# Rohde & Schwarz FSIQ7 20 Hz to 7 GHz Signal Analyzer (Pre-Owned)

# Rohde & Schwarz FSIQ7 20 Hz to 7 GHz Signal Analyzer (Pre-Owned)

## Key Features At A Glance

-   Three models spanning 20 Hz to 3.5 GHz, 7 GHz, or 26.5 GHz coverage across the series
-   Combined frequency-, time-, and modulation-domain analysis in a single instrument
-   Integrated vector signal analyzer covering 18 mobile radio standards from GSM and NADC through WCDMA/3GPP
-   13 digital demodulators from BPSK and QPSK through (G)MSK and 16QAM; symbol rate up to 6.4 Msymbol/s
-   Resolution bandwidths 1 Hz to 10 MHz in 1/2/3/5 steps; FFT filters 1 Hz to 1 kHz for fast measurements
-   True RMS detector for repeatable power measurements regardless of signal waveform
-   GPIB (IEC/IEEE 488.2) and RS-232-C standard for automation; Ethernet interface available as an option
-   24 cm (9.5") TFT colour display at VGA resolution with up to 25 sweeps/s update rate

The R&S FSIQ is a signal analyzer that combines spectrum analysis, time-domain analysis, and modulation analysis in a single unit. It performs comprehensive measurement functions across the frequency, time, and modulation domains, with an integrated vector signal analyzer that handles both digitally and analog modulated signals. The series comprises three models distinguished by frequency coverage: the R&S FSIQ3 (20 Hz to 3.5 GHz), the R&S FSIQ7 (20 Hz to 7 GHz), and the R&S FSIQ26 (20 Hz to 26.5 GHz).

The R&S FSIQ is built for measurements on digital and analog mobile radio systems. In the frequency domain it measures intermodulation and harmonics and is suited to adjacent-channel power ratio (ACPR) measurements across mobile radio systems, with particular emphasis on WCDMA — it attains a WCDMA ACPR value of 75 dB in 4.096 MHz integration bandwidth, already reached at a −12 dBm input level. The instrument also supports burst analysis in TDMA systems such as GSM and NADC, power ramp (power-time template) measurements, transmitter transient measurement, and measurement of incidental phase modulation and AM-to-φM conversion.

ValueTronics stocks and holds its own test-and-measurement inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is sourced and fulfilled directly from that facility; pre-owned units are inspected and functionally verified in-house by our technicians before they ship, while new units ship factory-sealed exactly as received from the manufacturer.

## Brand Heritage

Rohde & Schwarz is a privately held company headquartered in Munich, Germany. It has operated continuously under the Rohde & Schwarz name with no major corporate rebrand or acquisition affecting the brand.

## Compare Other Models in This Series

The R&S FSIQ is offered as three models that share the same one-box architecture for frequency-, time-, and modulation-domain analysis and differ primarily in upper frequency limit. The R&S FSIQ3 covers 20 Hz to 3.5 GHz, the R&S FSIQ7 covers 20 Hz to 7 GHz, and the R&S FSIQ26 covers 20 Hz to 26.5 GHz, with all three sharing the integrated vector signal analyzer, resolution bandwidths of 1 Hz to 10 MHz, and the 24 cm TFT colour display.

Because the three models share the same measurement functions, the selection between them is driven mainly by the frequency coverage an application requires, along with the model-specific performance values noted in the comparison table below.

Each model below links to its own dedicated product page with condition-matched pricing. Select the pre-owned listing for the specific frequency range your application requires, and review that page for its full per-model specifications.

The models differ in upper frequency limit (3.5 GHz, 7 GHz, and 26.5 GHz) and in several model-specific specifications. The minimum 5 ms full-span sweep time applies to the R&S FSIQ3 and R&S FSIQ7. The third-order intercept point is specified at +20 dBm (typ.) for the R&S FSIQ7 and +22 dBm (typ.) for the R&S FSIQ26, and total level uncertainty and frequency-response error limits widen at the higher frequency bands.

The R&S FSIQ26 uses an N/3.5 mm adapter connector system at its RF input and supports external and harmonic mixers (40 GHz to 110 GHz options) for measurements above its base range, and it ships with N-female and 3.5 mm-female test-port adapters. Power consumption and weight also rise across the series, from 195 VA / 24 kg for the R&S FSIQ3 to 245 VA / 26.5 kg for the R&S FSIQ26. Refer to the comparison table for the exact per-model values.

Model

Frequency Range

3rd-Order Intercept (TOI)

SSB Phase Noise (10 kHz offset, f ≤500 MHz)

**FSIQ7**

20 Hz to 7 GHz

TOI >15 dBm (20 dBm typ.)

< –114 dBc(1 Hz)

FSIQ3

20 Hz to 3.5 GHz

TOI >12 dBm (18 dBm typ.)

< –120 dBc(1 Hz)

FSIQ26

20 Hz to 26.5 GHz

TOI >17 dBm (22 dBm typ.)

