# RIGOL DSA710 1 GHz Spectrum Analyzer

# RIGOL DSA710 1 GHz Spectrum Analyzer

## Key Features At A Glance

-   All-Digital IF technology improves bandwidth precision and selectivity while reducing displayed average noise level
-   Frequency coverage from 100 kHz up to 1 GHz across the series
-   Displayed average noise level down to -130 dBm (typ.) with preamplifier on
-   SSB phase noise <-80 dBc/Hz @ 10 kHz offset
-   100 Hz minimum resolution bandwidth to distinguish closely spaced signals
-   LAN (LXI) and USB host/device standard; GPIB (IEEE 488.2) available via USB-GPIB option for automated test systems
-   8-inch WVGA (800x480) display in a compact, lightweight chassis
-   Optional advanced measurement kit, EMI filter & quasi-peak detector kit, and Ultra Spectrum PC software

The DSA700 Series is an all-digital IF spectrum analyzer covering a frequency range from 100 kHz up to 1 GHz. Built around an all-digital intermediate-frequency architecture, it presents spectral measurements on an 8-inch WVGA (800x480) TFT LCD display in a compact, lightweight chassis measuring 361.6 mm x 178.8 mm x 128 mm and weighing 4.25 kg.

A spectrum analyzer measures and displays signal amplitude across a range of frequencies, letting an engineer characterize spurious content, harmonics, and channel power in the frequency domain. With the optional Advanced Measurement Kit, the series performs harmonic distortion, third-order intercept (TOI), emission bandwidth, channel power, time-domain power, adjacent channel power, occupied bandwidth, carrier-to-noise ratio, and pass/fail measurements. Zero-span operation supports AM signal demodulation, while the peak table function reads out spectrum peak values.

ValueTronics International has supplied new and used test and measurement equipment since 1992 and stocks its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois.

## Compare Other Models in This Series

The DSA700 Series comprises two models that share the same all-digital IF architecture and differ principally in upper frequency limit: the DSA705 covers 100 kHz to 500 MHz, while the DSA710 extends coverage to 1 GHz. Both provide the same 100 Hz minimum resolution bandwidth, 1 Hz frequency resolution, and 8-inch WVGA display.

Across the series, amplitude and noise performance is shared, including displayed average noise level, SSB phase noise, and level measurement uncertainty, so model selection is driven primarily by the required frequency span.

Each new model in the DSA700 Series below is listed on its own dedicated product page with condition-matched pricing. Select the model whose frequency coverage matches your application, then choose the new unit that fits your requirements.

The defining difference between the two models is upper frequency limit: the DSA705 reaches 500 MHz and the DSA710 reaches 1 GHz, as shown in the comparison table. SSB phase noise is specified at fc = 500 MHz for the DSA705 and fc = 1 GHz for the DSA710, with both rated <-80 dBc/Hz at a 10 kHz offset.

Beyond frequency range, the two models share specifications such as displayed average noise level, frequency response, input attenuation range (0 dB to 30 dB in 1 dB steps), reference-level range, and a standard weight of 4.25 kg. Options including the EMI-DSA800 kit, the AMK-DSA800 advanced measurement kit, Ultra Spectrum PC software, and SSC-DSA signal seamless capture apply across the series.

Model

Frequency Range

DANL (Typ., PA On)

SSB Phase Noise @10 kHz

**DSA710**

100 kHz to 1 GHz

\-130 dBm (typ.)

<-80 dBc/Hz

DSA705

100 kHz to 500 MHz

\-130 dBm (typ.)

<-80 dBc/Hz

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Quick guide (hard copy)
-   Power cable

## Product Core & Specifications

Specification

Value

Frequency

Frequency range

100 kHz to 1 GHz

Frequency resolution

1 Hz

Internal Reference Frequency

Reference frequency

10 MHz

Accuracy

±\[ (time since last calibration × aging rate) + temperature stability + calibration accuracy\]

Initial calibration accuracy

<1 ppm

Temperature stability (0 °C to 50 °C, reference to 25 °C)

<2 ppm

Aging rate

<2 ppm/year

Frequency Readout Accuracy

Marker resolution

span/ (number of sweep points - 1)

Marker uncertainty

±(frequency indication × reference frequency accuracy + 1% × span + 10% × resolution bandwidth + marker resolution)

Frequency Counter

Resolution

1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz

Uncertainty

±(frequency indication × reference frequency accuracy + counter resolution)

Frequency Span

Range

0 Hz, 100 Hz to maximum frequency of instrument

Uncertainty

±span/ (number of sweep points - 1)

SSB Phase Noise (20 °C to 30 °C, fc = 1 GHz)

Carrier offset 10 kHz

<-80 dBc/Hz

Carrier offset 100 kHz

<-100 dBc/Hz (typ.)

Residual FM (20 °C to 30 °C, RBW = VBW = 1 kHz)

Residual FM

<50 Hz (nom.)

Bandwidths (Set "Auto SWT" to "Accy")

Resolution bandwidth (-3 dB)

100 Hz to 1 MHz, in 1-3-10 sequence

RBW uncertainty

<5% (nom.)

Video bandwidth (-3 dB)

1 Hz to 3 MHz, in 1-3-10 sequence

Resolution bandwidth (-6 dB) (EMI-DSA800 option)

200 Hz, 9 kHz, 120 kHz

Measurement Range

Range (fc ≥ 10 MHz)

DANL to +20 dBm

Maximum Input Level

DC voltage

50 V

CW RF power (attenuation = 30 dB)

+20 dBm (100 mW)

Max. damage level

+30 dBm (1 W)

Displayed Average Noise Level (DANL) (attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50, 20 °C to 30 °C, input impendence = 50 Ω)

PA off, 100 kHz to 1 MHz

<-90 dBm, <-110 dBm (typ.)

