**Additional Features:**

-   Dynamic Range: 0 to 200% of range up to maximum rated output
-   Maximum Output Current: 100A DC or RMS AC
-   DC Accuracy: (180 days 25°C ± 5°C) all ranges <100A ±0.015% of range ±0.03% of output 100 Amp Range: ± .03% of range ±0.03% of output
-   AC Accuracy: (180 days)( 25°C ± 5°C) ±0.15% of range ± 0.1% of output to 100 Hz ±0.2% of range ± 0.2% of output to 400 Hz ±0.3% of output ± 0.3% of range to 1 kHz
-   Input Impedance: 100 Kilohms
-   Input/Output Ratio: 2.00000 volt input provides full scale output (1.0000 volt input produces 100.0 amps output)
-   Stability: ±0.0005% - 15 min. (±0.005% for 24 Hours)
-   Maximum Input: 3 VDC or 2 VAC (RMS)
-   Response Time: 1 millisecond to ±0.01% of final value following input amplitude or frequency change
-   Maximum Isolation Voltage: ± 200 VDC or Peak AC
-   Input Common Mode Rejection Ratio: 60db @ DC linearly decreasing to 40 db @ maximum frequency
-   Load Regulation: Output current changes less than ± .002%
-   Temperature Range: 0°C to 50°C
-   Temperature Coefficient: ±0.001% of output ±0.002% of range/°C below 20°C and above 30°C (double above for temperature coefficient for AC)
-   Compliance Voltage: 4 Volts
-   Size: 266mm/10" L x 432mm/17" W x 584mm/23" D
-   Weight: 45 Kg/100 lbs. Net 66 Kg/140 lbs. shipping wt.
-   Power: 115/230 VAC ± 10% 50-60 Hz 2000 watts

## Details

- **Price:** 0.0 USD
- **Vendor:** Valhalla Scientific
- **Type:** Calibrator Oth
- **Tags:** Calibrator, Manufacturers, Other, Other Equipment, product_manual_group_1, Product_used, Valhalla Scientific

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 0.00 USD | In stock |

## Images

- 2555A Valhalla Scientific Calibrator Used

---

> Source: [ValueTronics](https://valuetronics.com/products/2555a-valhalla-scientific-calibrator-used)
> Updated: 2026-05-21