# Agilent N9030A PXA X-Series Signal Analyzer, 3 Hz to 3.6 GHz (Option 503) (Pre-Owned)

# Agilent N9030A PXA X-Series Signal Analyzer, 3 Hz to 3.6 GHz (Option 503) (Pre-Owned)

## Key Features At A Glance

-   Frequency coverage from 3 Hz up to 3.6, 8.4, 13.6, 26.5, 43, 44, or 50 GHz depending on option (503 through 550)
-   LAN 1000Base-T, GPIB (IEEE-488), and USB 2.0 interfaces for automated test system integration
-   Analysis bandwidth of 10 MHz standard, expandable to 25, 40, or 140 MHz with Options B25, B40, or B1X
-   Noise Floor Extension (NFE) lowers the displayed average noise level for low-level signal measurements
-   RF input connector by option: Type-N female, APC 3.5 mm (Option C35), or 2.4 mm male
-   PowerSuite measurements including channel power, ACP, occupied bandwidth, CCDF, and spectrum emission mask
-   Optional preamplifiers (Options P03 through P50) extend sensitivity across the frequency bands
-   Removable 80 GB solid-state drive with a 1024 x 768 XGA display, 8.4 in diagonal

The Agilent N9030A PXA is an X-Series signal analyzer positioned as a future-ready, evolutionary replacement for an existing high-performance analyzer. It spans frequencies from 3 Hz to as high as 50 GHz depending on the configured frequency option, and is built to reveal new levels of signal detail with outstanding RF performance while refreshing legacy systems with a highly compatible replacement.

The PXA is intended for demanding applications in aerospace, defense, and commercial communications. Its stated value is helping engineers sustain past achievements, enhance current designs, and accelerate future innovations — increasing test throughput and protecting existing system investments as measurement needs grow.

ValueTronics stocks test and measurement equipment in its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. That combination of real inventory and hands-on verification is what separates a stocking specialist from a listing aggregator.

## Brand Heritage

Agilent Technologies was spun out of Hewlett-Packard in 1999 as a dedicated test and measurement company. In 2014, Agilent's electronic measurement business was separated again to form Keysight Technologies, which carries the line forward today.

## Compare Other Models in This Series

The N9030A PXA is offered as an X-Series signal analyzer family in which frequency reach is set by the chosen frequency option. All configurations share the same 3 Hz low-frequency limit, while the upper frequency extends to 3.6 GHz (Option 503), 8.4 GHz (508), 13.6 GHz (513), 26.5 GHz (526), 43 GHz (543), 44 GHz (544), or 50 GHz (550).

Across the family, the core architecture — PowerSuite measurements, the I/Q analyzer, Noise Floor Extension, and the standard interface set — is consistent. What changes between configurations is primarily the frequency span and the associated RF front-end hardware.

Each pre-owned configuration below is its own dedicated product page with condition-matched pricing. Select the frequency option that matches your application's upper frequency requirement to reach the corresponding page.

Within the family, the configured frequency option determines not only the analyzer's upper frequency but also the available preamplifier coverage and the front-panel RF input connector type.

The configurations differ chiefly in upper frequency limit, which steps from 3.6 GHz through 50 GHz across Options 503 to 550. The standard RF input connector is Type-N female on the 3.6 to 26.5 GHz options (503, 508, 513, 526) and 2.4 mm male on the millimeter-wave options (543, 544, 550); an APC 3.5 mm input is available via Option C35 with Option 526.

Preamplifier coverage also tracks the frequency option: Options P03 through P50 extend from 9 kHz up to 3.6, 8.4, 13.6, 26.5, 43, 44, or 50 GHz respectively. See the comparison table for the per-option frequency ranges, connector types, and related differences.

Model

Max Frequency

Min Frequency

RF Input Connector

**N9030A**

3.6 GHz

3 Hz

Type-N female

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Product Core & Specifications

Specification

Value

Frequency and Time

Frequency range (DC coupled)

3 Hz to maximum frequency of instrument (option dependent; up to 50 GHz)

Frequency range (AC coupled)

10 MHz to maximum frequency of instrument

Precision frequency reference aging rate

± 1 x 10–7 / year; ± 1.5 x 10–7 / 2 years

Temperature stability (20 to 30 °C)

± 1.5 x 10–8

Temperature stability (full temperature range)

± 5 x 10–8

Achievable initial calibration accuracy

± 4 x 10–8

Marker frequency counter resolution

0.001 Hz

Frequency span range

0 Hz (zero span), 10 Hz to maximum frequency of instrument

Frequency span resolution

2 Hz

Sweep time range (span = 0 Hz)

1 µs to 6000 s

Sweep time range (span ≥ 10 Hz)

1 ms to 4000 s

Sweep trigger sources

Free run, line, video, external 1, external 2, RF burst, periodic timer

Sweep (trace) point range

1 to 40001

Resolution bandwidth range (–3.01 dB)

1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz

RBW selectivity (–60 dB/–3 dB)

4.1:1 nominal

EMI bandwidth (CISPR compliant)

200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required)

EMI bandwidth (MIL STD 461E compliant)

10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required)

Video bandwidth range

1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)

Maximum analysis bandwidth (standard)

10 MHz

Maximum analysis bandwidth (Option B25)

25 MHz

Maximum analysis bandwidth (Option B40)

40 MHz

Maximum analysis bandwidth (Option B1X)

140 MHz

Amplitude Accuracy and Range

Measurement range

Displayed average noise level (DANL) to maximum safe input level

Input attenuator range (3 Hz to 50 GHz)

0 to 70 dB in 2 dB steps

Electronic attenuator (Option EA3) range

0 to 24 dB, 1 dB steps; full range (mechanical + electronic) 0 to 94 dB, 1 dB steps

