# HIOKI BT3561A 60V Battery HiTESTER

# HIOKI BT3561A 60V Battery HiTESTER

## Key Features At A Glance

-   Maximum voltage measurement of 60 V (BT3561A), 100 V (BT3562A), or 300 V (BT3563A) — select the model matching the battery pack voltage being tested
-   AC four-terminal method at 1 kHz ±0.2 Hz test frequency for stable internal resistance measurement of lithium-ion cells and battery packs
-   Resistance measurement from 3 mΩ to 3 kΩ range with basic accuracy of ±0.5% rdg. ±5 dgt. (±10 dgt. on the 3 mΩ range for BT3562A/BT3563A)
-   Voltage measurement basic accuracy of ±0.01% rdg. ±3 dgt. across the series
-   Doubled allowable total line resistance versus the prior 3561 generation, reducing measurement errors when using long measurement cables on production lines
-   LAN (TCP/IP, 10BASE-T/100BASE-TX) and RS-232C standard on all three models; 37-pin EXT. I/O handler interface and analog output (DC 0 V to 3.1 V) for automated test system integration
-   Improved electrostatic resistance as a countermeasure against electrostatic charges during battery transport on a production line
-   Comparator (Hi/IN/Lo), statistical calculations up to 30,000 points, 126 panel save/load locations, and 400-point internal memory storage

The Hioki BT3561A, BT3562A, and BT3563A Battery HiTesters are AC four-terminal internal resistance and DC voltage testers designed for battery cell and battery pack production lines. Each model uses a 1 kHz ±0.2 Hz AC test signal to measure internal resistance, paired with a DC voltage measurement channel. The three models share the same measurement architecture but differ in their maximum DC voltage measurement range — 60 V for the BT3561A, 100 V for the BT3562A, and 300 V for the BT3563A — allowing the operator to select the model that matches the cell or pack voltage being tested.

These instruments are positioned for fully automated production line testing across the lithium-ion battery value chain. The BT3561A targets high-performance cell applications such as electric bikes and power tools and battery packs of up to 60 V. The BT3562A is positioned for large xEV and ESS battery cells and for 72 V and 96 V packs. The BT3563A is positioned for large battery packs up to 300 V used in xEV and energy storage applications.

## Brand Heritage

Hioki E.E. Corporation is headquartered in Ueda, Nagano, Japan, with origins dating to 1935. The company has operated continuously under the Hioki name and has not undergone the brand-lineage transitions seen elsewhere in test and measurement, so a customer's Hioki instrument from any era traces directly to the same manufacturer.

## Compare Other Models in This Series

The Hioki BT3560A series is a three-model family of battery internal-resistance and DC-voltage testers sharing a common AC four-terminal measurement architecture, a 1 kHz ±0.2 Hz test frequency, and a common interface set (LAN, RS-232C, EXT. I/O handler interface, and analog output). The three models are positioned by the maximum DC voltage the user needs to measure and by which low-end resistance ranges are required.

The BT3561A is intended for cells and packs up to 60 V. The BT3562A extends the voltage measurement range to 100 V and is positioned for large xEV and ESS cells and for 72 V and 96 V packs. The BT3563A extends the voltage measurement range to 300 V for large xEV and ESS packs. All three share resistance accuracy and voltage accuracy specifications on their overlapping ranges, so the model selection is primarily driven by the device under test's voltage class and the low-end resistance ranges the application requires.

Each model in the table below is a separate product page with condition-matched pricing. The BT3561A, BT3562A, and BT3563A are listed individually so that each new unit's specifications, included accessories, and availability can be reviewed before the order is placed.

The three models differ in three primary axes that appear on the comparison table. First, maximum DC voltage measurement: 60 V on the BT3561A, 100 V on the BT3562A, and 300 V on the BT3563A. Second, lowest resistance range: the BT3561A's lowest range is 30 mΩ, while the BT3562A and BT3563A add a 3 mΩ range with 0.1 μΩ resolution at 100 mA test current. Third, voltage range layout: the BT3561A offers 6 V and 60 V ranges; the BT3562A adds a 100 V range; the BT3563A replaces the 100 V range with a 300 V range.

