# Fluke 5800A 600 MHz Oscilloscope Calibrator (Pre-Owned)

# Fluke 5800A 600 MHz Oscilloscope Calibrator (Pre-Owned)

## Key Features At A Glance

-   Volt function delivers 0V to ±130V into 1 MΩ load with ±(0.025% of output + 25 µV) 1-year absolute uncertainty
-   Edge function provides 4.0 mV to 2.5V p-p amplitude with ≤300 ps rise time across 1 kHz to 10 MHz
-   Leveled sine wave function covers 50 kHz to 600 MHz with 5 mV to 5.5V p-p amplitude range
-   Time marker function spans 5s to 2 ns with ±1 ppm absolute uncertainty across most ranges
-   Pulse generator delivers ≤1.5 ns typical rise/fall times with 4 ns to 500 ns pulse width range
-   Oscilloscope input impedance measurement: 40Ω to 1.5 MΩ resistance and 5 pF to 50 pF capacitance
-   Standard IEEE-488 (GPIB) and RS-232 interfaces for automated calibration system integration
-   TV trigger signal output supports selectable NTSC, SECAM, and PAL-M frame formats

The Fluke 5800A Oscilloscope Calibrator is a precision calibration source designed to verify and calibrate the performance of oscilloscopes across voltage, time, frequency, and impedance parameters. The instrument provides discrete calibration functions including DC and square wave voltage outputs, edge signals for rise time verification, leveled sine waves for bandwidth measurement, time markers for timebase calibration, wave and pulse generators, and oscilloscope input impedance and capacitance measurement capabilities.

Oscilloscope calibrators are used by metrology laboratories, calibration service providers, and instrument manufacturers to verify oscilloscope performance against published specifications. The 5800A supports calibration of oscilloscope voltage accuracy, timebase accuracy, bandwidth and flatness, rise time response, input impedance, and trigger functionality. The included TV trigger signal output with selectable NTSC, SECAM, and PAL-M frame formats supports calibration of oscilloscopes used in broadcast and video applications.

## Brand Heritage

Fluke Corporation, headquartered in Everett, Washington, has manufactured precision electronic test instruments since 1948. Fluke was acquired by Danaher Corporation in 1998 and was subsequently transferred to Fortive Corporation when Danaher spun off its industrial businesses in 2016. The Fluke brand and product lines have remained continuous through these corporate transitions, with calibration and metrology products continuing under the established Fluke name.

## Product Core & Specifications

Function

Specification

Volt Function

DC Signal Amplitude Range (into 50Ω)

0V to ±6.6V

DC Signal Amplitude Range (into 1 MΩ)

0V to ±130V

Square Wave Signal Amplitude Range (into 50Ω)

±1 mV to ±6.6V p-p

Square Wave Signal Amplitude Range (into 1 MΩ)

±1 mV to ±130V p-p

DC 1-Year Absolute Uncertainty, tcal ±5°C (into 50Ω)

±(0.25% of output + 40 µV)

DC 1-Year Absolute Uncertainty, tcal ±5°C (into 1 MΩ)

±(0.025% of output + 25 µV)

Square Wave 1-Year Absolute Uncertainty (into 50Ω)

±(0.25% of output + 40 µV)

Square Wave 1-Year Absolute Uncertainty (into 1 MΩ)

±(0.05% of output + 5 µV)

Sequence

1-2-5 (e.g., 10 mV, 20 mV, 50 mV)

Frequency Range

10 Hz to 10 kHz

Edge Function (into 50Ω)

Amplitude Range (p-p)

4.0 mV to 2.5V

Frequency Range

1 kHz to 10 MHz

Rise Time

≤300 ps

1-Year Absolute Uncertainty: Amplitude

±(2% of output + 200 µV)

1-Year Absolute Uncertainty: Frequency

±(2.5 ppm of setting)

1-Year Absolute Uncertainty: Rise Time

(+0 ps / -100 ps)

Typical Jitter, Edge to Trigger

<5 ps (p-p)

Leading Edge Aberrations: Within 2 ns from 50% of rising edge

<(3% of output + 2 mV)

Leading Edge Aberrations: 2 ns to 5 ns

<(2% of output + 2 mV)

Leading Edge Aberrations: 5 ns to 15 ns

<(1% of output + 2 mV)

Leading Edge Aberrations: after 15 ns

<(0.5% of output + 2 mV)

Tunnel Diode Pulse Drive

Square wave at 100 Hz to 100 kHz, with variable amplitude of 60V to 100V p-p

Leveled Sine Wave Function (into 50Ω)

Amplitude Range (p-p)

5 mV to 5.5V

1-Year Absolute Uncertainty @ 50 kHz (reference)

±(2% of output + 300 µV)

1-Year Absolute Uncertainty, 50 kHz to 100 MHz

±(3.5% of output + 300 µV)

1-Year Absolute Uncertainty, 100 MHz to 300 MHz

±(4% of output + 300 µV)

1-Year Absolute Uncertainty, 300 MHz to 600 MHz

±(6% of output + 300 µV)

Flatness (relative to 50 kHz), 50 kHz to 100 MHz

±(1.5% of output + 100 µV)

Flatness, 100 MHz to 300 MHz

±(2% of output + 100 µV)

Flatness, 300 MHz to 600 MHz

±(4% of output + 100 µV)

Short-term Amplitude Stability

≤1% (within one hour after reference amplitude setting, ±5°C)

Frequency Resolution

10 kHz

Frequency 1-Year Absolute Uncertainty

±1 ppm

Distortion: 2nd Harmonic

≤-33 dBc

Distortion: 3rd and Higher Harmonics

≤-38 dBc

Time Marker Function (into 50Ω)

