 kΩ (max 350.00 kΩ)

FAST 0.040+0.010, MED/SLOW 0.040+0.007, current 5 μA

3 MΩ (max 3.500 0 MΩ)

FAST 0.200+0.010, MED/SLOW 0.200+0.007, current 500 nA

> **Important:** Proper operation of the RM3545 and RM3544 is not guaranteed when using test leads (probes) that lack guard terminals; use the specified guard-terminal test leads.

> **Important:** The base RM3544 does not include EXT I/O or communication interfaces (RS-232C or USB); select the RM3544-01 for these functions.

> **Important:** Temperature correction requires the Temperature Sensor Z2001 or a thermometer capable of generating analog voltage output (e.g. an infrared thermometer). The Z2001 is a bundled accessory with the RM3545/RM3545-01/RM3545-02 but is an option (not bundled) for the RM3544/RM3544-01.

> **Important:** The Multiplexer Unit Z3003 cannot be used in combination with a withstand voltage tester; doing so will cause an insulation breakdown of the Z3003 internal relay, resulting in electric shock or equipment damage.

> **Important:** RM3545 only: accuracy requires a warm-up time of 60 minutes or greater; if less than 60 minutes, double the figures in the accuracy table to obtain the measurement accuracy.

> Recommended pairing: install the optional Multiplexer Unit Z3003 in the RM3545-02 to enable 4-terminal scanning measurement of up to 20 locations.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 3220.0 USD
- **Vendor:** HIOKI
- **Type:** Resistance Meter
- **Tags:** Factory New, Hioki, Inc., Manufacturers, Meters & DAQ, Model 3545, Multiplexer, product_manual_group_3, Product_new, Resistance, Resistance Meter, RM3545

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 3220.00 USD | In stock |

## Images

- HIOKI RM3545-02 Resistance Meter with Multiplexer Support

---

> Source: [ValueTronics](https://valuetronics.com/products/rm3545-02-hioki-resistance-meter-new)
> Updated: 2026-06-03