# Ando AQ6317B Optical Spectrum Analyzer (Pre-Owned)

# Ando AQ6317B Optical Spectrum Analyzer (Pre-Owned)

## Key Features At A Glance

-   Wavelength accuracy of ±0.02 nm (1520–1580 nm) after calibration with the built-in reference light source
-   Dynamic range of 70 dB at peak ±0.4 nm and 60 dB at peak ±0.2 nm for 50 GHz-spacing WDM signals
-   Maximum wavelength resolution of 0.015 nm (1520–1620 nm, 0.01 nm resolution setting)
-   High sensitivity measuring light down to −90 dBm across 1200–1650 nm
-   Direct high-power measurement to +20 dBm (100 mW) without an optical attenuator
-   Polarization dependency suppressed to ±0.05 dB for accurate optical-amplifier gain measurement
-   GP-IB (2 ports) plus TLS control and trigger interfaces for automated measurement systems
-   Three independent trace memories with a 9.4-inch color LCD

The Ando AQ6317B is a high-accuracy, high-resolution optical spectrum analyzer built for evaluating dense wavelength-division-multiplexing (D-WDM) systems and components. It addresses a wide range of applications including light source evaluation, measurement of loss-wavelength characteristics in optical devices, and waveform analysis of WDM systems. The instrument achieves high wavelength accuracy and wavelength linearity in the C-band and L-band, suiting it to the evaluation of optical devices for WDM.

An optical spectrum analyzer resolves an optical signal into its component wavelengths and displays power as a function of wavelength, letting an engineer characterize the spectral content of light sources and optical components. The AQ6317B applies this to specific WDM tasks: optical multiplexer (MUX/DEMUX) evaluation, yielding insertion loss, passband central wavelength, and linearity; and optical fiber amplifier (EDFA) evaluation, where the ASE interpolation method measures gain and noise figure. It also provides analysis for LD, LED, and FBG devices.

ValueTronics stocks test and measurement equipment in its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois. Because we hold inventory ourselves, every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

## Brand Heritage

Ando Electric Co., Ltd. was acquired by Yokogawa Electric Corporation in 2002, and its optical test and measurement product line subsequently continued under the Yokogawa name.

## Product Core & Specifications

Specification

Value

Optical Specifications

Applicable fibers

SMF, GI (50/125 µm)

Measurement wavelength range

600 to 1750 nm

Wavelength accuracy

±0.02 nm (1520 to 1580 nm, after calibration with build-in reference light source); ±0.04 nm (1580 to 1620 nm, after calibration with build-in reference light source); ±0.5 nm (600 to 1750 nm)

Wavelength linearity

±0.01 nm (1520 to 1580 nm); ±0.02 nm (1580 to 1620 nm)

Wavelength repeatability

±0.005 nm (1 min)

Wavelength resolution

Max. resolution: 0.015 nm or better (1520 to 1620 nm, resolution setting: 0.01 nm); Resolution setting: 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm

Resolution accuracy

±5 % (1300 to 1650 nm, resolution: 0.05 nm or more, resolution correction: ON)

Measurement level range

\-90 to +20 dBm (1200 to 1650 nm, sensitivity: HIGH3); -80 to +20 dBm (1000 to 1200 nm, sensitivity: HIGH3); -60 to +20 dBm (600 to 1000 nm, sensitivity: HIGH3)

Level accuracy

±0.3 dB (1310/1550 nm, input: -30 dBm, sensitivity: HIGH1-3)

Level linearity

±0.05 dB (input: +10 to -50 dBm, sensitivity: HIGH1-3)

Level flatness

±0.1 dB (1520 to 1580 nm), ±0.2 dB (1580 to 1620 nm)

Polarization dependency

±0.05 dB (1550/1600 nm), ±0.05 dB typ. (1310 nm)

