# Fluke 734A DC Reference Standard (Pre-Owned)

# Fluke 734A DC Reference Standard (Pre-Owned)

## Key Features At A Glance

-   Four electrically and mechanically independent 732B DC Standards in a rack-width enclosure
-   10V output stability of ±2 ppm per year and 1.018V stability of ±0.8 ppm per month
-   10V output drives up to 12 mA to simplify use with low-input-impedance instrumentation
-   Each 732B operates 72 hours on internal battery; extendable to 130 hours with optional external battery and charger
-   Output protection: indefinite short-circuit protection; 10V input protected up to 1100V DC at 25 mA
-   Temperature coefficient ≤0.04 ppm/°C (10V) and ≤0.1 ppm/°C (1.018V) across 15–35°C
-   Compliance with ANSI ISA-S2, CSA C22.2 #231, IEC 348, IEC 1010, and UL 1244

The Fluke 734A DC Reference Standard is a direct voltage reference used to maintain the volt in primary and secondary standards laboratories. The system consists of four electrically and mechanically independent 732B DC Standards housed in a rack-width enclosure, with each 732B providing both 10V and 1.018V outputs from its own power supply, oven, supporting electronics, and packaging.

The 734A was designed for laboratories that need to maintain traceability to national standards and to distribute the volt to production, service, calibration laboratories, or other remote locations. Each 732B is small and highly portable, with a 72-hour internal battery life that supports transporting the volt across town or around the world while the reference is maintained undisturbed in the laboratory. The 1.018V output simplifies intercomparisons of saturated standard cell banks in applications where those are still in use.

ValueTronics International stocks, tests, and ships the 734A DC Reference Standard and individual 732B DC Standards from our 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Unlike drop-ship brokers and pass-through resellers, every unit in our inventory is held, inspected, and prepared in-house — so when a primary or secondary standards lab orders a 734A from us, the unit ships from a facility that has physically handled it, not from a listing routed to a third party.

## Brand Heritage

Fluke Corporation is a long-established US manufacturer of precision test, measurement, and calibration instrumentation, headquartered in Everett, Washington. Fluke became part of Danaher Corporation in 1998, and in 2016 Danaher spun off several of its industrial businesses — including Fluke — as Fortive Corporation, which remains Fluke's current parent.

## Compare Other Models in This Series

The 734A family covers the maintenance and dissemination of the DC volt at the primary and secondary standards laboratory level. The 734A is the full rack-width system containing four electrically and mechanically independent 732B DC Standards, and the 732B is the same standard offered individually for customers who need fewer cells or who want to add capacity to an existing 734A enclosure.

Each 732B provides both a 10V and a 1.018V output and includes its own power supply, oven, supporting electronics, and packaging. This independence is what allows the four cells in a 734A to be intercompared without common-mode error sources, and it is what allows a single 732B to be transported to a remote location while the rest of the reference remains undisturbed in the laboratory.

Each pre-owned model below is its own dedicated product page with condition-matched pricing, inventory status, and warranty terms. Click through to the model that matches your application — the 734A for a complete four-cell rack-width reference, or the 732B for an individual DC Standard.

The 734A is purchased when a laboratory needs the complete four-cell architecture in a single rack-width enclosure from the outset — for example, when establishing a new primary or secondary standards capability, or when the four-cell intercomparison approach described in NBS Technical Note 1239 is the operating model from day one. The 734A-7001 enclosure is also available separately for customers who are building up to four cells over time.

The two listings differ in form factor and quantity rather than in measurement performance. A single 732B delivers the same 10V output (±2 ppm per year stability, 12 mA drive capability) and the same 1.018V output (±0.8 ppm per month stability) as any one of the four cells inside a 734A — they are the same instrument. The 734A simply integrates four of them into the rack-width 734A-7001 enclosure and supports the four-cell intercomparison architecture from a single purchase.

The comparison table that follows lists the cell count, output voltages, output drive current, stability specifications, battery operating time, and physical dimensions for each configuration. Use the table to confirm which form factor matches your laboratory's intended reference architecture and rack space.

