# 9144-F-P-156 Fluke Temperature Calibrator New

**Features:**

-   Each process version provides built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24V loop supply to power the transmitter
-   Accuracy ranges from ±0.01°C to ±0.07°C depending on the measured temperature
-   Accepts an ITS-90 reference PRT that contains individual calibration constants that reside in a memory chip located inside the sensor housing, so sensors may be used interchangeably
-   Second channel is user-selectable for 2-, 3-, or 4-wire RTDs, thermo-couples, or 4-20 mA transmitters
-   With 24V loop power, you are able to power and measure the trans-mitter current while sourcing and measuring temperature thus allowing for the measurement of as-found and as-left data in one self-contained calibration tool
-   Onboard non-volatile memory for documentation of up to 20 tests
-   Each test can be given a unique alphanumeric ID and will record block temperature, reference temperature, UUT values, error, date, and time which can be viewed via the front panel or exported using Interface-it software (included)
-   Stability indicator visually and audibly tells you the Field Metrology Well is stable to the selectable criteria
-   Preprogrammed calibration routines are stored in memory for easy recall, and all inputs are easily accessible via the front panel of the instrument

**Applications:**

-   Ideal for the following applications: Laboratory, Education, R&D, Calibration Labs, Field Calibration, Industrial Process

**Specifications:**

-   Ref. Characterizations - 4-wire, ITS-90, CVD, IEC-751, Resistance
-   Ref. Therm. Accuracy ±0.010°C @ -25°C, ±0.015°C @ 0°C, ±0.020°C @ 50°C, ±0.025°C @ 150°C, ±0.030°C @ 200°C, ±0.40°C @ 350°C, ±0.050°C @ 420°C, ±0.70°C @ 660°C
-   Ref. Resistance Range - 0 to 400
-   Ref. Resistance Accuracy - 0-25: ±0.002, 25-400: ±60 ppm
-   RTD Characterizations - 4-wire RTD (2-, 3-wire RTD w/ jumpers only), PT-100 (385), (JIS), (3926), Ni-12, Resistance
-   RTD Therm. Accuracy Ni-120: ±0.1°C @ 0°C; Pt-100 (385): ±0.02°C @ 0°C; Pt-100 (3926): ±0.02°C @ 0°C;Pt-100 (JIS): ±0.02°C @ 0°C
-   RTD Resistance Range - 0 to 400
-   RTD Resistance Accuracy - 0-25: ±0.002, 25-400: ±80 ppm
-   T/C mV Range -10 to 75 mV
-   Voltage Accuracy -10 to 50 mV: ±0.01 mV; 50 to 75 mV: ±250 ppm
-   Cold Junction Comp. Accuracy - ±0.5°C (ambient of 13 to 33°C)

This unit has a built-in two channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20mA transmitters which includes the 24V loop supply to power the transmitter.

## Details

- **Price:** 9177.0 USD
- **Vendor:** Fluke
- **Type:** Temp Cali
- **Tags:** Calibrator, Fluke, Manufacturers, Other Equipment, product_manual_group_2, Product_new, Temperature

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 9177.00 USD | In stock |

## Images

- 9144-F-P-156 Fluke Temperature Calibrator

---

> Source: [ValueTronics](https://valuetronics.com/products/9144-f-p-156-fluke-temperature-calibrator-new)