# HP 54510A 250 MHz 1 GSa/s Digitizing Oscilloscope (Pre-Owned)

# HP 54510A 250 MHz 1 GSa/s Digitizing Oscilloscope (Pre-Owned)

## Key Features At A Glance

-   250 MHz bandwidth (DC-coupled, -3dB) with 1.4 ns rise time for general-purpose waveform analysis
-   Maximum sample rate of 1 GSa/s with 8-bit vertical resolution (10-bit via HP-IB with averaging)
-   Two simultaneous input channels with 8,000-point real-time waveform record length per channel
-   Five trigger modes — Edge, Pattern, State, Delay, and TV — for capturing complex digital and analog events
-   Selectable 1 MΩ or 50 Ω input impedance on Channel 1, Channel 2, and External Trigger inputs
-   Repetitive and real-time acquisition modes with pan-and-zoom across the full 8 k acquisition memory
-   Full HP-IB programmability per IEEE 488.2-1988 with hardcopy output to compatible HP printers and plotters
-   Four nonvolatile waveform memories, two volatile pixel memories, and four nonvolatile front-panel setup memories

The HP 54510A is a 250 MHz, two-channel digitizing oscilloscope designed as a general-purpose repetitive and real-time instrument with full HP-IB programmability. The instrument provides 1 GSa/s maximum sample rate, 8-bit vertical resolution, and 8,000 points of waveform memory per channel in real-time mode, along with an external trigger input and a comprehensive set of triggering, measurement, and waveform-storage features.

According to the published instrument description, the HP 54510A is suited to applications ranging from high-speed automated test environments on the HP-IB bus through device characterization in research and development laboratories. The datasheet calls out specific applications including capture of non-recurring events such as microprocessor start-up sequences, debugging of glitches that disrupt system operation, characterization of high-speed digital signals, and triggering on clamped composite video signals using the 60 Hz/525-line, 50 Hz/625-line, or user-defined TV trigger modes.

## Brand Heritage

The HP 54510A was originally manufactured by Hewlett-Packard. Hewlett-Packard spun out its test and measurement business as Agilent Technologies in 1999, and Agilent subsequently transferred the electronic measurement portfolio to Keysight Technologies in 2014. Original HP-branded 54510A units are functionally identical to those that may carry Agilent service documentation; service and calibration support for legacy HP instruments of this generation routes through the current Keysight lineage where available, and through accredited third-party calibration laboratories for instruments outside factory support.

## Accessories Supplied

-   Two HP 10441A miniature passive probes (2 meters)
-   One miniature probe to BNC male adapter (HP 1250-1454)
-   One 2.3 meter (7.5 feet) power cord
-   One Front-Panel Reference and Programming Reference Set
-   One Service Manual

## Product Core & Specifications

Category

Specification

Value

Vertical

Bandwidth (-3dB, dc coupled)

dc to 250 MHz

Vertical

Rise Time

1.4 ns

Vertical

Input R (selectable)

1 MΩ ±1% or 50 Ω ±1%

Vertical

Maximum Input Voltage (1 MΩ)

±250 V \[dc + peak ac(<10 kHz)\]

Vertical

Maximum Input Voltage (50 Ω)

5 Vrms

Vertical

Offset Accuracy

±(1% of channel offset + 2% of full scale)

Vertical

Voltage Measurement Accuracy (Dual Cursor)

±(1.25% of full scale + 0.032 × V/div)

Vertical

Voltage Measurement Accuracy (Single Cursor)

±(1.25% of full scale + offset accuracy + 0.016 × V/div)

Vertical

Number of Channels

2 (simultaneous acquisition)

Vertical

Vertical Sensitivity Range

1 mV/div to 5 V/div

Vertical

Vertical Gain Accuracy (dc)

±1.25% of full scale

Vertical

Vertical Resolution

8 bits over 8 divisions (±0.4%), 10 bits via HP-IB with averaging (±0.1%)

Vertical

Maximum Sample Rate

1 GSa/s

Vertical

Waveform Record Length

8001 points real time, 501 points repetitive

Vertical

Input C

7 pF nominal

Vertical

Input Coupling

ac, dc

Vertical

Channel-to-channel Isolation (dc to 50 MHz)

40 dB

Vertical

Channel-to-channel Isolation (50 to 250 MHz)

30 dB

Vertical

Bandwidth Limit (HF Reject)

30 MHz (-3 dB)

Vertical

LF Reject

450 Hz (-3 dB)

Vertical

ac Coupling

90 Hz (-3 dB)

Horizontal

Time Base Range

1 ns/div to 5 s/div

Horizontal

Time Base Resolution

20 ps

Horizontal

Delta-t Accuracy (Repetitive, ≥8 averages)

