# Fluke Calibration 5627A-9 Precision Industrial PRT, 9 in, -200 °C to 300 °C

# Fluke Calibration 5627A-9 Precision Industrial PRT, 9 in, -200 °C to 300 °C

## Key Features At A Glance

-   Thermometer readouts for calibrating PRTs, RTDs, thermistors, and thermocouples in field and laboratory environments
-   Readout options span single-channel handheld units to 40-channel benchtop systems with data logging and graphing
-   Reference probe selection covers Secondary Reference PRT, Precision Industrial PRT, Secondary Standard PRT, Full Immersion PRT, and Thermistor Secondary Probe
-   Probe temperature ranges extend from -200 °C to 661 °C depending on model selected
-   Combined system accuracy as tight as ±0.008 °C achievable with 1502A/1504 paired with the 5628 Secondary Standard PRT at -200 °C
-   Five-step selection workflow: choose readout, select probe, determine system accuracy, verify against sensor TAR requirement, evaluate calibration needs
-   Factory calibration traceable to national standards included as standard on most readouts and probes; ISO 17025 accredited calibration standard or optional depending on model

This Fluke Calibration selection guide covers a family of industrial temperature readouts and companion reference standard probes used to calibrate common temperature sensors such as PRTs, RTDs, thermistors, and thermocouples. The readout lineup spans handheld field units (1523 single-channel and 1524 two-channel), benchtop precision readouts (1502A and 1504), the 1529 Chub-E4 four-channel thermometer, and the 1586A Super-DAQ Precision Temperature Scanner with up to 40 channels via the DAQ-STAQ Multiplexer. Each readout pairs with a defined set of reference probes whose connector type matches the readout: P-type smart Lemo connectors for the 1523/1524, D-type DIN connectors for the 1502A/1504, and L-type gold-plated mini spade lugs for the 1529 and 1586A.

These instruments are used to calibrate industrial temperature sensors by placing the sensor under test in a stable temperature source — a dry-well, calibration bath, or furnace — and comparing its output to a reference standard probe connected to the readout. The guide identifies typical applications including improving calibration accuracy in baths and dry-wells and comparing against process instrumentation in-situ, such as a thermowell adjacent to a gauge or transmitter. The 1586A specifically supports temperature source control of Fluke Calibration dry-wells, baths, and furnaces to help automate sensor calibration workflows.

ValueTronics International maintains a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where the precision temperature instruments described in this guide are stored, inspected, and tested before shipment. Operating since 1992, ValueTronics handles each readout and reference probe through in-house verification procedures, providing a level of accountability that drop-shippers and pass-through brokers cannot match. For precision metrology instruments where measurement uncertainty traces directly to the condition of the equipment delivered, that hands-on inspection step is what separates a known-state instrument from an unknown-state one.

## Brand Heritage

Fluke Calibration is the calibration and metrology division of Fluke Corporation. The Hart Scientific temperature calibration product lines, which include several of the readout and probe families covered in this guide, became part of Fluke Calibration following Fluke's acquisition of Hart Scientific in 2005. Fluke Corporation itself has operated as a subsidiary of Danaher Corporation since 1998, and in 2016 Danaher spun off its test and measurement businesses — including Fluke — into Fortive Corporation, where the company continues to operate today.

## Compare Other Models in This Series

The readouts in this family span four distinct form factors and capability tiers. The 1523 and 1524 are handheld battery-powered units intended for field use, with the 1523 providing a single channel and the 1524 providing two channels (one of which may be a thermocouple). The 1502A and 1504 are single-channel benchtop readouts; the 1502A measures PRTs and RTDs, while the 1504 measures thermistors. The 1529 Chub-E4 is a four-channel benchtop readout offered in configurations supporting PRTs, thermistors, and thermocouples in various combinations.

All six readouts in the guide measure PRTs and RTDs. Thermistor measurement is supported on the 1523, 1524, 1504, 1529, and 1586A. Thermocouple measurement is supported on the 1523, 1524, 1529, and 1586A. Connector types differ by model: P-type smart Lemo (1523/1524), D-type DIN (1502A/1504), and L-type gold-plated mini spade lugs (1529/1586A).

Each new model below is its own product page with condition-matched pricing. Configuration, connector type, and channel count vary across the family, so the right model depends on the specific calibration workflow being supported.

The primary differentiators across the family are channel count, portability, and supported sensor types. The 1523 (single channel) and 1524 (two channels) are handheld battery-powered units with 25-reading and 15,000-reading data capacity respectively, with graphing for one channel on the 1523 and two channels on the 1524. The 1502A and 1504 are single-channel benchtop units; the 1502A handles PRTs and RTDs while the 1504 handles thermistors, with neither supporting data logging or graphing. The 1529 supports four channels with 8,000-reading data capacity, automated data logging, and statistical functions including min, max, and average.

System accuracy also varies meaningfully across the family. Paired with a 5628 Secondary Standard PRT at 0 °C, the 1502A/1504 achieves ±0.009 °C, the 1529 achieves ±0.009 °C, the 1586A with DAQ-STAQ achieves ±0.011 °C, and the 1523/1524 achieves ±0.035 °C. The comparison table below summarizes channel count, supported sensor types, and connector compatibility for the readouts in this guide.