Model

Output Voltages

10V Stability (1 Year)

Configuration

**734A**

10V and 1.018V

±2 ppm/year

Four 732Bs in rack-width enclosure

732B

10V and 1.018V

±2 ppm/year

Standalone portable standard

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Four 732B DC Standards
-   Rack-width enclosure (734A-7001)

## Product Core & Specifications

Specification

732B / 734A

Output Voltages

Outputs

10.0V and 1.018V at separate binding-post pairs

Stability (at Tcal ±1°C, 15°C ≤ Tcal ≤ 35°C)

10V — 30 Days

0.3 ppm

10V — 90 Days

0.8 ppm

10V — 1 Year

2.0 ppm

1.018V — 30 Days

0.8 ppm

Noise at Output Terminals

10V — Day-to-Day (S1)

0.065 ppm

10V — 0.01 Hz to 10 Hz

0.06 ppm rms

1.018V — Day-to-Day (S1)

0.1 ppm

1.018V — 0.01 Hz to 10 Hz

0.3 ppm rms

Output Current and Impedance

10V — Output Current

12 mA (limited to ≤0.1 mA for 72-hour battery operation)

10V — Output Impedance

≤1 milliohm

1.018V — Output Impedance

≤1 kohm

Output Adjustment

10V

0.15 ppm resolution

1.018V

Set at nominal ±1 mV; no adjustment provided (changes when 10V output is adjusted)

Retrace (Hysteresis) Error (Constant Temperature)

10 minute power off

≤0.1 ppm

1 to 24 hours power off

≤0.25 ppm

Stabilization Time (Constant Temperature)

Extended power-off period

14-day stabilization recommended

No power interruption

None required after moving environments

Power off <1 hour

1-hour warm-up

Power off 1 to 24 hours

24-hour warm-up

Temperature Coefficient of Output (15°C to 35°C)

10V

TC ≤0.04 ppm/°C

1.018V

TC ≤0.1 ppm/°C

Load Regulation (10V Output)

0 to 12 mA load change

<1 ppm

0 to 2 mA load change

<0.1 ppm

Line Regulation

10% line voltage change or full battery range

≤0.05 ppm

Output Protection

Short circuit

Indefinite, no damage

External voltage ≤220V dc

Protected up to 50 mA continuous

External voltage ≤1100V dc

Protected up to 25 mA continuous or up to 0.6 joules

Environment

Normal Operation

15 to 35°C, 15-80% RH, 0-6,000 ft

Safe Operation

0 to 50°C, 15-90% RH, 0-10,000 ft

Storage (no battery)

\-40 to 50°C, non-condensing, 0-40,000 ft

Line Power Requirements

Voltage

90-132V ac or 180-264V ac

Frequency

47-63 Hz

Battery Operation

Battery life (fully charged, 23°C ±5°C, output <0.1 mA)

Minimum 72 hours

Recharge time

<24 hours with self-contained charger

External DC input

Rear-panel; accepts 12V to 15V dc source

Isolation

Output terminal to earth ground or ac mains

At least 10,000 MΩ shunted by less than 1000 pF

Guard and Grounding

Chassis ground

Terminals on front and rear panels

Internal guard access

Front-panel terminal

Compliance with External Standards

Standards

ANSI ISA-S2, CSA C22.2 #231, IEC 348, IEC 1010, UL 1244

Mechanical — 732B

Height

5.3 inches (13.5 cm)

Width

3.9 inches (9.9 cm)

Depth

16.5 inches (41.9 cm)

Weight

13 lb (5.9 kg)

Mechanical — 734A

Height

7.5 inches (19.1 cm) with feet; 7.0 inches (17.8 cm) without feet

Width

17.0 inches (43.2 cm)

Depth

19.75 inches (50.2 cm)

Weight

65 lb (29.6 kg)

**Important:** Allow a 14-day stabilization period after extended power-off for best results. After power off less than 1 hour, a 1-hour warm-up is required; after power off 1 to 24 hours, a 24-hour warm-up is required.**Important:** The 732B-200 on-site calibration service is not available in all areas; contact your local Fluke representative for details.**Important:** For best results, allow a 14-day stabilization period after power has been off for an extended period of time. After power off <1 hour, a 1-hour warm-up is required; after power off 1 to 24 hours, a 24-hour warm-up is required.**Important:** 10V output current is limited to ≤0.1 mA to realize 72-hour operation under battery power.Recommended pairing: combine a 732B with the 742A-1 and 742A-10k Resistance Standards to build the artifact calibration support package for the Fluke 5700A or 5720A Calibrator or the HP 3458A Multimeter named in the datasheet.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Fluke
- **Type:** Dc Volt Std
- **Tags:** Fluke, Manufacturers, Other Equipment, product_manual_group_1, Product_used, Standard

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- Fluke 734A DC Reference Standard (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/734a-fluke-standard-used)
> Updated: 2026-05-30