Maximum safe input — average total power

+30 dBm (1 W)

Maximum safe input — peak pulse power

+50 dBm (100 W), < 10 µs pulse width, < 1 % duty cycle, input attenuation ≥ 30 dB

Maximum DC (DC coupled)

± 0.2 Vdc

Maximum DC (AC coupled)

± 100 Vdc

Display range (log scale)

0.1 to 1 dB/division in 0.1 dB steps; 1 to 20 dB/division in 1 dB steps (10 divisions)

Scale units

dBm, dBmV, dBµV, dBmA, dBµA, V, W, A

Total absolute amplitude accuracy at 50 MHz

± 0.24 dB (± 0.19 dB 95th percentile)

Reference level range (log scale)

–170 to +30 dBm in 0.01 dB steps

Reference level range (linear scale)

707 pV to 7.07 V with 0.11 % (0.01 dB) resolution

Display scale fidelity (–10 to –80 dBm input mixer level)

± 0.10 dB total (± 0.04 dB typical)

Trace detectors

Normal, peak, sample, negative peak, log power average, RMS average, voltage average

Preamplifier gain (9 kHz to 3.6 GHz)

+20 dB nominal

Preamplifier gain (3.6 to 26.5 GHz)

+35 dB nominal

Preamplifier gain (26.5 to 50 GHz)

+40 dB nominal

Displayed Average Noise Level (DANL) — Preamp off, Normal / LNP enabled (input terminated, 0 dB attenuation, 20 to 30 °C)

3 Hz to 9 kHz

–146 dBm

9 to 100 kHz

–150 dBm

100 kHz to 1 MHz

–155 dBm

1 to 10 MHz

–155 dBm

10 MHz to 1.2 GHz

–153 dBm

1.2 to 2.1 GHz

–152 dBm

2.1 to 3.0 GHz

–151 dBm

3.0 to 3.6 GHz

–147 dBm / –153 dBm

3.5 to 4.2 GHz

–150 dBm / –155 dBm

4.2 to 8.4 GHz

–149 dBm / –155 dBm

8.3 to 13.6 GHz

–145 dBm / –152 dBm

13.5 to 16.9 GHz

–143 dBm / –151 dBm

16.9 to 20.0 GHz

–137 dBm / –150 dBm

Dynamic Range

1 dB gain compression — 20 to 40 MHz (max power at input mixer)

–3 dBm (0 dBm typical)

1 dB gain compression — 40 to 200 MHz

+1 dBm (+3 dBm typical)

1 dB gain compression — 200 MHz to 3.6 GHz

+3 dBm (+5 dBm typical)

1 dB gain compression — 3.6 to 16 GHz

+1 dBm (+4 dBm typical)

1 dB gain compression — 16 to 26.5 GHz

–1 dBm (+2 dBm typical)

Third-order intermodulation (TOI) — 0.6 to 1.1 GHz

+20 dBm (+22 dBm typical)

Third-order intermodulation (TOI) — 1.1 to 3.6 GHz

+21 dBm (+23 dBm typical)

Third-order intermodulation (TOI) — 3.5 to 8.4 GHz

+15 dBm (+22 dBm typical)

Third-order intermodulation (TOI) — 8.3 to 13.6 GHz

+15 dBm (+23 dBm typical)

Second harmonic distortion (SHI) — 0.1 to 1.8 GHz

+45 dBm

Second harmonic distortion (SHI) — 1.75 to 2.5 GHz

+62 dBm / +80 dBm

Residual responses (input terminated, 0 dB attenuation, 200 kHz to 8.4 GHz)

–100 dBm

Phase Noise — typical (CF = 1 GHz, 20 to 30 °C)

10 Hz offset

–75 dBc/Hz nominal

100 Hz offset

–100 dBc/Hz typical

1 kHz offset

–125 dBc/Hz typical

10 kHz offset

–132 dBc/Hz typical

30 kHz offset

–132 dBc/Hz typical

100 kHz offset

–131 dBc/Hz typical

> **Important:** Specifications associated with Options 543, 544, and 550 (43, 44, and 50 GHz frequency ranges) are either preliminary or not yet available.

> **Important:** Option C35 (APC 3.5 mm RF input connector) is available with Option 526 only.

> **Important:** LNP (Low Noise Path) requires Option LNP; at frequency bands above 3.6 GHz, Preamp “On” supersedes “LNP enabled” — LNP cannot operate simultaneously with the preamp.

> **Important:** When Option MPB (microwave preselector bypass) is installed and enabled, some aspects of the analyzer performance change; refer to the PXA specification guide for details.

> **Important:** This data sheet is a summary of the complete specifications and conditions; the complete PXA Signal Analyzer Specification Guide is available from Agilent.

> Recommended pairing: for deep waveform capture, use the Agilent 89600B vector signal analysis (VSA) software or the N9064A, as noted in the I/Q analyzer data acquisition specifications.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 34971.0 USD
- **Vendor:** Agilent Technologies
- **Type:** Sig Ana
- **Tags:** Agilent, Agilent Keysight, Agilent Technologies, Analyzers, Inc., Keysight, Keysight Technologies, Manufacturers, Model 9030, N9030, N9030A, Pre-Owned, product_manual_group_1, product_manual_group_2, Product_used, Signal, Spec An, Updated Item Review

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 34971.00 USD | In stock |
| NIST Traceable | 0.00 USD | In stock |
| NIST Traceable with Full Data | 0.00 USD | In stock |
| ISO IEC 17025 Accredited | 0.00 USD | In stock |

## Images

- Keysight Agilent N9030A-550 PXA - Signal Analyzer - Used

---

> Source: [ValueTronics](https://valuetronics.com/products/n9030a-agilent-signal-analyzer-used)
> Updated: 2026-06-02