# 87 Fluke Multimeter Used

## Fluke 83V Digital Multimeter — Key Features at a Glance

-   **Average-responding rms indicating** — calibrated for sine-wave AC measurement on conventional electrical systems
-   **Basic DC voltage accuracy:** ± (0.1 % + 1) across all five DC ranges (6 V / 60 V / 600 V / 1000 V)
-   **6,000-count digital display** updating four times per second, with a 33-segment analog bar graph updating 40 times per second
-   **Measures voltage, current, resistance, capacitance, frequency, conductance, and continuity** — the complete field-DMM function set in one tool
-   **Safety:** 1000 V Overvoltage Category III / 600 V Overvoltage Category IV per ANSI/ISA, CSA, IEC 664, UL3111-1, TÜV EN61010-1
-   **Built to last:** 400-hour battery life with alkaline, 1-meter drop tested per IEC 61010-1:2001, MIL-PRF-28800 Class 2 vibration
-   **Same rugged Fluke 80V-series industrial form factor** as the 87V — the housing, holster, and flex-stand are common across the series
-   **Fluke 80V-series value tier** — for sine-wave AC work where true-rms response is not required (for non-linear loads, see the Fluke 87V)

## Overview

In summary, the Fluke 83V is the average-responding rms indicating digital multimeter in the Fluke 80V industrial series — a rugged 1000 V CAT III / 600 V CAT IV field DMM with ± (0.1 % + 1) basic DC voltage accuracy, 6,000-count digital display with 33-segment analog bar graph updating 40 times per second, and the full Fluke 80V-series industrial form factor including 1-meter drop tolerance per IEC 61010-1:2001 and MIL-PRF-28800 Class 2 vibration testing. The 83V is calibrated for sine-wave AC measurement; for true-rms measurement on distorted waveforms from variable-frequency drives, switching supplies, or non-linear loads, the Fluke 87V is the correct instrument.

## What is the Fluke 83V used for?

The Fluke 83V is used by commercial electricians, building maintenance technicians, HVAC service technicians, and electrical apprentices for general-purpose voltage, current, resistance, capacitance, and frequency measurement on sine-wave AC and DC systems. It is the value entry point into the Fluke 80V industrial series — the same rugged 1000 V CAT III / 600 V CAT IV form factor as the 87V at a lower price, calibrated for sine-wave AC accuracy rather than the true-rms response required by non-linear loads such as variable-frequency drives.

### How does the Fluke 83V compare to the Fluke 87V?

The 83V is an average-responding rms indicating meter; the 87V is true-rms responding — the 87V correctly measures distorted, non-sinusoidal waveforms common in industrial environments with variable-frequency drives and switching power supplies. The 87V also adds a selectable 4½-digit (19,999 count) high-resolution mode, K-type thermocouple temperature measurement (−200 °C to +1090 °C), a switchable low pass filter for clean VFD readings, 250 µs peak MIN MAX capture, and tighter accuracy across DC, resistance, and current ranges. Both meters share the same rugged industrial form factor and 1000 V CAT III / 600 V CAT IV safety rating.

## Recommended Accessories & Options

Every professional DMM owner replaces leads every 1–2 years, more frequently in harsh environments. The accessory list below covers the standard pairings for the Fluke 80V-series industrial meters; none of these are required to order the meter — they are listed as options to pair at quote time.

-   **Fluke TL75 Standard Test Leads:** Hard-point banana-plug test leads, rated CAT III 1000 V / CAT IV 600 V. The standard replacement lead for any Fluke DMM.
-   **Fluke TL224 SureGrip Insulated Test Leads:** Premium silicone-insulated leads with improved grip handles. CAT III 1000 V rating. The upgrade pick for field electricians who use their leads heavily.
-   **Fluke TPAK Magnetic Hanger:** Magnetic strap for hands-free meter mounting on enclosure doors and steel surfaces — frees both hands for probing.
-   **Fluke 80BK-A Integrated DMM Temperature Probe:** K-type thermocouple bead probe (87V only — required to use the 87V's onboard −200 °C to +1090 °C temperature measurement function).
-   **Fluke C35 Soft Carrying Case:** Padded zippered carrying case sized for the 80V-series meter with holster, leads, and accessories.
-   **Replacement holster and flex-stand:** The same yellow holster with integrated flex-stand referenced in the dimensions and weight specifications of the datasheet.
-   **Replacement fuses:** 44/100 A, 1000 V FAST Fuse for the mA/µA inputs; 11 A, 1000 V FAST Fuse for the A input — match the datasheet's fuse-protection specifications exactly.
-   **Replacement 9 V battery:** 9 V zinc, NEDA 1604 or 6F22 or 006P; 400 hours typical with alkaline (backlight off).
-   **Annual calibration service:** The published accuracy specification covers a one-year period after calibration; recommended for documentation-driven compliance environments.

