# Fluke 5502A Multi-Product Calibrator (Pre-Owned)

# Fluke 5502A Multi-Product Calibrator (Pre-Owned)

## Key Features At A Glance

-   Calibrates a wide variety of electrical test equipment including 3.5 and 4.5 digit digital multimeters
-   Direct voltage to 1020 V and direct current to 20.5 A, with AC voltage from 1 mV to 1020 V (10 Hz to 500 kHz)
-   Sources resistance from 0 Ω to 1100 MΩ and capacitance from 220 pF to 110 mF
-   Sources and measures thermocouples (B, C, E, J, K, L, N, R, S, T, U) and sources RTD temperatures across multiple types
-   Robust protection circuits including reverse power protection and output disconnection for external voltages up to ±300 V peak
-   Optional oscilloscope calibration to 300 MHz or 600 MHz with leveled sine wave from 5 mV to 5.5 Vpp and edge rise times <300 ps
-   RS-232 and IEEE-488 interfaces for full automation with MET/CAL Plus Calibration Management Software
-   Output current extendable to 120 A using the 52120A Transconductance Amplifier, and to 6000 A with 25- and 50-turn coils

The Fluke 5502A Multi-Product Calibrator is a full-functioned calibrator that addresses common workload items such as 3.5 and 4.5 digit digital multimeters. It sources direct voltage and current; alternating voltage and current with multiple waveforms and harmonics; simultaneous voltage and current outputs or dual voltage outputs to simulate DC and AC power with phase control; as well as resistance, capacitance, thermocouples, and RTDs. The 5502A can also measure thermocouples and thermocouple simulators, and two options add the capability to calibrate oscilloscopes to either 300 MHz or 600 MHz.

The 5502A is designed for metrology professionals who need a solution for calibrating low-to-medium accuracy electrical instrumentation. Typical workload covered includes handheld and bench meters (analog and digital) to 4.5 digits, current clamps and clamp meters, panel meters, electronic thermometers, chart recorders, oscilloscope recorders, XY recorders, and data loggers. Temperature measurement modes calibrate thermocouple simulators and can also document environmental conditions present at the time of calibration, as required by all quality standards.

ValueTronics operates from a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where every instrument is received, inspected, and tested by our in-house technicians before it ships. This direct handling model — rather than drop-shipping from third-party sources — is how ValueTronics has built its reputation since 1992 for matching the right instrument, in verified condition, to each customer's actual workload.

## Brand Heritage

Fluke Calibration is the calibration-focused brand within Fluke Corporation. The 5502A is part of Fluke Calibration's Multi-Product Calibrator family, which also includes the 5080A and the higher-accuracy 5522A — each designed to match different combinations of workload and budget.

## Compare Other Models in This Series

The Fluke Calibration Multi-Product Calibrator family was created to give metrology professionals a single instrument that calibrates the widest range of electronic test tools, with each model in the family matched to a different combination of workload coverage and accuracy. The family currently includes the 5080A High Compliance Multi-Product Calibrator, the 5502A Multi-Product Calibrator, and the 5522A Multi-Product Calibrator.

The 5502A sits between the 5080A and the 5522A in the family. It addresses common workload such as 3.5 and 4.5 digit digital multimeters, with optional oscilloscope calibration to 300 MHz or 600 MHz and optional high-current extension to 120 A through the 52120A Transconductance Amplifier.

Each pre-owned 5502A configuration listed below is its own dedicated product page with condition-matched pricing, separate inventory tracking, and a configuration-specific accessory list. Selecting the page that matches the desired oscilloscope calibration option ensures the quote reflects the actual hardware on hand.

The 5502A is offered in three configurations: the base 5502A without oscilloscope calibration, the 5502A/3 with the 300 MHz oscilloscope calibration option, and the 5502A/6 with the 600 MHz oscilloscope calibration option. All three configurations share the same electrical, resistance, capacitance, temperature, and high-current capabilities; the difference between them is the bandwidth of the integrated oscilloscope calibrator.

Across the Multi-Product Calibrator family, the 5080A provides 100 ppm basic DC voltage accuracy and is positioned for analog and digital workload with high voltage and current compliance. The 5502A provides 50 ppm basic DC voltage accuracy and addresses common workload including 3.5 and 4.5 digit DMMs, with optional oscilloscope calibration to 300 MHz or 600 MHz. The 5522A provides 11 ppm basic DC voltage accuracy, addresses 5.5 and 6.5 digit DMMs, and supports oscilloscope calibration to 1.1 GHz with optional power quality functions.

The comparison table that follows summarizes the workload coverage, basic DC voltage accuracy, and optional capabilities across the three Multi-Product Calibrator models so that the calibrator can be matched to the workload it will actually support.

Model

Configuration

Oscilloscope Calibration

DC Voltage Accuracy

**5502A**

Base unit (no scope option)

Not included

50 ppm

5502A/3

With 300 MHz scope option

300 MHz

50 ppm

5502A/6

With 600 MHz scope option

600 MHz

50 ppm

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Function

Range

Direct volts

0 to ± 1020 V

Direct current

0 to ± 20.5 A

Alternating volts

1 mV to 1020 V, 10 Hz to 500 kHz

Volt/hertz

1000 V@10 kHz / 330 V@100 kHz

Alternating current

29 µA to 20.5 A, 10 Hz to 30 kHz

Waveforms

Sine, square, triangle, truncated sine

Resistance

0 Ω to 1100 MΩ

Capacitance

220 pF to 110 mF

Power (phantom loads)

20.9 kW

Phase control

0.01°

Thermocouple (source and measure temperature)

B, C, E, J, K, L, N, R, S, T, U; 10 µV/°C and 1 mV/°C

RTD (source temperature)

Pt 385-100 Ω, Pt 3926-100 Ω, Pt 3916-100 Ω, Pt 385-200 Ω, Pt 385-500 Ω, Pt 385 1000 Ω, PtNi 385-120 Ω (Ni120), Cu 427 10 Ω

Interfaces

RS-232, IEEE 488

Frequency uncertainty

< 25 ppm

Oscilloscope calibrator (options)

Levelled sine wave from 5 mV to 5.5 Vpp max, frequencies 50 kHz to 600 MHz; edge rise times of < 300 ps, multiple trigger functions, lowest dc, square wave and timing uncertainty

Amplified current (accessory amplifier)

Extend from 20.5 A to a maximum of 100 A dc and 120 A ac from 10 Hz to 10 kHz

Basic dc V accuracy

50 ppm

**Important:** Output protection covers applied external voltages up to ±300 V peak.**Important:** Internal protection circuitry (reverse power protection, check before connect, immediate output disconnection, and fuse protection on the output terminals) protects against common user errors for applied external voltages up to ± 300 V peak.**Important:** Oscilloscope calibration capability is provided by optional configurations only (5502A/3 for 300 MHz or 5502A/6 for 600 MHz); the base 5502A does not include oscilloscope calibration.**Important:** To extend current output above 20.5 A (up to 100 A dc / 120 A ac, or up to 6000 A using 25 and 50 turn coils), the separately-ordered 52120A Transconductance Amplifier is required.Recommended pairing: use the Fluke 52120A Transconductance Amplifier with the 5502A to extend current output from 20.5 A up to 100 A DC and 120 A AC, with 25- and 50-turn coils available for workload requiring up to 6000 A.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Fluke Calibration
- **Type:** Calibrator
- **Tags:** Calibrator, Fluke, Manufacturers, Multifunction, Other Equipment, product_manual_group_3, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- Fluke 5502A Multi-Product Calibrator (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/5502a-fluke-multifunction-calibrator-used)
> Updated: 2026-05-30