On shared specifications — measurement frequency, response time (10 ms), sampling periods, allowable total line resistance on the 300 mΩ and 3 Ω ranges, voltage basic accuracy, resistance basic accuracy on ranges 30 mΩ and above, and the connectivity set — the three models behave identically. The comparison table below shows the per-model differences without restating the shared architecture.

Model

Max Voltage

Min Resistance Range

Max Resistance Range

**BT3561A**

60 V

30 mΩ

3 kΩ

BT3562A

100 V

3 mΩ

3 kΩ

BT3563A

300 V

3 mΩ

3 kΩ

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Parameter

BT3561A Specification

Measurement method

AC four-terminal method

Measurement frequency

1 kHz ±0.2 Hz

Resistance range: 30 mΩ

31.000 mΩ, 1 μΩ, 100 mA

Resistance range: 300 mΩ

310.00 mΩ, 10 μΩ, 10 mA

Resistance range: 3 Ω

3.1000 Ω, 100 μΩ, 1 mA

Resistance range: 30 Ω

31.000 Ω, 1 mΩ, 100 μA

Resistance range: 300 Ω

310.00 Ω, 10 mΩ, 10 μA

Resistance range: 3 kΩ

3.1000 kΩ, 100 mΩ, 10 μA

Basic accuracy (30 mΩ range or more)

±0.5% rdg. ±5 dgt.

Voltage range: 6 V

6.00000 V, 10 μV

Voltage range: 60 V

60.0000 V, 100 μV

Voltage basic accuracy

±0.01% rdg. ±3 dgt.

Response time (typical)

10 ms

Sampling period (Ω or V) EX.FAST/FAST/MEDIUM/SLOW

4 ms, 12 ms, 35 ms, 150 ms

Sampling period (ΩV) EX.FAST/FAST/MEDIUM/SLOW

8 ms, 24 ms, 70 ms, 253 ms

Open terminal voltage (30 mΩ or less / 300 mΩ / 3 Ω or more)

25 V, 7 V, 4 V peak

LAN (TCP/IP, 10BASE-T/100BASE-TX)

Yes

RS-232C (Max. 38.4 kbps)

Yes

USB

No

GP-IB

No

EXT. I/O (37-pin Handler interface)

Yes

Analog output (DC 0 V to 3.1 V)

Yes

Contact check

Yes

Zero adjustment (±1000 counts)

Yes

Pulse measurement

Yes

Comparator

Hi / IN / Lo

Statistical calculations

Max. 30,000

Delay

Yes

Average

2 to 16 times

Panel saving/loading

126

Memory storage

400

LabVIEW® driver

Yes

Applicable standards

Safety: EN61010; EMC: EN61326 Class A

CE

Yes

CSA

Certification in progress

**Important:** The doubled allowable total line resistance applies on the 300 mΩ and 3 Ω ranges; the 3 mΩ and 30 mΩ ranges have tighter line-resistance limits (4 Ω SENSE / 4 Ω SOURCE within accuracy). Confirm the fixture wiring against the range you intend to use.**Important:** Measurement leads / probes are not included with the base unit. The appropriate 4-terminal measurement lead must be selected and ordered separately for the cell, module, or pack geometry under test.**Important:** CSA certification status is listed as 'Certification in progress' on the datasheet for BT3561A, BT3562A, and BT3563A — confirm current certification status with HIOKI prior to deployment in regions requiring CSA compliance.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** HIOKI
- **Type:** Battery Tester
- **Tags:** Battery Tester, Hioki, Manufacturers, Other Equipment, product_manual_group_3, Product_new

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- BT3561A Hioki Battery Testers

---

> Source: [ValueTronics](https://valuetronics.com/products/bt3561a-hioki-battery-testers-new)