## Fluke 83V — Full Detail & Specifications

The digital multimeter is the most widely deployed test instrument in the world. It measures the three fundamental electrical quantities — voltage (AC and DC), current, and resistance — and it is carried daily by every electrician, every field service technician, every HVAC/R professional, every industrial maintenance engineer, and every facilities crew that touches an electrical system. When a professional says "I need a meter," the conversation almost always lands on the same name: Fluke. The brand has earned that position through decades of rugged, accurate, category-defining field instruments — a trust anchor that sets the standard against which every competing DMM is measured.

The Fluke 83V sits in the Fluke 80V industrial series — the line that established the modern field DMM form factor. The 83V is the value sibling in the Fluke 80V industrial series — the same rugged industrial form factor and CAT III 1000V / CAT IV 600V safety rating as the 87V, at a lower price point for technicians whose work does not require true-rms measurement of distorted waveforms or onboard thermocouple temperature. Both the 83V and the 87V share the same industrial-grade housing, the same 1000 V rms terminal rating, the same 1000 V Overvoltage Category III / 600 V Overvoltage Category IV safety compliance, and the same 1-meter drop tolerance per IEC 61010-1:2001 with the holster installed. The differences between the two models live in measurement architecture, accuracy class, and feature set — choose the model that matches the waveforms in your work and the documentation standard of your trade.

### Average-Responding rms Indicating — The Sine-Wave Workhorse

The Fluke 83V is an average-responding rms indicating meter. For pure or near-pure sinusoidal AC signals — utility-frequency 50/60 Hz line voltage, conventional resistive loads, lighting circuits with incandescent or magnetic-ballast fixtures, and most general electrical work — the 83V reads accurately and provides the same rugged Fluke industrial form factor as the true-rms 87V at a more accessible price point. Where the 83V is not the correct tool is on non-sinusoidal waveforms. Variable-frequency drives, switching power supplies, LED lighting, electronic ballasts, and other non-linear loads produce distorted current and voltage waveforms that an average-responding meter will under-read; for those measurements, the true-rms Fluke 87V is the correct instrument.

### CAT III 1000 V / CAT IV 600 V — The Safety Rating That Matters

Per the datasheet's general specifications: _"Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC 664 to 600 V Overvoltage Category IV. UL listed to UL3111-1. Licensed by TÜV to EN61010-1."_ Translated into practical terms: CAT III at 1000 V is the firm minimum rating for work on distribution wiring, panels, and bus bars — anything between the service entrance and the branch circuits. CAT IV at 600 V is the rating required for work at the utility service entrance, on outdoor conductors, and on the supply side of the main disconnect. A meter rated only CAT II should never be used on permanently installed distribution wiring; the Fluke 83V meets both the CAT III and CAT IV thresholds that professional electrical work demands.

### Accuracy Specification — How to Read It

Per the datasheet: _"Accuracy is given as ±(\[% of reading\] + \[number of least significant digits\]) at 18 °C to 28 °C, with relative humidity up to 90 %, for a period of one year after calibration."_ _"AC conversions are ac-coupled and valid from 3 % to 100 % of range. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add −(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3."_

### Fluke 83V AC Voltage Function Specifications (average responding rms indicating)

Accuracy is given as ±(\[% of reading\] + \[number of least significant digits\]) at 18 °C to 28 °C, with relative humidity up to 90 %, for a period of one year after calibration. AC conversions are ac-coupled and valid from 3 % to 100 % of range. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add −(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.