Dynamic range

60 dB (1523 nm, peak ±0.2 nm, resolution: 0.01 nm); 70 dB (1523 nm, peak ±0.4 nm, resolution: 0.01 nm); 45 dB (1523 nm, peak ±0.2 nm, resolution: 0.1 nm)

Sweep time

Approx. 500 ms (Span: 100 nm or less, sensitivity: NORM, HOLD, ave.: 1, 501 samples, resolution correction: OFF); Approx. 0.5 min (Span: 100 nm or less, sensitivity: HIGH2, ave.: 1, 501 samples, No signal)

Functions

Automatic measurement

Program function (20 program, 200 steps), Long-term measurement function

Setting of measuring conditions

Span setting: 0 to 1200 nm; Measuring sensitivity setting: NORMAL HOLD/AUTO, MID, HIGH1/2/3; Number of averaging setting: 1 to 1000 times; Sample number setting: 11 to 20001, AUTO; Automatic setting function of measuring conditions; Sweep-between-marker function; 0 nm sweep function; Pulse light measurement function; Air/vacuum wavelength measurement function; TLS synchronized measurement function

Trace display

Level scale setting: 0.1 to 10 dB/div, linear; Simultaneous display of 3 independent traces; Max./Min. hold display; Roll averaging display; Calculation-between-traces display; Normalized display; Curve-fit display; 3D display; Split display; Power density display, % display, dB/km display; Frequency display of horizontal axis scale

Data analysis

WDM waveform analysis (Wavelength/Level/SNR list display), Optical fiber amplifier analysis (GAIN/NF, Single/Multi channel), PMD analysis, Optical filter analysis, DFB-LD analysis, FP-LD analysis, LED analysis, SMSR analysis, Peak search, bottom search, spectral width search, notch width search; Delta marker (max. 200), line marker (analysis range specification); Graph display of long-term measurement result

Others

Self-wavelength calibration function (using built-in reference light source); Wavelength/Level compensation function, label function, help function

Memory

Build-in FDD

3.5-inch 2HD

Internal memory

32 traces, 20 programs

File format

Trace file, program file, measuring condition file, Text file (trace, analysis data, etc.), Graphics file (BMP, TIFF)

Data output

Printer

Built-in high speed printer

Interface

Remote control

GP-IB (2 ports), TLS control interfaces (TTL)

Others

Sweep trigger input (TTL), Sample enable input (TTL), Sample trigger input (TTL), Analog output (0 to 5 V), Video output (VGA)

General

Display

9.4-inch color LCD (Resolution: 640 x 480 dots)

Optical connector

FC (Standard)

Power requirement

AC 100 to 120/200 to 240 V, 50/60 Hz

Operating temperature

5 to 40 °C

Storage temperature

\-10 to +50 °C

Humidity

80 %RH or less (No condensation)

Dimensions

Approx. 425 (W) x 222 (H) x 450 (D) mm

Mass

Approx. 30 kg

Notes

1)

Horizontal scale: wavelength display mode

2)

Vertical scale: absolute power display mode, resolution: 0.05 nm or more, resolution correction: OFF

3)

At 15 to 30 °C, with 10/125 µm single mode fiber, after 2 hours of warm-up, after optical alignment

> **Important:** Maximum measurable input power: +20 dBm (100 mW).

> Recommended pairing: use the Ando AQ4321 Tunable Laser Source for high-speed synchronous sweep, which achieves higher wavelength resolution and wider dynamic range together with the AQ6317B.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 1275.0 USD
- **Vendor:** Ando
- **Type:** Spec An
- **Tags:** Analyzers, Ando, Ando Electric Co., AQ6317, AQ6317B, Ltd., Manufacturers, Model 6317, Optical, Optical Spectrum Analyzer, Pre-Owned, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 1125.00 USD | In stock |
| NIST Traceable with Full Data | 1275.00 USD | In stock |

## Images

- AQ6317B Ando Optical Analyzer

---

> Source: [ValueTronics](https://valuetronics.com/products/aq6317b-ando-optical-analyzer-used)
> Updated: 2026-06-02