< –114 dBc(1 Hz)

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Keyboard
-   Mouse
-   Power cable
-   Operating manual
-   Spare fuses

## Product Core & Specifications

Specification

Value

Frequency

Frequency range

20 Hz to 7 GHz

Display range for frequency axis

0 Hz, 10 Hz to 7 GHz

Spectral purity (dBc(1 Hz)) SSB phase noise, f ≤500 MHz

Carrier offset 100 Hz

< –81

Carrier offset 1 kHz

< –100

Carrier offset 10 kHz

< –114

Carrier offset 100 kHz

< –113

Carrier offset 1 MHz

< –132

Intermodulation

3rd-order Intercept (TOI), Δf >5 x RBW or 10 kHz

\>70 dBc for f >150 MHz (TOI >15 dBm, 20 dBm typ.)

Second harmonic intercept point (SHI)

\>25 dBm, >35 dBm typ. for f <150 MHz; >40 dBm, >45 dBm typ. for f >150 MHz

Displayed average noise level (DANL) (0 dB RF attenuation, RBW 10 Hz, VBW 1 Hz, 20 averages, trace average, span 0 Hz, termination 50 Ω)

20 Hz

<–74 dBm

1 kHz

<–104 dBm

10 kHz

<–119 dBm

100 kHz

<–129 dBm

1 MHz

<–142 dBm, –145 dBm typ.

10 MHz to 6 GHz

<–142 dBm, –147 dBm typ.

6 GHz to 7 GHz

<–139 dBm

Maximum dynamic range

1 dB compression to DANL (RBW 1 Hz)

165 dB

Immunity to interference

Image rejection

\>80 dB, >90 dB typ.

Intermediate frequency

\>75 dB

Other interfering signals (mixer level <10 dBm)

<–75 dB

Level (frequency response, 10 dB RF atten.)

Absolute error limit at 120 MHz

<0.3 dB

<2.2 GHz

<0.5 dB

2.2 GHz to 7 GHz

<1 dB

Total measurement error limit (20°C to 30°C, RBW 5 kHz to 1 MHz, signal 0 dB to 70 dB below ref. level)

<2.2 GHz

<1 dB

2.2 GHz to 7 GHz

<1.5 dB

Digital demodulation residual error (EVM and magnitude error, f <1 GHz)

Symbol rate ≤30 kHz

0.7% rms

30 kHz to 300 kHz

1.4% rms

300 kHz to 1 MHz

2.8% rms

1 MHz to 4.2 MHz

2% rms

4.2 MHz to 6.4 MHz

2.4% rms

Phase error (f <1 GHz)

Symbol rate ≤30 kHz

0.4° rms

30 kHz to 300 kHz

0.7° rms

300 kHz to 1 MHz

2° rms

1 MHz to 4.2 MHz

2° rms

4.2 MHz to 6.4 MHz

2.8° rms

Inputs and outputs (front panel)

RF input

N female, 50 Ω

VSWR (RF attenuation ≥10 dB), f <3.5 GHz

<1.5

VSWR (RF attenuation ≥10 dB), f <7 GHz

<2.0

General data

Power consumption

210 VA

Weight

24.5 kg

> **Important:** Recommended calibration interval: 1 year (2 years for operation with an external reference).

> **Important:** Options R&S FSE-B12 (Switchable Attenuator for Tracking Generator) and R&S FSE-B13 (1 dB Attenuator) cannot be installed simultaneously.

> **Important:** The Removable Hard Disk R&S FSE-B18 cannot be retrofitted; it is factory-fitted only.

> **Important:** The cdmaOne and WCDMA/3GPP application firmware (R&S FSIQ-K71/-K72/-K73) require the DSP and I/Q Memory Extension R&S FSIQ-B70; additional modifications may be required if the R&S FSIQ-B70 is retrofitted.

> **Important:** EDGE Application Firmware Extensions R&S FSE-K20/-K21 require R&S FSE-K10/-K11 respectively.

> Recommended pairing: add the optional Tracking Generator with I/Q Modulator — R&S FSE-B9 for the R&S FSIQ3, or R&S FSE-B11 for the R&S FSIQ7/26 — to generate AM signals with very low incidental FM/φM.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Rohde & Schwarz
- **Type:** Spec An
- **Tags:** Analyzers, FSIQ7, Manufacturers, Model 7, Pre-Owned, product_manual_group_1, Product_used, Rohde & Schwarz, Rohde & Schwarz GmbH & Co. KG, Spec An, Spectrum, Updated Item Review

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 0.00 USD | In stock |
| NIST Traceable with Full Data | 0.00 USD | In stock |
| ISO IEC 17025 Accredited | 0.00 USD | In stock |

## Images

- FSIQ7 Rohde & Schwarz Spectrum Analyzer

---

> Source: [ValueTronics](https://valuetronics.com/products/fsiq7-rohde-schwarz-spectrum-analyzer-used)
> Updated: 2026-06-03