Function

Range

Resolution

Accuracy

50 Hz – 60 Hz

30 Hz – 1 kHz

1 kHz – 5 kHz

V~1

600.0 mV

0.1 mV

± (0.5 % + 4)

± (1.0 % + 4)

± (2.0 % + 4)

6.000 V

0.001 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)

60.00 V

0.01 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)

600.0 V

0.1 V

± (0.5 % + 2)

± (1.0 % + 4)

± (2.0 % + 4)2

1000 V

1 V

± (0.5 % + 2)

± (1.0 % + 4)

unspecified

1.  Below a reading of 200 counts, add 10 counts.
2.  Frequency range: 1 kHz to 2.5 kHz.

### DC Voltage, Resistance, and Conductance Function Specifications

Function

Range

Resolution

Fluke 83V Accuracy

V—

6.000 V

0.001 V

± (0.1 % + 1)

60.00 V

0.01 V

± (0.1 % + 1)

600.0 V

0.1 V

± (0.1 % + 1)

1000 V

1 V

± (0.1 % + 1)

mV—

600.0 mV

0.1 mV

± (0.3 % + 1)

Ω

600.0 Ω

0.1 Ω

± (0.4 % + 2)1

6.000 kΩ

0.001 kΩ

± (0.4 % + 1)

60.00 kΩ

0.01 kΩ

± (0.4 % + 1)

600.0 kΩ

0.1 kΩ

± (0.7 % + 1)

6.000 MΩ

0.001 MΩ

± (0.7 % + 1)

50.00 MΩ

0.01 MΩ

± (1.0 % + 3)2

nS

60.00 nS

0.01 nS

± (1.0 % + 10)1

1.  When using the REL Δ function to compensate for offsets.
2.  Add 0.5 % of reading when measuring above 30 MΩ in the 50 MΩ range and 20 counts below 33 nS in the 60 nS range.

### Current Function Specifications

Function

Range

Resolution

Model 83 Accuracy

Burden Voltage (typical)

A~  
(45 Hz to 2 kHz)

60.00 mA

0.01 mA

± (1.2 % + 2)5

1.8 mV/mA

400.0 mA6

0.1 mA

± (1.2 % + 2)5

1.8 mV/mA

6.000 A

0.001 A

± (1.2 % + 2)5

0.03 V/A

10.00 A4

0.01 A

± (1.2 % + 2)5

0.03 V/A

A—

60.00 mA

0.01 mA

± (0.4 % + 4)

1.8 mV/mA

400.0 mA6

0.1 mA

± (0.4 % + 2)

1.8 mV/mA

6.000 A

0.001 A

± (0.4 % + 4)

0.03 V/A

10.00 A4

0.01 A

± (0.4 % + 2)

0.03 V/A

µA~  
(45 Hz to 2 kHz)

600.0 µA

0.1 µA

± (1.2 % + 2)5

100 µV/µA

6000 µA

1 µA

± (1.2 % + 2)5

100 µV/µA

µA—

600.0 µA

0.1 µA

± (0.4 % + 4)

100 µV/µA

6000 µA

1 µA

± (0.4 % + 2)

100 µV/µA

1.  AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sine wave input.
2.  AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range.
3.  Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error in later measurements.
4.  10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A unspecified.
5.  Below a reading of 200 counts, add 10 counts.
6.  400 mA continuous; 600 mA for 18 hours maximum.

### Capacitance and Diode Function Specifications

Function

Range

Resolution

Accuracy

Capacitance

10.00 nF

0.01 nF

± (1 % + 2)1

100.0 nF

0.1 nF

± (1 % + 2)1

1.000 µF

0.001 µF

± (1 % + 2)

10.00 µF

0.01 µF

± (1 % + 2)

100.0 µF

0.1 µF

± (1 % + 2)

9999 µF

1 µF

± (1 % + 2)

Diode

3.000 V

0.001 V

± (2 % + 1)

1.  With a film capacitor or better, using Relative mode to zero residual.

### MIN MAX Recording Specifications

Model

Nominal Response

Accuracy

83V

100 ms to 80 %

Specified accuracy ± 12 counts for changes > 200 ms in duration (± 40 counts in ac with beeper on)

