# Keysight N9010B EXA X-Series Multi-touch Signal Analyzer (Pre-Owned)

# Keysight N9010B EXA X-Series Multi-touch Signal Analyzer (Pre-Owned)

## Key Features At A Glance

-   Frequency coverage from 10 Hz up to 44 GHz across the N9010B option range
-   10.6-inch (269 mm) capacitive multi-touch display at 1280 x 768 resolution
-   LAN (1000Base-T), GPIB (IEEE-488), and USB 2.0/3.0 interfaces for automated test systems
-   25 MHz analysis bandwidth standard, expandable to 40 MHz with Option B40
-   PowerSuite power measurements: channel power, ACP, occupied bandwidth, CCDF, and spectrum emission mask
-   Maximum safe input of +30 dBm (1 W) average total power
-   Optional preamplifier, precision frequency reference (Option PFR), and Noise Floor Extension (Option NF2)
-   Removable solid-state drive, ≥80 GB nominal internal storage

The Keysight N9010B is the EXA X-Series signal analyzer with a capacitive multi-touch interface, positioned by Keysight as a cost-effective, economy-class instrument for signal analysis extending up to millimeter-wave frequencies. It is built around delivering maximum value in signal analysis—helping an engineer find the answer faster, whether the goal is tighter design margins or shorter test times.

A signal analyzer measures and characterizes signals in the frequency domain, resolving their amplitude, frequency, and spectral content. The N9010B carries a PowerSuite set of measurements—channel power, occupied bandwidth, adjacent channel power, power statistics CCDF, harmonic distortion, third-order intermod, burst power, spurious emission, and spectrum emission mask—with results referenced to communications formats including W-CDMA, IS-95, and cdma2000. With Option EMC the analyzer adds CISPR-compliant EMI bandwidths (200 Hz, 9 kHz, 120 kHz, 1 MHz) and MIL-STD-461E-compliant EMI bandwidths (10 Hz to 1 MHz).

ValueTronics stocks the equipment it sells from its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified in-house by our technicians before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

## Brand Heritage

Keysight Technologies traces its instrument heritage to Hewlett-Packard, whose test and measurement business was spun out as Agilent Technologies in 1999. That measurement business was then established as Keysight Technologies in 2014, which is the brand under which this analyzer is produced today.

## Product Core & Specifications

Specification

Value

Frequency Range (DC coupled)

Option 503

10 Hz to 3.6 GHz

Option 507

10 Hz to 7 GHz

Option 513

10 Hz to 13.6 GHz

Option 526

10 Hz to 26.5 GHz

Option 532

10 Hz to 32 GHz

Option 544

10 Hz to 44 GHz

Frequency Reference

Aging rate (Option PFR)

± 1 x 10–7 / year

Aging rate (Standard)

± 1 x 10–6 / year

Temperature stability, 20 to 30 °C (Option PFR)

± 1.5 x 10–8

Temperature stability, 20 to 30 °C (Standard)

± 2 x 10–6

Achievable initial calibration accuracy (Option PFR)

± 4 x 10–8

Achievable initial calibration accuracy (Standard)

± 1.4 x 10–6

Frequency Span

Range

0 Hz (zero span), 10 Hz to maximum frequency of instrument

Resolution

2 Hz

Accuracy (swept)

± (0.25 % x span + horizontal resolution)

Accuracy (FFT)

± (0.10 % x span + horizontal resolution)

Sweep Time and Triggering

Range, Span ≥ 10 Hz

1 ms to 4000 s

Accuracy, Span ≥ 10 Hz swept

± 0.01% nominal

Accuracy, Span ≥ 10 Hz FFT

± 40% nominal

Accuracy, Span = 0 Hz

± 0.01% nominal

Trigger sources

Free run, line, video, external 1, external 2, RF burst, periodic timer

Trigger delay resolution

0.1 μs

Time Gating

Gate methods

Gated LO; gated video; gated FFT

Gate length range (except FFT)

100.0 ns to 5.0 s

Gate delay range

0 to 100.0 s

Gate delay jitter

33.3 ns p-p nominal

Resolution Bandwidth (RBW)

Range (–3.01 dB bandwidth)

1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz

Bandwidth accuracy (power), 1 Hz to 750 kHz

± 1.0 % (± 0.044 dB)

Bandwidth accuracy (power), 820 kHz to 1.2 MHz (< 3.6 GHz CF)

± 2.0 % (± 0.088 dB)

Selectivity (–60 dB/–3 dB)

4.1:1 nominal

EMI bandwidth (CISPR compliant, Option EMC)

200 Hz, 9 kHz, 120 kHz, 1 MHz

EMI bandwidth (MIL STD 461E compliant, Option EMC)

10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz

Sweep (trace) point range

1 to 40001

Analysis Bandwidth

Maximum bandwidth (Option B40)

40 MHz

Maximum bandwidth (Standard)

25 MHz

Video Bandwidth (VBW)

Range

1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)

