# Agilent 54853A 2.5 GHz Infiniium Oscilloscope, 4-Channel, 20 GSa/s (Pre-Owned)

# Agilent 54853A 2.5 GHz Infiniium Oscilloscope, 4-Channel, 20 GSa/s (Pre-Owned)

## Key Features At A Glance

-   Bandwidth from 2 GHz to 6 GHz across the four-model 54850 Series (54852A, 54853A, 54854A, 54855A)
-   Sample rate of 20 GSa/s on all four channels simultaneously (10 GSa/s on the 54852A)
-   Standard 262 kpts per channel acquisition memory; optional 1 Mpts per channel at full sample rate and 32 Mpts per channel at 2 GSa/s or slower
-   Electronic attenuators replace mechanical attenuator relays to improve reliability and repeatability
-   Trigger jitter of 1.0 ps rms on the 54855A
-   AutoProbe interface and 18 GHz BNC-compatible input connector support InfiniiMax probing systems and SMA adapters
-   Windows XP Pro operating system with 10/100 Mbps LAN, GPIB (IEEE 488.2), USB, RS-232, and dual-monitor video output
-   Optional jitter analysis, serial data analysis, and protocol/compliance software (EZJIT, EZJIT Plus, SDA, USB, PCI Express, SATA, and others) available as licensed options

The Agilent Infiniium 54850 Series is a family of four-channel real-time digital oscilloscopes spanning 2 GHz to 6 GHz of analog bandwidth. The datasheet identifies four models in the series: the 54852A (2 GHz), 54853A (2.5 GHz), 54854A (4 GHz), and 54855A (6 GHz). All four models are four-channel instruments built on the same Infiniium platform, with the 54855A serving as the top-of-line bandwidth tier.

According to the datasheet, the 54850 Series targets applications driven by increased clock speeds and edge rates, high-speed serial differential bus design, and dense circuit boards that require probing in hard-to-reach locations. The datasheet's application table lists supported bus standards including Ethernet, USB 2.0, DDR I/II, SATA 1.5 Gbps, SAS 150, DVI, HDMI, Fibre Channel, PCI Express I, ExpressCard, InfiniBand, and Advanced TCA, with recommended bandwidths and corresponding compliance, decode, and jitter analysis software options.

## Brand Heritage

The Infiniium 54850 Series was introduced by Agilent Technologies, the test and measurement business spun out of Hewlett-Packard in 1999. In 2014, Agilent's electronic measurement business was transferred to a new entity, Keysight Technologies, which is the current corporate owner of the Infiniium oscilloscope product family. Used instruments bearing the Agilent badge are part of the same engineering lineage as current Keysight-branded products.

## Compare Other Models in This Series

The Infiniium 54850 Series consists of four four-channel real-time oscilloscope models that share a common Infiniium platform, user interface, and connectivity feature set. As the datasheet specification tables show, the four models are differentiated primarily by analog bandwidth and, on the 54852A, by maximum sample rate. All four share the same Windows XP Pro operating environment, the same AutoProbe interface, the same 18 GHz BNC-compatible input connector, and the same options for memory upgrade, jitter analysis, and serial data analysis software.

The 54855A sits at the top of the bandwidth ladder at 6 GHz (with the option 008 enhanced bandwidth software enabling 7 GHz typical, per the datasheet). The 54854A is the 4 GHz model, the 54853A is the 2.5 GHz model, and the 54852A is the 2 GHz model. Bandwidth tier is the primary application-driving difference within the family.

Each pre-owned model in the Infiniium 54850 Series is listed on its own product page with condition-matched pricing, installed options, and configuration details specific to the unit on hand. Buyers selecting between models in the series should first confirm the bandwidth required by their application, then review the per-model listing for the installed memory option, jitter and compliance software options, and accessory complement.

Within the family the 54855A is differentiated by its 6 GHz analog bandwidth, 70 ps rise time, and 1.0 ps rms trigger jitter, and by the inclusion of a high-performance calibration cable as standard. The datasheet also lists an option 008 enhanced bandwidth software for the 54855A that increases measurement bandwidth to 7 GHz typical or allows the user to reduce the scope bandwidth to 1 GHz to lower system noise.

Across the family, the key per-model specifications break out as follows in the datasheet: analog bandwidth at 6 GHz (54855A), 4 GHz (54854A), 2.5 GHz (54853A), and 2 GHz (54852A); rise time at 70 ps, 105 ps, 168 ps, and 210 ps respectively; trigger jitter at 1.0, 1.3, 1.7, and 1.8 ps rms respectively; and sample rate at 20 GSa/s on the three higher-bandwidth models versus 10 GSa/s on the 54852A. All four models share the same standard memory of 262 kpts per channel and access the same optional memory upgrade (1 Mpts at full rate, 32 Mpts at 2 GSa/s or slower).

The comparison table that follows this section reflects the per-model specification differences described in the datasheet. Application requirements such as the maximum signal frequency content, required edge rise time, and any compliance test bandwidth recommendation (the datasheet lists recommended bandwidths from 2 GHz to 6 GHz for the listed serial bus standards) should guide which model in the series is appropriate.