### Electrical Characteristics of the Terminals

Function

Overload Protection1

Input Impedance (nominal)

Common Mode Rejection Ratio (1 kΩ unbalance)

Normal Mode Rejection

V dc

1000 V rms

10 MΩ < 100 pF

\> 120 dB at dc, 50 Hz or 60 Hz

\> 60 dB at 50 Hz or 60 Hz

mV dc

1000 V rms

10 MΩ < 100 pF

\> 120 dB at dc, 50 Hz or 60 Hz

\> 60 dB at 50 Hz or 60 Hz

V ac (ac-coupled)

1000 V rms

10 MΩ < 100 pF

\> 60 dB, dc to 60 Hz

—

Function

Open Circuit Test Voltage

Full Scale Voltage

Typical Short Circuit Current

To 6.0 MΩ

50 MΩ or 60 nS

600 Ω

6 kΩ

60 kΩ

600 kΩ

6 MΩ

50 MΩ

Ω

1000 V rms

< 7.3 V dc

< 4.1 V dc

1 mA

100 µA

10 µA

1 µA

1 µA

0.5 µA

Diode

1000 V rms

< 3.9 V dc

3.000 V dc

0.6 mA typical

1.  106 V Hz maximum.

### Frequency Counter Specifications

Function

Range

Resolution

Accuracy

Frequency  
(0.5 Hz to 200 kHz, pulse width > 2 µs)

199.99

0.01 Hz

± (0.005 % + 1)

1999.9

0.1 Hz

± (0.005 % + 1)

19.999 kHz

0.001 kHz

± (0.005 % + 1)

199.99 kHz

0.01 kHz

± (0.005 % + 1)

\> 200 kHz

0.1 kHz

unspecified

#### Frequency Counter Sensitivity and Trigger Levels

Input Range1

Minimum Sensitivity (RMS Sine wave)

Approximate Trigger Level (DC Voltage Function)

5 Hz – 20 kHz

0.5 Hz – 200 kHz

600 mV dc

70 mV (to 400 Hz)

70 mV (to 400 Hz)

40 mV

600 mV ac

150 mV

150 mV

—

6 V

0.3 V

0.7 V

1.7 V

60 V

3 V

7 V (≤ 140 kHz)

4 V

600 V

30 V

70 V (≤ 14.0 kHz)

40 V

1000 V

100 V

700 V (≤ 1.4 kHz)

100 V

**Duty Cycle Range:** 0.0 to 99.9 % — **Accuracy:** Within ± (0.2 % per kHz + 0.1 %) for risetimes < 1 µs

1.  Maximum input for specified accuracy = 10X Range or 1000 V.

### General Specifications

**Maximum voltage between any terminal and earth ground**

1000 V rms

**Fuse protection for mA or µA inputs**

44/100 A, 1000 V FAST Fuse

**Fuse protection for A input**

11 A, 1000 V FAST Fuse

**Display — Digital**

6000 counts updates 4/sec; (Model 87V also has 19,999 counts in high-resolution mode)

**Display — Analog**

33 segments, updates 40/sec

**Display — Frequency**

19,999 counts, updates 3/sec at > 10 Hz

**Temperature — Operating**

−20 °C to +55 °C

**Temperature — Storage**

−40 °C to +60 °C

**Altitude — Operating**

2000 m

**Altitude — Storage**

10,000 m

**Temperature coefficient**

0.05 × (specified accuracy)/ °C (< 18 °C or > 28 °C)

**Electromagnetic compatibility**

In an RF field of 3 V/m total accuracy = specified accuracy

**Relative humidity**

0 % to 90 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 55 °C)

**Battery type**

9 V zinc, NEDA 1604 or 6F22 or 006P

**Battery life**

400 hours typical with alkaline (with backlight off)

**Vibration**

Per MIL-PRF-28800 for a Class 2 instrument

**Shock**

1 Meter drop per IEC 61010-1:2001

**Size (H × W × L)**

1.25 in × 3.41 in × 7.35 in (3.1 cm × 8.6 cm × 18.6 cm)

**Size with holster and flex-stand**

2.06 in × 3.86 in × 7.93 in (5.2 cm × 9.8 cm × 20.1 cm)

**Weight**

12.5 oz (355 g)

**Weight with holster and flex-stand**

22.0 oz (624 g)

**Safety**

Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC 664 to 600 V Overvoltage Category IV. UL listed to UL3111-1. Licensed by TÜV to EN61010-1.