Model

Type

Channels

Measures

**5627A-9**

Precision Industrial PRT

—

\-200 °C to 300 °C

1523

Handheld

1

PRT/RTD, Thermistor, Thermocouple

1524

Handheld

2

PRT/RTD, Thermistor, Thermocouple

1502A

Benchtop

1

PRT/RTD

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Probe Model

Probe Range

Transition Junction Range

Size

Basic Accuracy

Minimum Immersion Depth

**Secondary Reference PRT (17025 accredited calibration included)**

5615-6

\-200 °C to 300 °C

\-50 °C to 200 °C

152 mm x 4.76 mm (6.0 in x 0.188 in)

± 0.013 °C at 0.010 °C

100 mm (4.0 in)

5615-9

\-200 °C to 420 °C

\-50 °C to 200 °C

229 mm x 4.76 mm (9.0 in x 0.188 in)

± 0.013 °C at 0.010 °C

100 mm (4.0 in)

5615-12

\-200 °C to 420 °C

\-50 °C to 200 °C

305 mm x 6.35 mm (12.0 in x 0.250 in)

± 0.013 °C at 0.010 °C

127 mm (5.0 in)

**Precision Industrial PRT (17025 accredited calibration included)**

5627A-6

\-200 °C to 300 °C

0 °C to 150 °C

152 mm x 4.7 mm (6.0 in x 0.187 in)

± 0.05 °C at 0 °C

100 mm (4.0 in)

5627A-9

\-200 °C to 300 °C

0 °C to 150 °C

229 mm x 4.7 mm (9.0 in x 0.187 in)

± 0.05 °C at 0 °C

100 mm (4.0 in)

5627A-12

\-200 °C to 420 °C

0 °C to 150 °C

305 mm x 6.35 mm (12.0 in x 0.250 in)

± 0.05 °C at 0 °C

127 mm (5.0 in)

**Secondary Standard PRT (17025 accredited calibration included)**

5628

\-200 °C to 661 °C

0 °C to 80 °C

305 or 381 mm x 6.35 mm (12.0 or 15.0 in x 0.250 in)

± 0.006 °C at 0 °C

127 mm (5.0 in)

**Full Immersion PRT (no calibration included; calibration optional)**

5606

\-200 °C to 160 °C

\-200 °C to 160 °C

50 mm x 3.1 mm (2.0 in x 0.125 in)

± 0.05 °C

76 mm (3.0 in)

**Thermistor Secondary Probe (NIST traceable calibration included; NVLAP accredited optional)**

5610

0 °C to 100 °C

0 °C to 100 °C

152 or 229 mm x 3.2 mm (6.0 or 9.0 in x 0.125 in)

± 0.01 °C

76 mm (3.0 in)

**Important:** The 5615-12 Secondary Reference PRT has a probe range of -200 °C to 420 °C but a transition junction range of -50 °C to 200 °C. The probe will be damaged if the handle is exposed to temperatures outside the -50 °C to 200 °C range. For applications below -50 °C, the 5606 Full Immersion PRT operates with a transition junction range matching its full -200 °C to 160 °C probe range.**Important:** Reference probes must be terminated with the correct connector for the chosen readout: type P (smart Lemo) for the 1523/1524, type D (DIN) for the 1502A/1504, and type L (gold-plated mini spade lug) for the 1529 and 1586A. A Lemo-to-Universal Thermocouple adapter (2373-LTC) is available for connection to thermocouples on the 1523/1524.**Important:** The 1524 can measure two channels at a time, but only one channel can be a thermocouple.**Important:** Probe transition-junction temperature limits must be observed independently of probe-tip temperature range. For example, the 5615-12 Secondary Reference PRT has a probe range of -200 °C to 420 °C but a transition-junction range of -50 °C to 200 °C; exceeding the transition-junction limits can damage the probe even when the tip is within its rated range.**Important:** Minimum probe immersion is approximately 15 times the probe diameter plus the sensor length. Insufficient immersion of the reference probe or unit under test will produce inaccurate measurements.**Important:** The 5606 Full Immersion PRT does not include a factory calibration as standard; calibration is available as an option. Until calibrated, the chain of traceability is broken.**Important:** Type T and Type K thermocouples are typically supplied uncalibrated by the manufacturer. Check with the distributor regarding thermocouple calibration options.**Important:** "Basic Accuracy" values listed for probes include calibration uncertainty and short-term repeatability but do not include long-term drift.Recommended pairing: connector type must match the readout. Use 5615-x-P, 5627A-x-P, 5628-12-P, 5606-50-P, and 5610-9-P probes with the 1523/1524. Use the -D termination variants with the 1502A/1504. Use the -L termination variants with the 1529 and 1586A.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 1199.0 USD
- **Vendor:** Fluke Calibration
- **Type:** Data Log
- **Tags:** Fluke, Manufacturers, Other, Other Equipment, Probe, product_manual_group_2, Product_new

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 1199.00 USD | In stock |

## Images

- 5627A-9-A Fluke Probe

---

> Source: [ValueTronics](https://valuetronics.com/products/5627a-9-a-fluke-probe-new)
> Updated: 2026-05-30