# Anritsu MS4623A 6 GHz Transmission/Reflection Vector Network Measurement System (Pre-Owned)

# Anritsu MS4623A 6 GHz Transmission/Reflection Vector Network Measurement System (Pre-Owned)

## Key Features At A Glance

-   Frequency coverage from 10 MHz to 9 GHz across the MS4622x (3 GHz), MS4623x (6 GHz), and MS4624x (9 GHz) family
-   Available in 2-port, 3-port, and 4-port configurations to address transmission/reflection, S-parameter, and balanced/differential measurements
-   Measurement speed of 150 microseconds per point with dynamic range of 125 dB and S-parameter uncertainty under 0.05 dB
-   Source power to +10 dBm and receiver noise as low as -115 dBm for active component characterization
-   Mixed-mode S-parameters, embedding/de-embedding, and arbitrary impedance features standard on 3 and 4-port models
-   Integrated measurement suite spans S-parameters, harmonics (2nd through 9th), IMD (3rd/5th/7th/9th), noise figure (50 MHz to 6 GHz), and frequency-translated group delay
-   AutoCal modules simplify 2, 3, and 4-port calibrations; N-type, 3.5 mm, or GPC-7 connector options available
-   External SCSI interface, sequence automation, and enhanced markers support manufacturing workflows

The Anritsu MS4624D is part of the Scorpion MS462xx family of Vector Network Measurement Systems, a complete RF measurement platform spanning 10 MHz to 9 GHz. The Scorpion family is offered in three frequency ranges — the MS4622x to 3 GHz, the MS4623x to 6 GHz, and the MS4624x to 9 GHz — with each frequency range available in multiple port configurations to match the device under test. The MS4624D specifically is the 4-port, 9 GHz configuration, designed to characterize balanced and differential components alongside conventional single-ended devices.

Anritsu positions the Scorpion as a single instrument capable of testing the full range of wireless components and systems, including antennas, isolators, filters, duplexers, couplers, SAW filters, baluns, amplifiers, mixers, power amplifiers, tower mount amplifiers, and multiport integrated components. With a single connection to the device under test, the instrument can perform S-parameter measurements, time domain and distance-to-fault testing, compression and 20 dB power sweeps, harmonics from the 2nd through the 9th order, noise figure across 50 MHz to 6 GHz, intermodulation distortion at the 3rd, 5th, 7th, and 9th orders, and group delay through frequency translation. Typical applications described by the manufacturer include handset front-end module characterization, base station infrastructure components, mixer and amplifier development, and multiport SAW filter and integrated circuit production test.

ValueTronics International stocks and tests Anritsu Scorpion-family vector network measurement systems at our 20,000 square foot secure facility at 1675 Cambridge Drive in Elgin, Illinois. Every {{CONDITION}} MS4624D that ships from our facility is inspected and performance-verified by our technicians before it leaves the warehouse, with condition documentation describing the specific unit a customer will receive. Customers ordering through ValueTronics work directly with a named contact who handles the transaction from quote through delivery and post-sale support, rather than navigating an anonymous catalog process.

## Brand Heritage

Anritsu Corporation is a Japanese precision measurement company with roots extending to 1895. The Scorpion vector network analyzer line traces in part to Wiltron Company, a US RF instrument maker that Anritsu acquired in 1993, which gave Anritsu critical RF and microwave measurement technology and a North American commercial presence. Anritsu has continued to develop the platform independently as part of its broader RF and microwave measurement portfolio.

## Compare Other Models in This Series

The Scorpion MS462xx family is organized along two independent axes: frequency range and port count. The frequency-range axis is denoted by the second digit of the model number — MS4622x covers 10 MHz to 3 GHz, MS4623x covers 10 MHz to 6 GHz, and MS4624x covers 10 MHz to 9 GHz. The port-count axis is denoted by the trailing letter — the A suffix denotes a 2-port transmission/reflection configuration, B denotes a full-featured 2- or 3-port S-parameter configuration, C denotes the configuration used in the ME7840A Power Amplifier Test System, and D denotes the 4-port balanced/differential configuration.

Within each frequency range, Anritsu describes the family as upgradable: a customer can specify only the frequency range and port count required today and add capability later as requirements change. The A-family targets economical transmission/reflection production test, the B-family adds 2- and 3-port S-parameter measurements with embedding/de-embedding and arbitrary impedance, and the D-family extends to 4-port balanced/differential measurements with the full mixed-mode S-parameter matrix.

Each pre-owned Scorpion configuration listed below is its own product page with condition-matched pricing and inventory. The comparison table on this page is intended to help locate the configuration that matches the frequency range and port count required for the application; the linked product pages provide the unit-specific detail.

The MS4624D is the 4-port, 9 GHz configuration at the top of the Scorpion range. Among the family, it is the configuration that supports the full set of balanced and differential measurement features described in the datasheet — including mixed-mode S-parameters across the complete D1D1 through C2C2 matrix, 4-port AutoCal-based calibration, and time domain analysis on 4-port devices — while also providing the widest frequency coverage offered in the family.

Across the family, the differences are primarily in frequency coverage and port count rather than in measurement architecture. All configurations share the same measurement speed, the same source power capability, the same AutoCal compatibility, and the same general feature set within the limits of their port count. Stepping from a 2-port to a 3-port or 4-port configuration adds standard mixed-mode S-parameter measurement, true 3- or 4-port calibration, and the full embedding/de-embedding feature set as documented in the datasheet.

Stepping up in frequency range (3 GHz to 6 GHz to 9 GHz) extends the upper measurement frequency without changing the lower limit of 10 MHz. Customers should refer to the comparison table that follows for a side-by-side view of the configurations and select the model whose frequency range and port count match the device under test.