1-Year Absolute Uncertainty, 5s to 50 ms

±(2.5 ppm + 5 µHz)

1-Year Absolute Uncertainty, 20 ms to 100 ns

±1 ppm

1-Year Absolute Uncertainty, 50 ns to 20 ns

±1 ppm

1-Year Absolute Uncertainty, 10 ns

±1 ppm

1-Year Absolute Uncertainty, 5 ns to 2 ns

±1 ppm

Wave Shape, 5s to 50 ms

Pulsed, sawtooth or square

Wave Shape, 20 ms to 100 ns

Pulsed, sawtooth, square, 1/5 pulse

Wave Shape, 50 ns to 20 ns

Pulsed, sawtooth or square

Wave Shape, 10 ns

Square or sine

Wave Shape, 5 ns to 2 ns

Sine

Typical Jitter (p-p), 5s to 50 ms

<10 ppm

Typical Jitter (p-p), all other ranges

<1 ppm

Sequence

5-2-1 from 5s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)

Wave Generator (Square, Sine, Triangle into 50Ω or 1 MΩ)

Amplitude Range (into 1 MΩ)

1.8 mV to 55V p-p

Amplitude Range (into 50Ω)

1.8 mV to 2.5V p-p

1-Year Absolute Uncertainty, tcal ±5°C, 10 Hz to 10 kHz

±(3% of p-p output + 100 µV)

Typical DC Offset Range

0 to ±(≥40% of p-p amplitude); dc offset plus wave signal must not exceed 30V rms

Frequency Range

0.01 Hz to 100 kHz

Pulse Generator (Positive Pulse into 50Ω)

Typical Rise/Fall Times

≤1.5 ns

Amplitude Available

Discrete steps: 1V, 250 mV, 100 mV, 25 mV, 10 mV

Pulse Width Range

4 ns to 500 ns (not to exceed 40% of period)

Pulse Width Uncertainty (typical)

5% + 2 ns

Pulse Period Range

20 ms to 200 ns (50 Hz to 5.0 MHz)

Pulse Period 1-Year Absolute Uncertainty

±1 ppm

Trigger Functions

External Trigger Functions

Available for pulse, time mark, edge and voltage generator functions

TV Trigger Frame Formats

Selectable NTSC, SECAM, PAL, PAL-M

TV Trigger Polarity

Selectable inverted or uninverted video

TV Trigger Line Marker

Selectable Line Video Marker

Measurement Functions (Oscilloscope Input Impedance)

Resistance Range / Uncertainty: 40Ω to 60Ω

0.1%

Resistance Range / Uncertainty: 500 kΩ to 1.5 MΩ

0.1%

Capacitance Range / Uncertainty: 5 pF to 50 pF

5% ±0.5 pF

Overload Measurement Function

Source Voltage

5V to 9V

Typical Maximum Time Limit dc or ac 1 kHz ac

Settable from 5 to 60 sec

Auxiliary Input

Frequency Range

Up to 2 GHz

Voltage Range

0-40V p-p

VSWR

<1.7 @ 2 GHz

External Frequency Reference Input

Reference Frequency

10 MHz

General Specifications

Warm-up Time

Twice the time since last warmed up, to a maximum of 30 minutes

Settling Time

5 seconds or faster for all functions and ranges

Standard Interfaces

IEEE-488 (GPIB), RS-232

Temperature: Operating

0 °C to 50 °C

Temperature: Calibration (tcal)

15 °C to 35 °C

Temperature: Storage

\-20 °C to 70 °C

Temperature Coefficient

0.1X/°C of 1-year specification for temperatures outside tcal +5 °C

Relative Humidity: Operating

<80% to 30 °C, <70% to 40 °C, <40% to 50 °C

Relative Humidity: Storage

<95%, non-condensing

Altitude: Operating

3,050m (10,000 ft) maximum

Altitude: Non-operating

12,200m (40,000 ft) maximum

**Important:** Designed to operate in Standard Laboratory environments where the Electromagnetic environment is highly controlled. If used in areas with Electromagnetic fields greater than 1 V/m, there could be errors in output values.**Important:** The dc offset plus the wave signal must not exceed 30V rms (Wave Generator).**Important:** Pulse width not to exceed 40% of period (Pulse Generator).**Important:** Short-term Amplitude Stability of ≤1% applies within one hour after reference amplitude setting, provided temperature varies no more than ±5°C.**Important:** The Overload Measurement Function applies dc or ac (1 kHz square wave) power into the 50Ω oscilloscope input and monitors the current. In order to prevent oscilloscope front end damage, a limited amount of energy is applied by a user-settable time limit.Recommended pairing: the 5800A/TDP Tunnel Diode Pulser accessory provides a square wave at 100 Hz to 100 kHz with variable amplitude of 60V to 100V p-p for tunnel diode pulse drive applications listed in the datasheet's edge function specifications.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 2520.0 USD
- **Vendor:** Fluke
- **Type:** Calibrator
- **Tags:** Calibrator, Fluke, Manufacturers, Oscilloscope, Other Equipment, product_manual_group_2, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 0.00 USD | In stock |
| NIST Traceable with Full Data | 2100.00 USD | In stock |
| ISO IEC 17025 Accredited | 2520.00 USD | In stock |

## Images

- Fluke 5800A 600 MHz Oscilloscope Calibrator (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/5800a-fluke-oscilloscope-calibrator-used)
> Updated: 2026-05-30