### Factual Grounding — Specifications and Industry Context

The Fluke 83V delivers ± (0.1 % + 1) basic DC voltage accuracy across all five DC voltage ranges (6.000 V, 60.00 V, 600.0 V, 1000 V), ± (0.4 % + 1) basic resistance accuracy on the mid-decade ranges, ± (0.4 % + 2) basic DC current accuracy on the 400 mA and 10 A ranges, and 0.005 % + 1 frequency counter accuracy from 0.5 Hz to 200 kHz. Battery life is 400 hours typical with an alkaline 9 V zinc battery (NEDA 1604 / 6F22 / 006P), with backlight off. Operating temperature is −20 °C to +55 °C; storage temperature is −40 °C to +60 °C; operating altitude is 2000 m. The instrument is vibration-tested per MIL-PRF-28800 for a Class 2 instrument and shock-tested to a 1 Meter drop per IEC 61010-1:2001. Net weight is 12.5 oz (355 g); with holster and flex-stand, 22.0 oz (624 g). Total accuracy in an RF field of 3 V/m equals the specified accuracy — a robust electromagnetic compatibility specification for noisy industrial environments.

### Entity Authority — Fluke Corporation and the Field DMM Standard

Fluke Corporation was founded in 1948 in Seattle by John Fluke Sr. and is now part of Fortive Corporation (previously Danaher). Fluke is the dominant global brand in field test instruments and the unambiguous reference standard for handheld digital multimeters in industrial, electrical contracting, and facilities maintenance markets. The Fluke 80V industrial series carries safety compliance to ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 (1000 V Overvoltage Category III), IEC 664 (600 V Overvoltage Category IV), UL3111-1 listing, and TÜV licensing to EN61010-1. Vibration testing follows MIL-PRF-28800 for a Class 2 instrument; shock testing follows IEC 61010-1:2001 for a 1 Meter drop. The CAT III and CAT IV ratings define the IEC transient overvoltage protection level required for work on distribution wiring, panels, bus bars (CAT III), and the utility service entrance (CAT IV).

### How does the Fluke 83V compare to industry baseline digital multimeters?

-   **Safety rating:** 1000 V Overvoltage Category III and 600 V Overvoltage Category IV — the firm minimum for industrial panel and utility service-entrance work, exceeding the CAT II rating typical of consumer-grade DMMs.
-   **Basic DC voltage accuracy:** ± (0.1 % + 1) across all five DC ranges — professional field accuracy consistent with the Fluke industrial standard for general electrical work.
-   **AC measurement architecture:** Average-responding rms indicating — calibrated for sine-wave AC signals; correct for general electrical work but not for non-linear loads.
-   **Display resolution:** 6,000 counts digital with a 33-segment analog bar graph updating 40 times per second — exceeding the 4,000-count basic and 6,000-count entry-professional thresholds typical of the field DMM category.
-   **Fuse protection:** 44/100 A, 1000 V FAST Fuse on mA/µA inputs and 11 A, 1000 V FAST Fuse on the A input — sized for the high-energy fault currents present at industrial panels rather than the lower-energy faults seen at branch-circuit outlets.
-   **Drop protection:** 1 Meter drop per IEC 61010-1:2001 with the holster installed — appropriate for daily tool-bag use in a field service environment.
-   **Battery life:** 400 hours typical with alkaline — multi-week continuous-use life that minimizes downtime in the field.

## Details

- **Price:** 0.0 USD - 91.0 USD
- **Vendor:** Fluke
- **Type:** Multi Met
- **Tags:** Fluke, Manufacturers, Meters & DAQ, Multimeter, product_manual_group_2, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 65.00 USD | In stock |
| NIST Traceable with Full Data | 80.00 USD | In stock |
| ISO IEC 17025 Accredited | 91.00 USD | In stock |