Model

Frequency Range

Number of Ports

Configuration Type

**MS4623A**

10 MHz to 6 GHz

2

Transmission/Reflection (T/R)

MS4622A

10 MHz to 3 GHz

2

Transmission/Reflection (T/R)

MS4622B

10 MHz to 3 GHz

2 or 3

Full-Featured S-Parameter

MS4622C

10 MHz to 3 GHz

n (ME7840A PA Test System component)

Power Amplifier Test System Configuration

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Specification

Value

General Performance

Frequency Range (MS4622x)

10 MHz to 3 GHz

Frequency Range (MS4623x)

10 MHz to 6 GHz

Frequency Range (MS4624x)

10 MHz to 9 GHz

Measurement Speed

150 µsec/point

Dynamic Range

125 dB (typical)

Source Power

Up to +10 dBm; adjustable between +7 and –85 dBm (B/D models)

Receiver Noise

As low as –115 dBm (typical, B/D models)

S-Parameter Uncertainty

< 0.05 dB

Connector Options

N-Type, 3.5 mm, or GPC-7

Corrected Test Port Performance

Directivity, 10–1000 MHz

\> 46 dB

Source Match, 10–1000 MHz

\> 44 dB

Load Match, 10–1000 MHz

\> 46 dB

Directivity, 1000–3000 MHz

\> 44 dB

Source Match, 1000–3000 MHz

\> 41 dB

Load Match, 1000–3000 MHz

\> 44 dB

Directivity, 3000–6000 MHz

\> 38 dB

Source Match, 3000–6000 MHz

\> 39 dB

Load Match, 3000–6000 MHz

\> 38 dB

Directivity, 6000–9000 MHz

\> 37 dB

Source Match, 6000–9000 MHz

\> 36 dB

Load Match, 6000–9000 MHz

\> 37 dB

Raw Test Port Performance

Directivity Raw, 10–1000 MHz

23 dB

Port Match Raw, 10–1000 MHz

15 dB (MS462xD with Noise Figure Options 4F or 4G degraded below 100 MHz)

Directivity Raw, 1000–3000 MHz

23 dB

Port Match Raw, 1000–3000 MHz

15 dB

Directivity Raw, 3000–6000 MHz

20 dB

Port Match Raw, 3000–6000 MHz

15 dB

Directivity Raw, 6000–9000 MHz

15 dB

Port Match Raw, 6000–9000 MHz

9 dB

Amplifier Measurement Performance

S-Parameters Accuracy

< ± 0.05 dB

Gain Compression

As high as +16 dBm

Noise Figure

Less than 0.5 dB, 50 MHz to 6 GHz

Third Order Intercept

Up to +40 dBm

Mixer Measurement Performance

Compression

20 dB Power Sweep, +7 dBm Source Power

Group Delay Trace Noise

Up to 100 nS peak-to-peak Resolution

IF Frequency Range (with option)

Receivers extendable down to 10 kHz

Calibration / Error Correction

Error Models

12-term (2-port), 24-term (3-port), 40-term (4-port)

Calibration Method

Manual (opens, shorts, loads, thru-lines) or AutoCal (3658 series modules)

Measurement Capabilities

S-Parameters

10 MHz to 9 GHz (frequency-range dependent)

Time Domain / Distance to Fault

Supported

Compression

20 dB Power Sweeps

Harmonics

2nd through 9th, automatic

Noise Figure

50 MHz to 6 GHz

IMD

3rd, 5th, 7th, and 9th order, automatic

FTGD

Group Delay through Frequency Translation

Mixed-Mode S-Parameters

Standard on 3 and 4-port configurations

Embedding/De-embedding

Supported (transmission lines, L-C circuit primitives, SnP data files)

Arbitrary Impedance

Supported

Interfaces & Software

External Storage

External SCSI Interface

Software

Scorpion Navigator (Windows PC); ActiveX modules for Visual C++, Visual Basic, LabView, TestStand

**Important:** Specified noise figure coverage is 50 MHz to 6 GHz. For MS462xD models with Noise Figure Options 4F or 4G, raw port match performance is degraded below 100 MHz.**Important:** MS462xD models with Noise Figure Options 4F or 4G have degraded port match raw performance below 100 MHz (per the datasheet's Corrected and Raw VNA performance footnote).**Important:** For Intermodulation Distortion (IMD) measurements with an external power combiner, the Scorpion uses its two internal sources; no external synthesizer is required for basic IMD setups.**Important:** Mixer characterization measurements requiring LO drive use an external Anritsu MG3690A synthesizer (sold separately).**Important:** The ME7840A Power Amplifier Test System is the integrated turnkey solution based on the MS462xC; the MS462xC is intended to be used as part of this system.**Important:** For frequencies higher than 9 GHz, the datasheet directs users to the Anritsu 37000 Family of Microwave Vector Network Analyzers.**Important:** Calibration kits, AutoCal modules, and high-power test sets required for full measurement capability (e.g., power amplifier testing up to 100 W) are not part of the base instrument and must be ordered separately.Recommended pairing: for mixer, up/down converter, and frequency-translating measurements requiring external LO drive or multi-tone stimulus, the datasheet specifies the Anritsu MG3690A synthesizer; the Scorpion's Multiple Source Control can coordinate up to four such synthesizers.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Anritsu
- **Type:** Network Analyzer
- **Tags:** Analyzers, Anritsu, Manufacturers, Network, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- Anritsu MS4623A 6 GHz Transmission/Reflection Vector Network Measurement System (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/ms4623a-anritsu-network-analyzer-used)
> Updated: 2026-05-30