±(0.005% × delta-t + 2E-6 × delay setting + 100 ps)

Horizontal

Delta-t Accuracy (Real Time, single acquisition)

±(0.005% × delta-t + 2E-6 × delay setting + 150 ps)

Horizontal

Delay Range (post-trigger)

10,000 × (s/div)

Trigger

Trigger Sensitivity (Internal, dc to 50 MHz)

0.5 division

Trigger

Trigger Sensitivity (Internal, 50 MHz to 250 MHz)

1.0 division

Trigger

Trigger Sensitivity (External, dc to 250 MHz)

100 mVp-p into 50 Ω

Trigger

Trigger Pulse Width (Internal)

1.75 ns minimum

Trigger

Trigger Pulse Width (External)

2.8 ns minimum

Trigger

Trigger Level Range (Internal)

±1.5 × full scale from center of screen

Trigger

Trigger Level Range (External)

±2 V

Trigger

Trigger Modes

Edge, Pattern, State, Delay, TV

Trigger

Trigger Holdoff (time)

40 ns - 320 ms

Trigger

Trigger Holdoff (events)

2 - 16,000,000

Display

Data Display Resolution

451 points horizontally by 256 points vertically

Display

Number of Screens

1 or 2

Display

Display Modes

Normal, Averaged, Envelope

Display

Averaging Range

1 to 2048 in powers of 2

Display

Variable Persistence

500 ms to 10 seconds, or infinite

Memory

Nonvolatile Setup Memories

4

Memory

Nonvolatile Waveform Memories

4

Memory

Volatile Pixel Memories

2

Memory

HP-IB Multiple Single-shot Capability

300 k RAM, up to 500 Waveforms per Second

Programming

Interface

Full HP-IB Programmability (IEEE-488), IEEE 488.2-1988 compliant

Programming

Data Transfer Rate

Approximately 120 kBytes per second

Environmental

Operating Temperature

0°C to +55°C (32°F to +131°F)

Environmental

Non-operating Temperature

\-40°C to +70°C (-40°F to +158°F)

Environmental

Operating Humidity

up to 95% relative humidity (non-condensing) at +40°C

Environmental

Non-operating Humidity

up to 90% relative humidity at +65°C

Environmental

Operating Altitude

up to 4,600 meters (15,000 ft)

Environmental

Non-operating Altitude

up to 15,300 meters (50,000 ft)

Power

Voltage

115/230 V ac, -25% to +15%, 48-66 Hz

Power

Power

350 VA maximum

Physical

Net Weight

approximately 10 kg (22 lb)

Physical

Shipping Weight

approximately 20 kg (44 lb)

**Important:** Maximum input voltage: ±250 V (DC + peak AC <10 kHz) on the 1 MΩ setting, and 5 Vrms on the 50 Ω setting. On vertical sensitivity ranges of 50 mV/div or below, maximum input overdrive must not exceed 100 V.**Important:** This product is marked discontinued by the manufacturer (Agilent/HP). Agilent no longer sells or supports this product; service centers may be able to perform calibration if no repair parts are needed, but no other support from Agilent is available.**Important:** This is a Safety Class 1 instrument with an exposed chassis that is directly connected to earth via the power supply cord to meet IEC Standard 348. A three-wire grounded power cable must be used.**Important:** Before applying power to the instrument, be sure the fuse module is set to the correct line voltage position. Severe damage will occur if the line voltage is not properly set.**Important:** On vertical ranges ≤50 mV/div, the maximum overdrive of the input must not exceed 100 V.**Important:** All TV trigger modes require a clamped video signal for stable triggering. Use the HP 1133A TV/Video Sync Pod to provide clamped video output for use with the HP 54510A's TV triggering capabilities.**Important:** Voltage measurement accuracy decreases 0.08% per °C from firmware calibration temperature; specification is valid for a temperature range ±10° from software-calibration temperature.**Important:** Upper bandwidth reduces by 2.5 MHz for each °C above 35°C.Recommended pairing: the HP 1133A TV/Video Sync Pod, which the datasheet identifies as the means of providing the clamped video output required for stable triggering across all TV trigger modes on the HP 54510A.

## About this used unit

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 262.5 USD
- **Vendor:** HP
- **Type:** Dig Oscilloscope
- **Tags:** Agilent, Agilent Keysight, Digital, Manufacturers, Oscilloscopes, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 187.50 USD | In stock |
| NIST Traceable with Full Data | 225.00 USD | In stock |
| ISO IEC 17025 Accredited | 262.50 USD | In stock |

## Images

- HP 54510A 250 MHz 1 GSa/s Digitizing Oscilloscope (Pre-Owned)

---