PA off, 1 MHz to 500 MHz

<-100 dBm, <-110 dBm (typ.)

PA off, 500 MHz to 1 GHz

<-100 dBm, <-110 dBm (typ.)

PA on, 100 kHz to 1 MHz

<-110 dBm, <-130 dBm (typ.)

PA on, 1 MHz to 500 MHz

<-120 dBm, <-130 dBm (typ.)

PA on, 500 MHz to 1 GHz

<-120 dBm, <-130 dBm (typ.)

Level Display

Logarithmic level axis

1 dB to 200 dB

Linear level axis

0 to reference level

Number of display points

601

Number of traces

3 + math trace

Trace detectors

normal, positive-peak, negative-peak, sample, RMS, voltage average; quasi-peak (with EMI-DSA800 option)

Trace functions

clear write, max hold, min hold, average, view, blank

Units of level axis

dBm, dBmV, dBµV, nV, µV, mV, V, nW, µW, mW, W

Frequency Response

PA off (fc ≥ 100 kHz, attenuation = 10 dB, relative to 50 MHz, 20 °C to 30 °C), 100 kHz to 500 MHz

<0.7 dB

PA off, 500 MHz to 1 GHz

<0.7 dB

PA on (fc ≥ 1 MHz, attenuation = 10 dB, relative to 50 MHz, 20 °C to 30 °C), 100 kHz to 500 MHz

<1.0 dB

PA on, 500 MHz to 1 GHz

<1.0 dB

Input Attenuation Switching Uncertainty

Setting range

0 dB to 30 dB, in 1 dB step

Switching uncertainty (fc = 50 MHz, relative to 10 dB, 20 °C to 30 °C)

<0.5 dB

Absolute Amplitude Uncertainty

Uncertainty (fc, 20 °C to 30 °C)

<0.4 dB

RBW Switching Uncertainty

Uncertainty (relative to 1 kHz RBW)

<0.1 dB

Reference Level

Range

\-100 dBm to +20 dBm, in 1 dB step

Resolution, log scale

0.01 dB

Resolution, linear scale

4 digits

Preamplifier Gain (standard)

Gain, 100 kHz to 500 MHz

20 dB (nom.)

Gain, 500 MHz to 1 GHz

20 dB (nom.)

Level Measurement Uncertainty

Level measurement uncertainty (attenuation = 10 dB, -50 dBm < input level ≤ 0 dBm, fc > 10 MHz, 20 °C to 30 °C)

<1.5 dB (nom.)

RF Input VSWR (attenuation ≥ 10 dB)

VSWR, 300 kHz to 500 MHz

<1.5 (nom.)

VSWR, 500 MHz to 1 GHz

<1.5 (nom.)

Distortion

Second harmonic intercept (SHI) (fc ≥ 50 MHz, input signal level = -20 dBm, attenuation = 10 dB)

+40 dBm

Third-order intercept (TOI) (fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 10 dB)

+10 dBm

1 dB compression of input mixer (P1dB) (fc ≥ 50 MHz, attenuation = 0 dB)

\>0 dBm

Spurious Response (input terminated 50 Ω, attenuation = 0 dB, 20 °C to 30 °C)

Spurious response, inherent

<-88 dBm (typ.)

Intermediate frequency

<-60 dBc

System related sidebands

<-60 dBc

Input related spurious (mixer level = -30 dBm)

<-60 dBc

Trigger

Trigger source

free run, video, external

External trigger level

5 V TTL level

Signal Seamless Capture (SSC)

Measurement bandwidth

1.5 MHz

Sweep

Sweep time, span ≥ 100 Hz

10 ms to 1000 s

Sweep time, zero span

20 µs to 1000 s

> **Important:** Maximum input: 50 V DC; +20 dBm (100 mW) CW RF power at 30 dB attenuation; +30 dBm (1 W) maximum damage level.

> **Important:** When fc ≥ 10 MHz, input level > +25 dBm and the preamplifier (PA) is Off, the input protection switch will be on.

> **Important:** Standard delivery includes only the quick guide (hard copy) and power cable. RF cables, adaptors, antennas, attenuators, the RF demo (TX1000/RX1000) training kit, and the near field probe are optional accessories ordered separately.

> **Important:** Advanced measurement functions, the EMI Filter & Quasi-Peak Detector Kit (EMI-DSA800), and PC software (Ultra Spectrum) are optional and not included with the standard instrument.

> **Important:** Recommended calibration interval is 18 months.

> Recommended pairing: the optional TX1000 (transmitter) and RX1000 (receiver) RF demo kits for hands-on RF transmit/receive training and demonstrations with the analyzer.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 799.0 USD
- **Vendor:** RIGOL
- **Type:** Spec An
- **Tags:** Analyzers, DSA710, Factory New, Inc., Manufacturers, Model 710, product_manual_group_1, Product_new, Rigol, RIGOL Technologies, Spectrum, Spectrum Analyzer

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 799.00 USD | In stock |

## Images

- DSA710 Rigol Spectrum Analyzer

---

> Source: [ValueTronics](https://valuetronics.com/products/dsa710-rigol-spectrum-analyzer-new)