Accuracy

± 6 % nominal

Amplitude Range

Measurement range

Displayed average noise level (DANL) to +23 dBm

Input attenuator range, 10 Hz to 44 GHz (Standard)

0 to 60 dB in 10 dB steps

Input attenuator range (Option FSA)

0 to 60 dB in 2 dB steps

Electronic attenuator (Option EA3), range

0 to 24 dB, 1 dB steps

Electronic attenuator (Option EA3), full range (mechanical + electronic)

0 to 84 dB, 1 dB steps

Maximum Safe Input Level

Average total power (with and without preamp)

+30 dBm (1 W)

Peak pulse power (< 10 μs pulse width, < 1 % duty cycle, input attenuation ≥ 30 dB)

+50 dBm (100 W)

DC volts (DC coupled)

± 0.2 Vdc

DC volts (AC coupled)

± 100 Vdc

Display Range

Log scale

0.1 to 1 dB/division in 0.1 dB steps; 1 to 20 dB/division in 1 dB steps (10 display divisions)

Linear scale

10 divisions

Scale units

dBm, dBmV, dBμV, dBmA, dBμA, V, W, A

Frequency Response (Specification, 10 dB input attenuation, 20 to 30 °C, RF/MW Opt 503/507/513/526)

9 kHz to 10 MHz

± 0.8 dB

10 MHz to 3.6 GHz

± 0.6 dB

3.5 to 7.0 GHz

± 2.0 dB

7.0 to 13.6 GHz

± 2.5 dB

13.5 to 22.0 GHz

± 3.0 dB

22.0 to 26.5 GHz

± 3.2 dB

Frequency Response (Specification, Millimeter-wave Opt 532/544)

9 kHz to 10 MHz

± 0.6 dB

10 to 50 MHz

± 0.45 dB

50 MHz to 3.6 GHz

± 0.45 dB

3.5 to 5.2 GHz

± 1.7 dB

5.2 to 8.4 GHz

± 1.5 dB

8.3 to 13.6 GHz

± 2.0 dB

13.5 to 17.1 GHz

± 2.0 dB

17.0 to 22.0 GHz

± 2.0 dB

22.0 to 26.5 GHz

± 2.5 dB

26.4 to 34.5 GHz

± 2.5 dB

34.4 to 44 GHz

± 3.2 dB

Total Absolute Amplitude Accuracy (10 dB attenuation, 20 to 30 °C)

At 50 MHz

± 0.40 dB

At all frequencies

± (0.40 dB + frequency response)

Input VSWR (≥ 10 dB input attenuation)

10 MHz to 3.6 GHz (Opt 503/507/513/526)

< 1.2:1 nominal

3.6 to 26.5 GHz (Opt 503/507/513/526)

< 1.9:1 nominal

10 MHz to 3.6 GHz (Opt 532/544)

1.2:1 nominal

3.6 to 26.5 GHz (Opt 532/544)

1.5:1 nominal

26.5 to 44 GHz (Opt 532/544)

< 1.8:1 nominal

Reference Level

Range (Log scale)

–170 to +23 dBm in 0.01 dB steps

Range (Linear scale)

707 pV to 3.16 V

Display scale fidelity (between –10 dBm and –80 dBm input mixer level)

± 0.15 dB total

Trace detectors

Normal, peak, sample, negative peak, log power average, RMS average, voltage average

Preamplifier (Option P03/P07/P13/P26/P32/P44)

Frequency range (Option P03)

100 kHz to 3.6 GHz

Frequency range (Option P07)

100 kHz to 7 GHz

Frequency range (Option P13)

100 kHz to 13.6 GHz

Frequency range (Option P26)

100 kHz to 26.5 GHz

Frequency range (Option P32)

100 kHz to 32 GHz

> **Important:** Maximum safe input level: +30 dBm (1 W) average total power; peak pulse power +50 dBm (100 W) for <10 μs pulse width and <1% duty cycle with input attenuation ≥ 30 dB. DC volts: ±0.2 Vdc DC coupled, ±100 Vdc AC coupled.

> **Important:** External mixing (Option EXM) is available only with the EXA millimeter-wave configurations, Option 532 or 544.

> **Important:** If Auto Align is set to Light, performance is not warranted, and nominal performance will degrade to become a factor of 1.4 wider for any specification subject to alignment, such as amplitude tolerances.

> **Important:** When Option MPB (microwave preselector bypass) is installed and enabled, some aspects of the analyzer performance change; refer to the EXA specification guide for details.

> Usage tip: the rear-panel +28 V (pulsed) noise source drive is designed for use with Keysight SNS Series noise sources.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Keysight
- **Type:** Spec An
- **Tags:** Agilent, Agilent Keysight, Analyzers, Inc., Keysight, Keysight Technologies, Manufacturers, Model 9010, N9010, N9010B, Pre-Owned, product_manual_group_2, Product_used, Signal, Spec An, Spectrum

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- Keysight N9010B EXA X-Series Multi-touch Signal Analyzer (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/n9010b-agilent-spectrum-analyzer-used)
> Updated: 2026-06-02