Model

Bandwidth

Channels

Sample Rate

**54853A**

2.5 GHz

4

20 GSa/s

54855A

6 GHz

4

20 GSa/s

54854A

4 GHz

4

20 GSa/s

54852A

2 GHz

4

10 GSa/s

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

## Accessories Supplied

-   Optical USB mouse
-   Compact keyboard
-   User's quick-start guide
-   Documentation CD (service guide, programmer's guide, programmer's quick reference guide)
-   Accessory pouch
-   Power cord
-   E2655B probe deskew and performance verification kit
-   One-year warranty

## Product Core & Specifications

Category

Specification

Value

Vertical

Input channels

4

Vertical

Input impedance

50 Ω ± 2.5%

Vertical

Sensitivity

1 mV/div to 1 V/div

Vertical

Input coupling

DC

Vertical

Vertical resolution

8 bits, ≥ 12 bits with averaging

Vertical

Channel to channel isolation

DC to 100 MHz: 40 dB; 100 MHz to 1 GHz: 28 dB; > 1 GHz to 6 GHz: 24 dB

Vertical

DC gain accuracy

± 1% of full scale at full resolution channel scale

Vertical

Maximum input voltage

± 5 V

Vertical

Offset range

\> ± 12 div or ± 4 Volts, whichever is smallest

Vertical

Offset accuracy

± (2% of channel offset + 1% of full scale)

Vertical

Dynamic range

± 4 div from center screen

Horizontal

Main timebase range

54855A and 54854A: 5 ps/div to 20 s/div; 54853A and 54852A: 10 ps/div to 20 s/div

Horizontal

Main timebase delay range

−200 s to 200 s

Horizontal

Delayed timebase range

1 ps/div to current main time scale setting

Horizontal

Channel deskew

−50 µs to 150 µs range, 100 fs resolution

Horizontal

Time scale accuracy

± 1 ppm pk

Acquisition

Standard memory depth

262,144 points per channel at all sample rates

Acquisition

Option 001 memory depth

1,025,000 at all sample rates; 32,800,000 at ≤ 2 GSa/s sample rate

Trigger

Auxiliary sensitivity

DC to 500 MHz: 500 mV p-p

Trigger

Internal level range

± 8 div from center screen or ± 4 Volts, whichever is smallest

Trigger

Auxiliary level range

± 5 V

Trigger

Sweep modes

Auto, triggered, single

Trigger

Holdoff range

80 ns to 320 ms

Display

Display

8.4 inch diagonal color TFT-LCD

Display

Resolution

640 x 480 pixels

Computer

Operating system

Windows XP Pro

Computer

CPU

Intel Pentium III 1 GHz

Computer

PC system memory

512 MB

Computer

Drives

≥ 20 GB internal hard drive, CD-R drive on rear panel, 3.5 inch 1.44 MB floppy drive

I/O Ports

LAN

RJ-45, 10Base-T and 100Base-T

I/O Ports

GPIB

IEEE 488.2, fully programmable

I/O Ports

RS-232

COM1

I/O Ports

Parallel

Centronics printer port

I/O Ports

PS/2

2 ports

I/O Ports

USB

2 ports

I/O Ports

Video output

15 pin VGA

I/O Ports

Dual-monitor video output

15 pin XGA

I/O Ports

Trigger output

5 V, 50 Ω back-terminated

I/O Ports

Time base reference output

10 MHz, 5 V, 50 Ω back-terminated

Environment

Operating temperature

5°C to +40°C

Environment

Non-operating temperature

−40°C to +70°C

Environment

Operating humidity

Up to 95% RH (non-condensing) at +40°C

Environment

Operating altitude

Up to 4,600 meters (15,000 feet)

Power

Line input

100-240 VAC, ± 10%, CAT II, 47 to 440 Hz

Power

Maximum power dissipated

475 W

Physical

Net weight

13 kg (28.5 lbs.)

Physical

Shipping weight

16 kg (35.2 lbs.)

Physical

Height

216 mm (8.5 in)

Physical

Width

437 mm (17.19 in)

**Important:** Maximum input voltage: ±5 V (warranted specification per the 54850 Series datasheet). Confirm the InfiniiMax 1130 Series probe head or other accessory in use matches the signal voltage range for the application; the 1130 Series probe amplifier itself specifies a maximum input of 30 V peak, CAT I.**Important:** No probes are included with the 54850 Series oscilloscopes. The InfiniiMax 1130 Series probes must be purchased separately.**Important:** Maximum input voltage is ± 5 V. Exceeding this limit may damage the oscilloscope.**Important:** Specifications are valid after a 30-minute warm-up period and ±5°C from annual calibration temperature.Recommended pairing: the InfiniiMax 1130 Series probe amplifiers (1134A, 1132A, 1131A, 1130A) and their probe heads are listed in the 54850 Series datasheet as the matched probing system. The datasheet also states no probes are included with the oscilloscope and must be ordered separately.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD
- **Vendor:** Agilent
- **Type:** Dig Oscilloscope
- **Tags:** Agilent, Agilent Keysight, Digital, Manufacturers, Oscilloscopes, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 0.00 USD | In stock |
| NIST Traceable with Full Data | 0.00 USD | In stock |
| ISO IEC 17025 Accredited | 0.00 USD | In stock |

## Images

- Agilent 54853A 2.5 GHz Infiniium Oscilloscope, 4-Channel, 20 GSa/s (Pre-Owned)

---

> Source: [ValueTronics](https://valuetronics.com/products/54853a-agilent-digital-oscilloscope-used)
> Updated: 2026-05-30